YS/T 683-2008 is the English-translated version of 压力(差压)变送器现场校准规范.
YS/T 683-2008 — This standard applies to include positive and negative pressure, differential and absolute pressure transmitters for field calibration.
Found what you need?
Document preview — YS/T 683-2008
National Standard of the People's Republic of China
- ICS
- 17.100
- Classification
- N 11
Issued by: National Development and Reform Commission
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions
- 4 Overview
- 5 Metrological characteristics
- 5.1 Measurement error
- 5.2 Return Error
- 5.3 Static error
- 5.4 Stability
- 5.5 Insulation resistance
- 6 Calibration conditions
- 6.1 Environmental conditions
- 6.2 Staff
Foreword
This standard by the National Standardization Technical Committee of non-ferrous metals and centralized.
The Standard Appendix A, Appendix B and Appendix C are informative appendices.
This standard by the Jinchuan Group Ltd. is responsible for drafting.
This standard by the Aluminum Corporation of China Henan Branch, Liuzhou China Tin Group Co., Ltd., Chinalco Luoyang Copper Co.
Division, Yunnan Copper Co., Ltd., Southwest Aluminum (Group) Co., Ltd. and Zhuzhou Smelter Group Co., Ltd. participated in the drafting.
1 Scope
This standard applies to include positive and negative pressure, differential and absolute pressure transmitters for field calibration.
2 Normative references
The following documents contain provisions which, through reference in this standard and become the standard terms. For dated references, subsequent Amendments (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement are based on research Whether the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.
JJF1059-1999 Evaluation and Expression of Uncertainty in Measurement
3 Terms and Definitions
In measured on the same point, the absolute value of the difference between the actual value of the down stroke, the return is an error at this point.
4 Overview
Pressure variable pressure transmitter is converted to the corresponding normalized output signal can be transmitted meter. Mainly used for industrial process pressure (Or differential pressure) measured parameters.
Pressure transmitter consists of a pressure-sensitive components, signal processing and conversion units, according to the principle can be divided into capacitive, piezoresistive and so on.
Electric and pneumatic pressure transmitter has two categories. Normalized output signal electric pressure transmitter is 0mA ~ 10mA or DC signal 4mA ~ 20mA; the pneumatic pressure transmitter output signal is standardized gas pressure signal is 20kPa ~ 100kPa.
5 Metrological characteristics
5.1 Measurement error
The maximum pressure transmitter measurement error shall not exceed the maximum permissible error.
5.2 Return Error
Return error pressure transmitter must not exceed the maximum permissible error of absolute value.
5.3 Static error
Static differential pressure transmitter errors shall not exceed the maximum permissible error.
5.4 Stability
Pressure (DP) transmitter drift must not exceed the range specified in the instructions.
5.5 Insulation resistance
Insulation resistance of each terminal pressure (DP) transmitter (including housing) should be not less than 20Mohm. Of the terminals is not isolated Do insulation test.
6 Calibration conditions
6.1 Environmental conditions
Environmental conditions should be corrected to meet the normal requirements of the instrument, the standard working conditions should be consistent with the site environmental conditions, on-site calibration should be considered due to environmental values deviate from the uncertainty of the impact of the amount shown.
6.2 Staff
Calibration should be qualified and trained personnel hold the appropriate qualifications.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 9 pages — is available in the English PDF.
How to Download YS/T 683-2008
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YS/T 1115.10-2026 — Chemical analysis methods of copper, lead and zinc ores and tailings - Part 10: Determination of gold, silver, molybdenum, arsenic, tungsten, bismuth, gallium, indium, thallium, selenium, tellurium and germanium contents - Inductively coupled plasma mass spectrometry
YS/T 1115.7-2026 — Methods for chemical analysis of copper, lead, zinc ores and tailings - Part 7: Determination of molybdenum content - Thiocyanate spectrophotometric method
YS/T 1115.8-2026 — Methods for chemical analysis of copper, lead, zinc ores and tailings - Part 8: Determination of fluorine content - Ion-selective electrode method and ion chromatography
Secure payment via Stripe
Payments accepted
YS/T 683-2008
$195.00