Scope

WS/T 648-2019 (General hygienic requirement for air disinfecting machine) is available as an English-translated PDF.

WS/T 648-2019 — This standard specifies the classification, naming and model, technical requirements, inspection methods, of use, nameplates and instructions for air sterilizers.

This standard applies to air sterilizers that are sterilized using physical factors, chemical factors and other factors.

Found what you need?

Document preview — WS/T 648-2019

National Standard of the People's Republic of China

ICS
11.080
Classification
C 50

Issued by: National Health Commission

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Classification
  • 4.1 Classification by working principle
  • 4.2 Classification by installation method
  • 5 Name and model
  • 5.1 Naming
  • 5.2 Model
  • 6 Technical requirements
  • 6.1 General requirements
  • 6.2 Physical factor air disinfection machine technical requirements
  • 6.3 Technical requirements for air disinfection of chemical factors
  • 6.4 Other factors of air sterilizer technical requirements
  • 6.5 Disinfection effect requirements
  • 6.6 Security requirements
  • 7 Detection methods
  • 7.1 Air disinfection simulation field test
  • 7.2 Air disinfection field test
  • 7.3 Toxicological safety test
  • 7.4 Working noise
  • 7.5 Sterilizer electrical safety indicators
  • 7.6 Circulator air volume
  • 8 How to use
  • 9 Nameplates and instructions

Foreword

This standard was drafted in accordance with the rules given in GB/T 1.1-2009.

This standard was drafted. Heilongjiang Provincial Center for Disease Control and Prevention, China Center for Disease Control and Prevention, Environmental and Health-related Product Safety Institute, Chinese People's Liberation Army Center for Disease Control and Prevention, Jiangsu Provincial Center for Disease Control and Prevention, Jiangsu Provincial Health Supervision Institute, Hebei Province Disease Prevention and Control Center, Shandong Provincial Center for Disease Control and Prevention.

1 Scope

This standard specifies the classification, naming and model, technical requirements, inspection methods, of use, nameplates and instructions for air sterilizers.

This standard applies to air sterilizers that are sterilized using physical factors, chemical factors and other factors.

2 Normative references

The following documents are indispensable for the application of this document. For dated references, only the dated version applies to this document.

For undated references, the latest edition (including all amendments) applies to this document.

GB 9706.1 Medical electrical equipment - Part 1. General requirements for safety GB/T 14294 modular air conditioning unit GB 28232 Ozone Generator Safety and Hygiene Standard GB 28235 UV air sterilizer safety and hygiene standards GBZ/T 189.8 workplace physical factor measurement noise Disinfection Technical Specification (2002 Edition) Ministry of Health (Wei Fa Jian Fa [2002] No. 282) Management Specification for Labeling Instructions for Disinfecting Products (2005 Edition) Ministry of Health (Wei Supervision and Development [2005] No. 426)

3 Terms and definitions

The following terms and definitions apply to this document.

Air disinfecting machine Use physical, chemical or other methods to kill or remove microorganisms in indoor air, and to achieve disinfection requirements, with independent power, energy Stand-alone device.

Natural decay rate

The percentage of bacteria in the air that naturally decrease or naturally die.

Extinction rate

The percentage of bacteria in the air that are naturally decaying and disinfected.

Killing rate

In the microbial killing test, the percentage indicates the decrease in the number of microorganisms.

4 Classification

4.1 Classification by working principle

4.1.1 Air disinfection machine with physical factors. use electrostatic adsorption, filtration technology and ultraviolet light to kill or remove microorganisms in the air.

An air sterilizer that meets the disinfection requirements can be used for indoor air disinfection in some cases. Such as electrostatic adsorption air sterilizer, high efficiency filtration (HEPA), ultraviolet air sterilizer, etc.

4.1.2 Air disinfection machine for chemical factors. use the chemical factors produced to kill microorganisms in the air to achieve the disinfection requirements of the air disinfection machine, it is only used for disinfection of indoor air in the absence of people. Such as chlorine dioxide air disinfection machine, ozone air disinfection machine, hydrogen peroxide air disinfection machine, Peracetic acid air disinfection machine, etc.

4.1.3 Other factors of air disinfection machine. use other factors to kill microorganisms in the air, to achieve the disinfection requirements of the air disinfection machine, such as Ion body air sterilizer, photocatalyst air sterilizer, etc.

