JJG 714-2012Verification Regulation of Blood Cell Analyzers (English PDF)
This is a limited preview
Buy now to download the full PDF (10 pages)
Need quotation for your management?
Get a formal quotation for JJG 714-2012 in 30 minutes.
Issued by
General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China 3.
Level / Type
National · Mandatory
Issue date
June 18, 2012
Implementation date
December 18, 2012
Scope
JJG 714-2012 (Verification Regulation of Blood Cell Analyzers) is available as an English-translated PDF.
JJG 714-2012 is the Chinese standard "Verification Regulation of Blood Cell Analyzers". Its scope clause reads: This Regulation applies to the initial verification, subsequent verification and in-use verification of blood cell analyzers.
Its clauses include general. It was issued by the General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China 3. on 2012-06-18 and took effect on 2012-12-18.
Found what you need?
Document preview — JJG 714-2012
National Standard of the People's Republic of China
Issued by: General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China 3.
Contents
- 1 Scope...5
- 2 Normative References...5
- 3 General...5
- 4 Metrological Performance Requirements...6
- 5 General Technical Requirements...6
- 6 Control of Measuring Devices...6
- 7 Handlings of Verification Results...9
- 8 Verification Cycle...9
- Annex A Verification Original Data Log Sheet...10
1 Scope
This Regulation applies to the initial verification, subsequent verification and in-use verification of blood cell analyzers.
2 Normative References
The following document is referenced in this Regulation.
YY/T 0653-2008, Hematology Analyzer
For dated references, only the edition dated applies to this document. For undated references, the latest edition of the referenced documents (including all amendments) applies to this Regulation.
3 General
Blood cell analyzers (hereinafter referred to as analyzers) are capable of measuring the red blood cell count (RBC), white blood cell count (WBC), blood platelet count (PLT) and hemoglobin concentration (HGB) in human blood, which are mainly used for blood tests by hospitals and scientific research departments and other departments.
The measuring methods for the red blood cell count in the analyzers include electrical impedance method, laser scattering method, fluorescent staining method and so on.
The method widely used is the electrical impedance method; gem hole is used as the sensor; after the sensor absorbs a fixed quantity of blood cell specimen, the blood cell count is converted into the corresponding electrical pulse count. The electrical pulse is amplified, voltage discriminated and shaped; then the blood cell count is determined by measuring the electrical pulse count. The measurement of hemoglobin concentration is generally carried out in accordance with the colorimetric principle; the photoelectric element is used as the sensor; the sensor converts the variation of hemoglobin concentration into the variation of the corresponding voltage signal; the hemoglobin concentration is determined by measuring the variation of voltage after the voltage signal goes through amplification operation.
The apparatus consist of mechanical system, electrical system, testing system and control system and so on.
apparatus to preheat them; preset all switches and rotary knobs. The apparatus are verified in accordance with the routine procedures if their self-inspection is normal. For the apparatus having mode selection functions, set them to the whole blood measurement mode. Reagents, which are from the manufacturer or whose effectiveness can be proved, shall be used for the apparatus.
6.4.3 Blank value verification
Take an appropriate amount of apparatus blank diluent to pour into a clean measuring cup; repeat the measurement for 4 times. Discard the measured value for the first time;
record the maximum values of the other 3 measured results, i.e. the blank values of
RBC, WBC, PLT, HGB.
6.4.4 Carryover verification
Select one kind of standard substances of high and low values; mix up before use. In accordance with the sequence from high values to low values, carry out measurement continuously for 3 times. The measured values are respectively i1, i2, i3 and j1, j2, j3.
Calculate the carryovers of RBC, WBC, PLT and HGB in accordance with Formula (1).
where,
CO - the carryover;
i3 - the value of standard substance of a high value measured for the third time;
j1 - the value of standard substance of a low value measured for the first time;
j3 - the value of standard substance of a low value measured for the third time.
6.4.5 Verification of indication error
Take 3 ~ 5 kinds of blood cell standard substances of different concentrations within the commonly-used measuring range of the apparatus (take 5 kinds for the first verification); carry out measurement continuously for 5 times after mixing up respectively; calculate respectively the relative error deltaT of indicating value of the apparatus in accordance with Formula (2); take deltaT having the maximum absolute value as the verification results of the indication error of RBC, WBC, PLT and HGB of the apparatus.
where,
Ts - the standard value of standard substance;
Annex B
Format of Inner-Pages of Verification Certificate/Verification Result
Letter
B.1 Format of the second page of verification certificate/verification result letter
Certificate No. xxxx - xxxx
Verification body authorization instructions.
Environmental conditions and site for verification
Temperature. °C Site.
Relative humidity. % Others.
Metrological standard apparatus used for verification
Name Measuring range
Uncertainty/accuracy grade/maximum allowable error
Metrological standard certificate no.
Valid until
Standard instruments used for verification
Name Measuring range
Uncertainty/accuracy grade/maximum allowable error
Metrological standard certificate no.
Valid until
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 10 pages — is available in the English PDF.
Referenced standards
Normative references
How to Download JJG 714-2012
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0653-2008 — Hematology analyzer
JJG 1201-2024 — Digital Tyre Pressure Gauges
JJG 1202-2024 — Long Distance Transmissible Pressure Gauges
Secure payment via Stripe
Payments accepted
JJG 714-2012
$125.00