HJ 770-2015Water quality. Determination of Phenoxy carboxylic acids herbicide by High Performance Liquid Chromatography. Tandem Mass Spectrometry (English PDF)

水质 苯氧羧酸类除草剂的测定 液相色谱串联质谱法

Preview PDF

This is a limited preview

Buy now to download the full PDF (11 pages)

Need quotation for your management?

Get a formal quotation for HJ 770-2015 in 30 minutes.

Issued by

Ministry of Ecology and Environment

Level / Type

National · Mandatory

Issue date

November 20, 2015

Implementation date

December 15, 2015

Scope

HJ 770-2015 is the English-translated version of 水质 苯氧羧酸类除草剂的测定 液相色谱串联质谱法.

HJ 770-2015 — This standard specifies the determination of phenoxy carboxylic acid herbicides in water by liquid chromatography/tandem mass spectrometry.

This standard applies to surface water, groundwater and waste water 2-methyl-4-chlorophenoxyacetic acid, 2,4-dichlorophenoxyacetic acid, 2- (2-methyl-Chlorophenoxy) propionic acid, 2- (2,4-dichlorophenoxy) -propionic acid (or 2,4-propionic acid), 2,4,5-trichlorophenoxyacetic acid, 2- ( 2,4,5-trichloro Phenoxy) -propionic acid (or 2,4,5-propionic acid), 4- (2,4-dichlorophenoxy) -butyric acid and 4- (2-methyl-4-chlorophenoxy ) Butyric acid Determination of phenoxy carboxylic acid herbicides.

When the injection volume is 10 µl, the detection limit of the method is 0.3 µg/L ~ 0.5 µg/L, and the detection limit is 1.2 µg/L ~ 2.0 Mg/L, see Appendix A. When the volume of the sample was 100 ml, the volume of the volume was 1.0 ml after concentration and the volume of the sample was 10 µl.

Limited to 0.006 µg/L to 0.009 µg/L, with a lower limit of 0.024 µg/L to 0.036 µg/L, as described in Appendix A.

Found what you need?

Document preview — HJ 770-2015

National Standard of the People's Republic of China

Issued by: Ministry of Ecology and Environment

Contents

  • Preface
  • 1 Scope of application
  • 2 Normative reference documents
  • 3 Principle of the method
  • 4 Reagents and materials
  • 5 Instruments and equipment
  • 6 Samples
  • 7 Analysis steps
  • 8 Results calculation and presentation
  • 9 Precision and accuracy

Foreword

To implement the "Environmental Protection Law of the People's Republic of China" and "Water Pollution Prevention Law of the People's Republic of China" to protect the environment and protect the human body Health, standard water phenoxy carboxylic acid herbicides monitoring methods, the development of this standard.

This standard specifies the determination of phenoxy carboxylic acid herbicides in water by liquid chromatography/tandem mass spectrometry.

This standard is the first release.

Appendix A is a normative appendix, Appendix B is an informative appendix.

This standard is organized by the Ministry of Environmental Protection Science and Technology Standards Division.

The main drafting unit of this standard. Jiangsu Province Environmental Monitoring Center.

The standard method of verification unit. Suzhou City Environmental Monitoring Center Station, Dalian Environmental Monitoring Center Station, Jiangsu Entry-Exit Inspection and Quarantine Bureau Food Laboratory, Changzhou Environmental Monitoring Center, Jinan Environmental Monitoring Center, Shimadzu Enterprise Management (China) Co., Ltd. Shanghai Analysis center.

The Environmental Protection Department of this standard approves on November 20, 2015.

This standard is implemented as of December 15, 2015.

This standard is explained by the Ministry of Environmental Protection.

Water quality - Determination of phenoxy carboxylic acid herbicides Liquid chromatography/tandem mass spectrometry Warning. Phenoxy carboxylic acid herbicides are toxic organic compounds, the experimental process of the use of organic solvents with strong volatility. Minute Analysis of personnel in the experimental operation, should be in the fume hood operation; should be required to wear protective equipment, to avoid contact with skin and clothing.

1 Scope of application

This standard specifies the determination of phenoxy carboxylic acid herbicides in water by liquid chromatography/tandem mass spectrometry.