4.2 Classification by installation method

Can be divided into wall-mounted, cabinet, mobile cabinet, embedded air disinfection machine.

5 Name and model

5.1 Naming

It shall comply with the provisions of the "Management Specifications for Labeling Instructions for Disinfecting Products".

5.2 Model

Product Name - B --1--2 Product Name - G --1--2 Product Name - Y --1--2 Product Name - Q --1--2 Description. B--wall coded air sterilizer code G-cabinet air disinfection machine code Y-mobile air sterilizer code Q-embedded air sterilizer code-1-Common name-2-property name Example. XX card-Q--1--2 Indicates XX brand embedded 300 ozone air disinfection machine

6 Technical requirements

6.1 General requirements

6.1.1 Appearance

The appearance of the air sterilizer should be smooth and flat, and the components should be firmly connected. It can work safely in normal use, and will not cause people and weeks.

Environmental hazards.

6.1.2 Effective life

Ultraviolet air sterilizer should meet the requirements of life in GB 28235; ozone air sterilizer should meet the effective life of GB 28232 Seeking; other air sterilizers should meet the requirements of the corresponding standards, and there should be no corresponding standards to meet the requirements of product quality standards.

6.1.3 Working noise

The working noise of the UV air sterilizer should meet the noise requirements of GB 28235; the air sterilizer without the corresponding standard runs.

It should be stable and reliable, without vibration, and the noise is <=60dB (A-weighted).

6.1.4 Circulating air volume

An air sterilizer that relies on circulating air volume for disinfection purposes, the circulating air volume should be more than 8 times the applicable volume;

The air sterilizer should meet the requirements of circulating air volume in GB 28235.

6.1.5 Applicable volume

The volume must not be less than 30 m3.

6.1.6 Electrical safety

Should meet the requirements of GB 9706.1.

6.2 Physical factor air disinfection machine technical requirements

6.2.1 The technical requirements for raw materials, and components of UV air sterilizers shall comply with the requirements of GB 28235; other air disinfection the raw materials, and components of the machine should meet the requirements of the corresponding standards. If there is no corresponding standard, the requirements of the product quality standards should be met.

6.2.2 The ultraviolet lamp irradiation intensity of the UV air sterilizer should meet the requirements of GB 28235.

6.2.3 The amount of ultraviolet leakage and ozone leakage of the UV air sterilizer should meet the requirements of GB 28235.

6.3 Technical requirements for air disinfection of chemical factors

6.3.1 The technical requirements for the ozone generator and its ozone generating unit of the ozone air disinfection machine shall comply with the requirements of GB 28232; other air the raw materials and technical requirements of the sterilizer should meet the requirements of the corresponding standards. If there is no corresponding standard, the requirements of the product quality standards should be met.

6.3.2 Ozone air disinfection machine The concentration of ozone in the disinfection of air should meet the requirements of GB 28232; the chemical causes of other air disinfection machines the sub-concentration should meet the requirements of the corresponding standard, and the corresponding standard should meet the requirements of the product quality standard.

6.3.3 The disinfection time should be <=1h.

6.4 Other factors of air sterilizer technical requirements

6.4.1 The raw materials, components and technical requirements of the air sterilizer should meet the corresponding standards. If there is no corresponding standard, the product quality should be achieved.

The requirements of the quantity standard.

6.4.2 The plasma air disinfection machine shall not be equipped with medium and high efficiency filters and ultraviolet germicidal lamps.

6.4.3 Leakage of harmful substances to human body shall comply with the requirements of the corresponding standards. If there is no corresponding standard, the requirements of product quality standards shall be met.

6.4.4 The disinfection time should be <= 2h.

6.5 Disinfection effect requirements

6.5.1 Air disinfection simulation field test

Air disinfection simulation field test with air sterilizer, under the condition of 20 ° C ~ 25 ° C, relative humidity 50% ~ 70%, booting to the killing rate for Staphylococcus aureus should be >=99.9% at the time specified in the instructions.

6.5.2 Air disinfection field test

Under the natural conditions of the site, the air disinfection machine is used for air disinfection field test, and the start-up effect is to the time specified in the manual, in the air.

The mortality rate of natural bacteria should be >=90.0%.

6.6 Security requirements

6.6.1 The toxicological safety of the air disinfection machine of chemical factors shall comply with the relevant requirements of the Technical Specifications for Disinfection (2002 Edition).

