Valid

HS/T 62-2019Determination of copper content in printed circuit board waste and scrap -- Wavelength dispersive X-Ray fluorescence spectrometry (English PDF)

Preview PDF

This is a limited preview

Buy now to download the full PDF (5 pages)

Need quotation for your management?

Get a formal quotation for HS/T 62-2019 in 30 minutes.

Issued by

General Administration of Customs

Level / Type

National · Recommended

Issue date

December 19, 2019

Implementation date

June 1, 2020

Scope

HS/T 62-2019 (Determination of copper content in printed circuit board waste and scrap -- Wavelength dispersive X-Ray fluorescence spectrometry) is available as an English-translated PDF.

HS/T 62-2019 — This standard specifies test methods for quantitative analysis of copper content in PCB waste and scrap. This standard is applicable to the determination of copper content in PCB waste and scrap. The measurement range is 10.0 g/100g ~ 56.0 g/100g.

Found what you need?

Document preview — HS/T 62-2019

National Standard of the People's Republic of China

ICS
71.040.50
Classification
G 85/89

Issued by: General Administration of Customs

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 Principle
  • 4 Reagents and materials
  • 4.6 Platinum-gold alloy crucible (95% P 5% Au)
  • 5 Equipment
  • 6 Analysis steps
  • 6.1 Sample preparation
  • 6.2 Determination
  • 7 Result representation
  • 8 Precision
  • 28 Vac 30 100 55.1 0.46 F-PC 50 ~ 150 20.756 9.893

Foreword

This standard was drafted in accordance with the rules given in HS/T 1-2011 and HS/T 39-2013.

This standard was proposed by the Customs Administration Department of the General Administration of Customs of the People's Republic of China.

This standard is under the jurisdiction of the Department of Policies and Regulations of the General Administration of Customs of the People's Republic of China.

Appendix A and Appendix B of this standard are informative appendices.

This standard was drafted. Chongqing Customs, People's Republic of China.

1 Scope

This standard specifies test methods for quantitative analysis of copper content in PCB waste and scrap.

This standard is applicable to the determination of copper content in PCB waste and scrap. The measurement range is 10.0 g/100g ~ 56.0 g/100g.

2 Normative references

The following documents are essential for the application of this document. For dated references, only the dated version applies to this article Pieces. For undated references, the latest version (including all amendments) applies to this document.

GB/T 6682 National Standard for Analytical Laboratory Water JY/T 016 General rules for wavelength-dispersive X-ray fluorescence spectroscopy

3 Principle

The crusher and ring grinder are used to crush and pre-process the PCB. After homogenization, the muffle furnace is used for burning. After burning, the ball is ground.

The instrument performs secondary pulverization, and then prepares as a fuse. The measured value is obtained by X-fluorescence quantitative analysis, and then the weighing value of the sample before and after burning, the elemental ratio of copper in copper oxide and X-fluorescence quantitative measurement calculate the copper content in the PCB.

4 Reagents and materials

Unless otherwise specified, the reagents used in this method are analytical grade, and the water is grade III water specified in GB/T 6682.

4.1 Lithium tetraborate (LiB4O7).

4.2 Lithium metaborate (LiBO2).

4.3 LiBr. Prepared as a solution with a mass fraction of 18 g/100g.

4.4 Lithium nitrate (LiNO3). formulated as a solution with a mass fraction of 22 g/100g.

4.5 Corundum crucible. 100 mL volume.

4.6 Platinum-gold alloy crucible (95% P 5% Au)

4.7 Desiccator. An effective desiccant is included.

4.8 Reference materials. copper oxide (CuO), silicon dioxide (SiO2), aluminum oxide (Al2O3), iron oxide (Fe2O3), oxygen Calcium (CaO), magnesium oxide (MgO).

5 Equipment

5.1 Wavelength dispersion X-ray fluorescence spectrum. The composition and performance of the instrument should meet the requirements of JY/T 016.

5.2 Pulverizer. Sieve with 5 cm hole size.

5.3 Ring grinder. with tungsten carbide grinding head.

5.4 Ball mill. The maximum speed is not less than 500 rpm.

5.5 Electronic balance. Sensitivity 0.1 mg.

5.6 Muffle furnace. the maximum operating temperature is not less than 850 °C, and the cavity volume is not less than 30 L.

5.7 Automatic temperature control oven. The maximum use temperature is not lower than 110 °C.

5.8 Melt sample machine. The maximum operating temperature is not less than 1150 °C, with a crucible.

5.9 Standard sieve. 100 mesh.

HS/T XXXX-XXXX

5.10 Microsampler.

6 Analysis steps

6.1 Sample preparation

6.1.1 Crush

A 1 kg sample was weighed and pulverized into an irregular block and powder with a maximum diameter of 5 cm by a pulverizer (5.2), and recorded as sample a.

6.1.2 Ring grinding

The sample a was put into a ring mill, and the obtained powder was passed through a 100-target sieve, and recorded as sample b.

6.1.3 Burn

The following steps are started, and two samples are tested in parallel.

6.1.3.1 Take sample b in an oven at 101 °C ~ 105 °C, dry it for 2 hours, take it out, cool to room temperature in a desiccator, and weigh about 3 g.

Record the weight as m1 (to the nearest 0.1 mg).

