JJG 376-2007Verification Regulation of Electrolytic Conductivity Meters (English PDF)
Also coversJJG376-2007
This is a limited preview
Buy now to download the full PDF (14 pages)
Need quotation for your management?
Get a formal quotation for JJG 376-2007 in 30 minutes.
Issued by
State Administration of Quality Supervision, Inspection and Quarantine Verification Regulation of
Level / Type
National · Mandatory
Issue date
November 21, 2007
Implementation date
May 21, 2008
Scope
JJG 376-2007 (Verification Regulation of Electrolytic Conductivity Meters) is available as an English-translated PDF.
JJG 376-2007 is the Chinese standard "Verification Regulation of Electrolytic Conductivity Meters". Its scope clause reads: This regulation applies to the first-time verification, subsequent verification, and in-use inspection of electrolytic conductivity meters.
Calibration of resistivity meter, salinity meter based on the electrical conductivity measurement principle, and TDS measurement instrument can reference to this regulation.
Its clauses include terminology and measuring unit; overview; measuring performance requirements; general technical requirements; control of measuring instruments.
It was issued by the State Administration of Quality Supervision, Inspection and Quarantine Verification Regulation of on 2007-11-21 and took effect on 2008-05-21.
Found what you need?
Document preview — JJG 376-2007
National Standard of the People's Republic of China
Issued by: State Administration of Quality Supervision, Inspection and Quarantine Verification Regulation of
Contents
- 1 Scope...4
- 2 Normative references...4
- 3 Terminology and measuring unit...4
- 4 Overview...6
- 5 Measuring performance requirements...6
- 5.1 The electronic unit repeatability...6
- 5.2 Electronic unit cited error...6
- 5.3 Indication error of conductance cell's constant...6
- 5.4 Indication error of temperature coefficient...6
- 5.5 Indication error of temperature measurement...6
- 5.6 Instrument cited error...6
- 5.7 Repetitiveness of Instrument...7
- 6 General technical requirements...7
- 7 Control of measuring instruments...7
- 7.1 Verification conditions...7
- 7.2 Verification items...9
- 7.3 Verification method...9
- 7.4 Processing of verification results...14
- 7.5 Verification period...14
1 Scope
This regulation applies to the first-time verification, subsequent verification, and in-use inspection of electrolytic conductivity meters. Calibration of resistivity meter, salinity meter based on the electrical conductivity measurement principle, and TDS measurement instrument can reference to this regulation.
2 Normative references
OIML R68 Edition 1985 Calibration method for conductivity cell
BS EN 60746-3.2002 Expression of performance of electrochemical analyzers - Part 3.
3 Terminology and measuring unit
3.1 Electrolytic conductance
The ratio of current and electric potential difference when ionic charge of conductivity cell moves in the electrolyte solution.
3.3 Cell constant
Conductance cell constant is calculated by the formula (3).
3.4 Temperature coefficient
The relative change of the conductivity of the electrolyte solution that is caused by temperature change of 1 °C. For strong electrolyte whose electrical conductivity is greater than 1 x 10-4 S·m-1, the temperature coefficient alpha can be approximately expressed by
Formula (5).
4 Overview
Conductivity meter is used to measure the electrolyte solution conductivity. The measuring principle of conductivity meter is based on the relationship [formula (4)] of conductivity, electric conductance, and conductance cell’s constant. Apply alternating electrical signals between electrodes of conductance cell to measure the solution conductivity; obtain the conductivity according to the imputed conductance cell’s constant.
5 Measuring performance requirements
5.1 The electronic unit repeatability
Verification results shall comply with the provisions of Table 1.
5.2 Electronic unit cited error
Verification results shall comply with the provisions of Table 1.
5.3 Indication error of conductance cell’s constant
Verification results shall comply with the provisions of Table 1.
6 General technical requirements
Instrument appearance shall be bright, clean and flat. Instrument function keys shall be able to work properly. All fasteners shall not be loose. The indication of instrument panel shall be clear and intact. Digital display instrument shall be clear and complete.
7 Control of measuring instruments
Control of measuring instruments includes first-time verification, subsequent verification and in-use inspection.
7.1 Verification conditions
7.1.1 Environmental conditions
Verification results shall comply with the provisions of Table 2.
7.1.2 Solution conductance simulator
It is the standard device for verifying electronic unit’s metrological performance. It is consisted of a set of non-inductive a.c. (alternate current) resistance. Setting of resistance value shall be able to meet the requirements of "verification method".
7.1.3 Conductivity standard solution
The relative uncertainty of reference value of standard solution used in verification shall be less than or equal to 0.25 0.25% (k = 2). Potassium chloride solution conductivity standard material may be used. It may also use KCl conductivity solid standard material prepared according to the provisions of the appendix.
7.3.8 Instrument repeatability
Repeatedly measure standard solution 1 or standard solution 2 for 6 times. Calculate the ratio of single-measurement standard deviation AND full-range according to Formula (6).
It represents the repeatability of instrument measurement results.
7.4 Processing of verification results
After verification, the instrument complying with this regulation is deemed as qualified instrument. A verification certificate shall be issued. The certificate shall provide each item’s verification result and the instrument level. The instrument at certain level shall meet the requirements of all metrological performance of that level. If some range can not meet the requirements, it shall indicate the limitation range. If the instrument is not qualified, a notice of verification result shall be issued.
7.5 Verification period
The verification period generally shall not be more than a year.
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 14 pages — is available in the English PDF.
How to Download JJG 376-2007
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJG 1192-2023 — Calibrator of Off-board Chargers for Electric Vehicles
JJG 1193-2023 — Calibrator of AC Charging Piles for Electric Vehicles
JJG 908-2023 — Automobile Side Slip Testers
Secure payment via Stripe
Payments accepted
JJG 376-2007
$125.00