6.6.2 The air disinfection machine of physical factors and other factors shall not release any toxic and harmful substances during operation, and the concentration of ozone in indoor air <=0.16 mg/m3.

7 Detection methods

7.1 Air disinfection simulation field test

Execute according to Appendix A.

7.2 Air disinfection field test

Execute according to Appendix B.

7.3 Toxicological safety test

According to the relevant provisions of the "Disinfection Technical Specifications".

7.4 Working noise

Determined according to the method specified in GBZ/T 189.8.

7.5 Sterilizer electrical safety indicators

Determine according to the method specified in GB 9706.1.

7.6 Circulator air volume

Determine according to the method specified in GB/T 14294.

8 How to use

8.1 According to the instructions of the use of various air disinfection machines, the disinfection effect should meet the requirements of 6.5.

8.2 When using an air sterilizer to sterilize the intended disinfection site, it should be carried out in a closed environment to avoid circulation with outdoor air.

Ensure disinfection.

8.3 When using the ozone air disinfection machine to disinfect indoor air, the indoor relative humidity should be >=70% to ensure the disinfection effect.

8.4 When disinfecting indoor air with an air sterilizer that produces harmful substances such as chlorine dioxide, hydrogen peroxide, peracetic acid, or ozone, it should be carried out under unmanned conditions in the room. After the disinfection is finished, the indoor disinfection factor should be reduced until there is no impact on people (generally more than 30 minutes) can enter the person; when the situation permits, the window can be ventilated, so that the disinfection factor can be diffused and neutralized as soon as possible; pay attention to the protection of indoor items and avoid strong Damage to the item by the oxidant.

9 Nameplates and instructions

It shall comply with the provisions of the "Management Specifications for Labeling Instructions for Disinfecting Products".

Appendix A

(normative appendix) Air disinfection simulation field test A.1 Purpose the simulated test site air is contaminated by artificial spraying of microbial aerosols, and the air disinfection machine is used for air disinfection.

Low safe use of the dose.

A.2 Test equipment and equipment A.2.1 Test strain. Staphylococcus aureus 8032 strain.

A.2.2 Medium. nutrient broth medium, nutrient agar medium, nutrient agar medium containing neutralizer (containing neutralizing agent for attachment) those who have passed the C identification are sampled after disinfection of the air disinfection machine for chemical disinfection factors).

A.2.3 Aerosol chamber for disinfection test. The aerosol chamber should be constructed of smooth, corrosion-resistant and easy-to-clean materials such as stainless steel, aluminum alloy and glass.

A pair of neighbors (both volumes of 20 m3), one for the disinfection test and the other for the test control. The environment in which a pair of aerosol chambers are located (including temperature Degree, relative humidity, light, airtightness, ventilation, etc.) should be consistent. Temperature and relative humidity regulators and ventilator filters should be installed Sterile or other disinfection devices and corresponding pipelines, such as sleeve operation and sample delivery for spray sterilization, drug administration, sampling, etc.

A.2.4 Spraying equipment, including air compressors, pressure gauges, gas flow meters and aerosol sprayers. Spraying bacteria aerosol micro the diameter of the granules should be 90% or more and should be between 1 µm and 10 µm.

A.2.5 Air microbial sampling device. Six-stage mesh air impact sampler, air extraction equipment, gas flow meter, timer, etc.

A.2.6 Environmental monitoring equipment. thermometers, hygrometers, etc.

A.3 Preparation of test bacterial suspension Take fresh stabbed cultures of Staphylococcus aureus from 3rd to 7th generations at 36°C±1°C for 18-24 hours, using TPS (trypeptone saline) Solution) Wash the lawn, filter it with sterile cotton wool, and dilute it to the desired concentration with nutrient broth.

A.4 operating procedures A.4.1 Installation of the air sterilizer to be tested Before the test starts, install the air disinfection machine to be tested at the far end of the test aerosol chamber according to the installation instructions of the air disinfection machine to be tested.

The source is confirmed to be able to work normally, and in the control aerosol chamber, the air disinfection capability is removed and compared with the equipment of the same type to be tested.

Then close the door. Thereafter, all operations and equipment are operated outdoors through a window or remote control with a sealed sleeve. straight the door can only be opened until the end of the test.

A.4.2 Test environment condition setting Turn on the computer control system (or turn on the temperature and humidity adjustment device), and adjust the temperature and relative humidity of the two aerosol chambers to the test.