6.1.3.2 After weighing, put it into the corundum crucible (4.5) and place it in a muffle furnace for burning. The burning temperature is 775 °C ~ 825 °C. The burning time for 10 h. After taking it out, cool it to room temperature in a desiccator and weigh it. Record the weight as m2 (accurate to 0.1 mg).

6.1.4 Ball milling

The sample c was ball-milled. The rotation speed was set to 500 r/min and the ball-milling time was 2 minutes. The sample obtained by the ball-milling was recorded as sample d.

6.1.5 Fuse

6.1.5.1 Put sample d in an oven at 101 °C ~ 105 °C, dry it for 2 hours, and take it out. Cool it in a desiccator to room temperature before weighing.

6.1.5.2 Prepare lithium tetraborate and lithium metaborate as a mixture of lithium borate in a ratio of 67.33; weigh 0.4 g of dry sample d (standard to the nearest 0.1 mg), weigh 6 g of lithium borate mixed flux (accurate to 0.1 mg), mix uniformly, and drop in lithium bromide with a micro sampler.

The solution and lithium nitrate solution are each 1 mL. All weighing and mixing processes should be completed in a platinum-gold alloy crucible (4.6).

6.1.5.3 Put the crucible (4.6) with the sample into the melting sample machine. The recommended melting temperature is 1050 °C and the melting time is 10 minutes.

After the melting is completed, the sample in the crucible is poured into the mold, taken out after cooling, and directly formed or poured into a measurable sample piece.

6.2 Determination

6.2.1 Draw a standard curve

Refer to Table A.1 in Appendix A for the proportions of the five groups. Weigh the reference material (4.8) (accurate to 0.1 mg), mix well, and follow 6.1.5 the method was used to prepare a standard fuse, and a standard curve was measured and drawn.

6.2.2 Sample determination

Select an appropriate standard fuse (6.2.1) for instrument drift correction. The calibration interval can be determined according to the stability of the instrument. Test the tablet is measured, and the result is calculated as copper oxide and recorded as W (four significant digits are retained). The test conditions of the instrument are shown in Table B.1 in Appendix B.

Or for those who use more advanced instruments, they should choose according to the optimization method of the instrument manufacturer.

7 Result representation

7.1 The copper content in the PCB is calculated according to formula (1).

XWK= x x x (1)

In the formula.

W-copper content (calculated as copper oxide) using an X-ray fluorescence spectrometer;

K-the coefficient of copper oxide converted to copper, calculated as 0.799 (calculated according to the latest international atomic weight table);

m1-the amount of sample weighed before burning, in grams (g);

m2-the amount of sample weighed after burning, the unit is gram (g);

X-The copper content (in terms of copper) in the PCB, the unit is grams per hundred grams (g/100g).

7.2 The result is retained to one digit after the decimal point. The arithmetic mean of the results of the two determinations is taken as the final result.

8 Precision

The analysis results are as follows. in the same laboratory, the same operator uses the same equipment, according to the same test method, and in a short time the absolute difference between independent test results obtained from the same measured object must not exceed 0.8 g/100g; in different laboratories, different operators the author uses different equipment and according to the same test method, the absolute difference between the test results obtained from the same test object must not exceed 1.5 g/100g.

HS/T XXXX-XXXX

Appendix A

(Informative appendix) Standard Fuse Raw Material Proportion Table

Table A.1 Proportion of oxide raw materials used to configure standard fuses

Unit. g

Group

Oxide type

CuO SiO2 Al2O3 Fe2O3 CaO MgO 10% CuO 0.0400 0.2772 0.0478 0.0167 0.0116 0.0066 30% CuO 0.1200 0.1624 0.0736 0.0296 0.0052 0.0093 40% CuO 0.1600 0.1240 0.0055 0.0152 0.0050 0.0903 50% CuO 0.2000 0.0996 0.0053 0.0130 0.0113 0.0708 70% CuO 0.2800 0.0642 0.0162 0.0174 0.0110 0.0113 Appendix B (Informative appendix) Measurement condition table

Table B.1 Table of measurement conditions

Elemental analysis line Mask mm mode Voltage kV Current mA Spectroscopic crystal 2d (Å) Collimator detector PHA Peak position 2 theta (degrees) wavelength (Å) Cu KB1-Maj 28 Vac 50 24 4.026 0.23 SC 40 ~ 250 40.461 1.392 4 Mg KA1-HR-M in

28 Vac 30 100 55.1 0.46 F-PC 50 ~ 150 20.756 9.893

Al KA1-Maj 28 Vac 30 100 8.752 0.23 F-PC 50 ~ 150 144.639 8.339 3 Si KA1-Maj 28 Vac 30 100 8.752 0.46 F-PC 50 ~ 150 109.022 7.125 4 Ca KA1-HS -M in

28 Vac 50 60 4.026 0.46 F-PC 50 ~ 150 113.141 3.358 4 Fe KA1-Maj 28 Vac 50 10 4.026 0.23 SC 50 ~ 150 57.517 1.936 Ba LA1-HR-T

28 Vac 50 60 4.026 0.23 F-PC 50 ~ 150 87.177 2.777 7 Note. The measurement conditions in Table B.1 are Bruker S8 Tiger type instruments. The requirements of the instrument for the test environment are a temperature of 17 °C ~ 29 °C and a humidity of <= 60%;

Of users recommend using their own optimal measurement conditions.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 5 pages — is available in the English PDF.

Referenced standards

How to Download HS/T 62-2019

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
5 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

HS/T 62-2019

$150.00