JJG 821-2005Verification Regulation of Total Organic Carbon Analyzer (English PDF)
This is a limited preview
Buy now to download the full PDF (16 pages)
Need quotation for your management?
Get a formal quotation for JJG 821-2005 in 30 minutes.
Issued by
General Administration of Quality Supervision, Inspection and Quarantine of PRC
Level / Type
National · Mandatory
Issue date
September 5, 2005
Implementation date
March 5, 2006
Scope
JJG 821-2005 (Verification Regulation of Total Organic Carbon Analyzer) is available as an English-translated PDF.
JJG 821-2005 is the Chinese standard "Verification Regulation of Total Organic Carbon Analyzer". Its scope clause reads: This Regulation applies to the initial verification, subsequent verification, in-use inspection, of the total organic carbon analyzer.
Its clauses include overview; measurement performance requirements; general technical requirements; control of measuring instrument. It was issued by the General Administration of Quality Supervision, Inspection and Quarantine of PRC on 2005-09-05 and took effect on 2006-03-05.
Found what you need?
Document preview — JJG 821-2005
National Standard of the People's Republic of China
Issued by: General Administration of Quality Supervision, Inspection and Quarantine of PRC
Contents
- 1 Scope...5
- 2 Overview...5
- 3 Measurement performance requirements...5
- 3.1 Indication error for inorganic carbon detection...5
- 3.2 Indication error of organic carbon detection...5
- 3.3 Repeatability for inorganic carbon detection...5
- 3.4 Repeatability for organic carbon detection...5
- 4 General technical requirements...6
- 4.1 Visual inspection...6
- 4.2 Insulation resistance and dielectric strength...6
- 5 Control of measuring instrument...6
- 5.1 Verification conditions...6
- 5.2 Verification Items...7
- 5.3 Verification method...8
- 5.4 Processing of verification results...11
- 5.5 Verification period...11
- Appendix A Preparation method of water for standard solution...12
- Appendix B Preparation method of standard solutions...13
- Appendix C Verification record format...14
- Appendix D Inner page format of verification certificate...15
- Appendix E Format of the inner page of the verification result notification letter...16
1 Scope
This Regulation applies to the initial verification, subsequent verification, in-use inspection, of the total organic carbon analyzer.
2 Overview
For the total organic carbon analyzer (hereinafter referred to as TOC analyzer), when measuring the carbon content in a solution or solid state sample, it oxidizes the carbon element in the sample to carbon dioxide; uses the corresponding relationship between carbon dioxide and carbon mass, to obtain the content of carbon in the sample.
The TOC analyzer is mainly composed of the following parts: Sampling device, reactor, gas-liquid separator, non-dispersive infrared carbon dioxide detector, data processing unit.
3 Measurement performance requirements
3.1 Indication error for inorganic carbon detection
The indication error limit of the inorganic carbon detection is ±4% (relative error).
3.2 Indication error of organic carbon detection
The indication error limit of organic carbon detection is ±5% (relative error).
3.3 Repeatability for inorganic carbon detection
The repeatability of inorganic carbon detection shall be better than 3%.
3.4 Repeatability for organic carbon detection
The repeatability of organic carbon detection shall be better than 3%.
4 General technical requirements
4.1 Visual inspection
4.1.1 The instrument has no damage, deformation and other phenomena, that affect its work. The instrument name, model, exit-factory serial number, manufacturing time, manufacturer name, etc. shall be complete and clear. Dangerous parts (high temperature, strong light, etc.) shall be clearly marked. The manufacturer's instruction manual is attached.
4.1.2 The structure of the instrument is complete. The components and cable plug-ins are connected reliably. The accessories are complete. All adjustment knobs, buttons, switches can work normally. The indication scale or digital display of the instrument shall be clear. The performance of the instrument, which has a recorder, shall meet the relevant technical requirements.
4.1.3 Stainless steel pipes or PTFE pipes shall be used for gas pipelines. Each joint shall be tight and firm. There is no leak under the operating pressure.
4.2 Insulation resistance and dielectric strength
4.2.1 Insulation resistance
For the instrument using 220 V AC power supply, the insulation resistance of the phase line of the instrument power supply to ground shall not be less than 20 Mohm.
4.2.2 Dielectric strength
For instruments, which use 220 V AC power supply, the dielectric strength of the phase line of the power supply to the ground shall be able to withstand the test of AC voltage
1500 V, frequency 50 Hz, for a duration of 1 min, without breakdown and arcing.
5 Control of measuring instrument
The control of the instrument includes initial verification, subsequent verification, in- use inspection.
5.1 Verification conditions
5.1.1 Environmental conditions
Ambient temperature is (20 ± 10) °C, relative humidity is <= 85%, power supply voltage is (220 ± 22) V.
5.1.2 Standard materials and equipment for verification
5.1.2.l Standard material for sodium carbonate purity
National grade-2 standard material. Uncertainty of purity value: 0.02%, k = 2.
5.1.2.2 Standard material for potassium hydrogen phthalate purity
National grade-2 standard material. Uncertainty of purity value: 0.02%, k = 2.
5.1.2.3 Analytical balances
The minimum division value is 0.1 mg.
5.1.2.4 Glass measuring device
Grade A.
5.1.2.5 Blank water used for calibrating the standard solution of the instrument (see
Appendix A for the preparation method)
5.1.2.6 Insulation resistance gauge
500 V, grade 10.
5.1.2.7 Dielectric strength tester (voltage greater than 1.5 kV)
5.1.3 Other requirements
5.1.3.1 Instrument placement
The instrument shall be placed on the workbench stably. There is no strong light directly on the instrument, strong magnetic field, electric field interference, or vibration around the instrument. The indoor air is fresh and well ventilated.
5.1.3.2 Instrument carrier gas
The pressure and flow of the carrier gas source are stable at the specified values of the instrument.
5.2 Verification Items
The verification items are as shown in Table 1.
Appendix B
Preparation method of standard solutions
B.1 Equipment needed to prepare the solution
The balances, pipettes, volumetric flasks used shall be subject to quantitative verification.
B.2 Processing of chemical reagents
B.2.1 Processing of anhydrous sodium carbonate
Place the anhydrous sodium carbonate in a porcelain crucible. Dry it at 270 °C, for 4 h.
Then place it in a desiccator to cool, for later use.
B.2.2 Processing of anhydrous potassium hydrogen phthalate
Put potassium hydrogen phthalate in a weighing bottle. Then dry it in an oven at 118 °C, for 2 h. Take it out and put it in a desiccator to cool it, for later use.
B.3 Preparation of standard solutions
B.3.1 Inorganic carbon standard solution
Accurately weigh a certain amount of anhydrous sodium carbonate. Dissolve it in pure water. Transfer it to a volumetric flask. Dilute it to the mark. Shake well. Accurately calculate the concentration of the solution.
B.3.2 Organic carbon standard solution
B.3.2.1 Potassium hydrogen phthalate standard solution
Accurately weigh a certain amount of potassium hydrogen phthalate. Dissolve it in pure water. Transfer it to a volumetric flask. Dilute it to the mark. Shake well. Accurately calculate the concentration of the solution.
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.
How to Download JJG 821-2005
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJG 1151-2025 — Verification Regulation of Liquid Chromatograph-Atomic Fluorescence Spectrometers
JJG 1204-2025 — Verification Regulation of Electronic Price Computing Scales(for Trial Implementation)
JJG 579-2025 — Trial Case Lenses
Secure payment via Stripe
Payments accepted
JJG 821-2005
$175.00