Required temperature (20 ° C ~ 25 ° C) and relative humidity (50% ~ 70%).

A.4.3 Aerosol spray staining the microbial aerosol generator was fixed on the sampling vehicle in the control and test group aerosol chambers and placed in the center of the aerosol chamber.

The position is 1.0 m from the ground, and the bacteria are sprayed according to different aerosol generating devices to set the pressure, gas flow rate and spray time. Spraying Stir the bacteria with a fan (mixer). After the spray was dyed, the mixture was stirred for 5 min and allowed to stand for 5 min.

A.4.4 Sampling before disinfection After 5 min of rest, the aerosol chambers of the control group and the test group were separately sampled before disinfection, as the control group before the start of the test and the test.

The positive control (ie, the amount of contaminated bacteria) before the group was disinfected. The number of positive control bacteria in the indoor air of aerosol should reach 5x104 CFU/m3~5x105 CFU/m 3 (The number of positive control bacteria in the control group at the last time of the disinfection test shall not be less than 5x104 CFU/m3).

A.4.5 Disinfection treatment According to the instructions for use of the air sterilizer to be tested, start up and run.

A.4.6 Sampling after disinfection the air sterilizer is applied to the first scheduled time (0.5 times of the specified time), and is immediately the same for the test group and the control group.

Sampling is performed; the action is continued until the second predetermined disinfection time (the time specified in the manual), and sampling is performed again as described above.

A.4.7 Sampling requirements the test uses a six-stage sieve air impingement sampler for sampling. When sampling, a six-stage sieve air impingement sampler is placed in the center of the aerosol chamber.

At a height of 1.0 m, the sampling flow rate is 28.3 L/min, and the sampling time is determined according to the prediction test (the general control group and the test group are sampled before the disinfection treatment).

5 s ~ 10 s, the disinfection effect of the test group after disinfection, such as disinfection qualified sampling 5 min ~ 10 min).

A.4.8 Cultivation and observation of results After sampling, the plate was taken out aseptically and placed in a 36 ° C ± 1 ° C incubator for 48 h for viable culture counting. After completing the test group and being positive After the control group was sampled, the same batch of the unused medium was cultured simultaneously with the above two groups of samples as a negative control. If the negative control group has bacteria Growth, indicating that the medium used is contaminated, the test is invalid, and the replacement of the sterile equipment is repeated.

A.4.9 Disinfection of aerosol chamber After the whole test is completed, after the final disinfection of the bacteria on the surface of the aerosol chamber and the air, the fan is turned on, and the bacteria are removed and filtered.

In addition to polluted air trapped inside the aerosol chamber.

A.5 Data Processing

A.5.1 Calculation of bacteria content in the air the amount of bacteria in the air is calculated according to formula (A.1).

T3.28

X (A.1)

In the formula.

X--Air bacteria, in units of colony forming units per cubic meter (CFU/m3);

S--the total number of bacteria on the six-stage sampling plate, the unit is the colony forming unit (CFU);

t--Sampling time in minutes (min).

A.5.2 Calculation of kill rate the disinfection effect of the air disinfection machine on the bacteria is determined by the kill rate (or clearance rate) Kt, and the value is expressed by (%), and is calculated according to the formula (A.3).

N tt (A.2)

VNV

K (A.3)

In the formula.

Kt--killing rate (or clearance rate) of bacteria in the air, %;

V0' and Vt'--the air-containing bacteria in the test group before and during the disinfection process, the unit is the colony forming unit per Cubic meters (CFU/m3);

Nt--the natural decline rate of bacteria in the air;

V0 and Vt--the amount of bacteria in the air before the start of the test and at different times during the test, in units of colony forming units Square meter (CFU/m3).

A.6 Repeat test

The test was repeated 3 times in the same condition.

A.7 Result determination the killing rate (or clearance rate) of the three test results was >=99.9%, which was judged to be qualified for disinfection.

A.8 Precautions

A.8.1 In the test, because the conditions for uniform control are difficult, it is necessary to set the test group and the control group at the same time in each test.

Consistent.

A.8.2 Note the temperature and relative humidity during the test to record the comparison.

A.8.3 The sample should be microbiologically tested as soon as possible to avoid affecting the accuracy of the results.

A.8.4 The aerosol chamber should be adequately ventilated after each test. Disinfect and rinse if necessary, and the second test can be done after 4 hours.

A.8.5 During the test, the aerosol chamber must be kept closed and equipped with an air filter to prevent the contaminated air from polluting the environment.

A.8.6 During the test, the aerosol chamber should be protected from direct sunlight to avoid bactericidal instability.

A.8.7 The filter material in the air chamber exhaust filter device should be replaced regularly. The replaced filter material should be sterilized before other treatment.

Appendix B

(normative appendix) Air disinfection field test B.1 Purpose at the applicable site, in the absence of human conditions, natural bacteria are used as indicator microorganisms, and disinfection sites (such as wards, bedrooms, offices, etc.) can be sealed.

The location of the air disinfection or microbial removal treatment, verify the practical disinfection effect of the air disinfection machine.

B.2 Test equipment and equipment B.2.1 medium Nutrient agar medium, nutrient agar medium containing neutralizer (the neutralizing agent included is qualified according to Appendix C, used for chemical disinfection) Factor air sterilizer is sampled after disinfection).

B.2.2 Air microbial sampling device Six-stage mesh air impact sampler, extraction equipment, gas flow meter, timer, etc.

B.2.3 Environmental monitoring equipment Thermometer, hygrometer, etc.

B.3 Operating procedures B.3.1 Test site selection According to the requirements of the air disinfection machine, select a representative test site (such as ward, bedroom, office, ambulance, etc.) Closed place), and the amount of bacteria in the test site should be >=1000 CFU/m3, and the test should be carried out in the absence of indoors.

B.3.2 Installation of air disinfection machine to be tested Before the start of the test, install the air sterilizer to be tested at the remote end of the selected test site according to the installation instructions of the air sterilizer to be tested.

Good power and confirm that it works.

B.3.3 Sampling before disinfection After the air in the selected closed test site is static for 5 min, the natural bacteria in the air are sampled with a six-stage mesh air impact type sampler.

Pre-disinfection sample (positive control), when sampling, the test site <=10 m2 set a sampling point, the test site >=10 m2, each increase 10 m

2 Add one sampling point and set up to 5 sampling points. One sampling point sampling, the six-stage mesh air impact type sampler is placed in the test site at a height of 1.0m in the center, multiple sampling points place the six-stage mesh air impingement sampler on the diagonal or evenly distributed in the plum blossom, and away from disinfection.

The air outlet of the machine should be >0.5 m from the wall and sampled at a height of 1.0 m. The sampling flow rate is 28.3 L/min, and the sampling time is based on air.

The amount of bacteria is determined, generally not more than 10 minutes.

B.3.4 Disinfection treatment According to the instructions for use of the air sterilizer to be tested, start up and run.

B.3.5 Sampling after disinfection the air sterilizer is applied to the predetermined time (the time specified in the manual), and the sampling point is pre-sterilized with six-stage mesh air impingement sampling.

The same method is used to sample natural bacteria in the air as a test sample after disinfection.

B.3.6 Cultivation and observation of results After sampling, the plate was taken out aseptically and placed in a 36 ° C ± 1 ° C incubator for 48 h for viable culture counting. At the same time, the same batch of unused the nutrient base was cultured simultaneously with the above two groups of samples as a negative control. If the negative control group has bacteria growth, the medium used is contaminated.

The test is invalid and will be re-executed after replacement.

B.3.7 Repeated test

The test was repeated 3 times.

B.4 Data Processing

B.4.1 The amount of bacteria in the air is calculated as in A.5.1.

B.4.2 Calculation of the death rate the disinfection effect of the air disinfection machine on the natural bacteria in the air is measured by the death rate. The value is expressed in (%) and is calculated according to the formula (B.1).

100x

X (B.1)

In the formula.

X--death rate, %;

A--the average number of bacteria in the sample before disinfection, the unit is the colony forming unit per cubic meter (CFU/m3);

B--The average number of bacteria in the sample after disinfection, the unit is the colony forming unit per cubic meter (CFU/m3).

B.5 Result determination the rate of death of natural bacteria in each test was >=90.0%.

B.6 Precautions

B.6.1 In the test, the conditions for uniform control are difficult, so the environmental conditions before and after disinfection and between different times should be kept as consistent as possible.

B.6.2 Note the temperature and relative humidity during the test to record the comparison.

B.6.3 The on-site room should be protected from direct sunlig...

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 10 pages — is available in the English PDF.

Referenced standards

How to Download WS/T 648-2019

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
10 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

WS/T 648-2019

$285.00