This standard applies to surface water, groundwater and waste water 2-methyl-4-chlorophenoxyacetic acid, 2,4-dichlorophenoxyacetic acid, 2- (2-methyl-Chlorophenoxy) propionic acid, 2- (2,4-dichlorophenoxy) -propionic acid (or 2,4-propionic acid), 2,4,5-trichlorophenoxyacetic acid, 2- ( 2,4,5-trichloro Phenoxy) -propionic acid (or 2,4,5-propionic acid), 4- (2,4-dichlorophenoxy) -butyric acid and 4- (2-methyl-4-chlorophenoxy ) Butyric acid Determination of phenoxy carboxylic acid herbicides.

When the injection volume is 10 µl, the detection limit of the method is 0.3 µg/L ~ 0.5 µg/L, and the detection limit is 1.2 µg/L ~ 2.0 Mg/L, see Appendix A.

When the volume of the sample was 100 ml, the volume of the volume was 1.0 ml after concentration and the volume of the sample was 10 µl.

Limited to 0.006 µg/L to 0.009 µg/L, with a lower limit of 0.024 µg/L to 0.036 µg/L, as described in Appendix A.

2 Normative reference documents

The contents of this standard refer to the terms of the following documents. For undated references, the valid version applies to this standard.

Technical specification for surface water and wastewater monitoring Technical specification for groundwater environmental monitoring

3 Principle of the method

The phenoxy carboxylic acid herbicides in water were enriched by direct injection or solid phase extraction and separated by liquid chromatography/tandem mass spectrometry. according to Retention time and characteristic ion peak characterization, internal standard method quantification.

4 Reagents and materials

Unless otherwise stated, analytical reagents and experimental water are used in accordance with national standards.

4.1 Standard solution

4.1.1 Phenoxy carboxylic acid herbicide standard stock solution. rho = 100 µg/ml.

Can be directly purchased certified standard solution can also be prepared with standard materials. The stock solution is stored in the dark at 0 ° C to 4 ° C or referenced to the manufacturer product manual. Use should be restored to room temperature, and shake.

4.1.2 phenoxy carboxylic acid herbicides standard use of liquid. rho = 10.0 µg/ml (reference concentration).

The standard stock solution (4.1.1) of the phenoxy carboxylic acid compound is diluted with acetonitrile (4.2) as required. Standard use of liquid at 4 °C cold, avoid Light preservation, shelf life of two months. Use should be restored to room temperature, and shake.

4.1.3 Internal standard stock solution. rho = 100 µg/ml.

The internal standard is 2,4-dichlorophenoxyacetic acid-13C6, can be directly purchased certified standard solution can also be prepared with the standard material. The stock solution is at zero °C ~ 4 °C Keep away from light or refer to the manufacturer's product description. Use should be restored to room temperature, and shake.

4.1.4 Internal standard liquid. rho = 10.0 µg/ml (reference concentration).

The internal standard stock solution (4.1.3) was diluted with acetonitrile (4.2) as needed. Internal standard use of liquid at 0 °C ~ 4 °C dark preservation, shelf life Two months. Use should be restored to room temperature, and shake.

4.2 acetonitrile (CH3CN). pesticide residues.

4.3 methanol (CH3OH). pesticide residues.

4.4 Methanol/water mixed solution. 1 4.

4.5 acetonitrile/water mixed solution. 1 4.

4.6 Ammonium acetate solution. C (CH3COONH4) = 0.002 mol/L.

Weigh 0.077 g of ammonium acetate in 500 ml of experimental water and 0.22 µm filter (4.13).

4.7 Sulfuric acid (H2SO4). rho (H2SO4) = 1.84 g/ml.

4.8 Sodium hydroxide (NaOH).

4.9 Sulfuric acid solution. 1 1.

Measure 50 ml of concentrated sulfuric acid (4.7) and slowly add to 50 ml of water.

4.10 Sodium hydroxide solution. rho (NaOH) = 0.4 g/ml.

Weigh 40 g of sodium hydroxide (4.8) in water and set to 100 ml.

4.11 solid phase extraction column. filler for the divinyl benzene and N-vinyl pyrrolidone copolymer (HLB) or the same column efficiency of the extraction column, specifications for the 6 Ml/500 mg.

4.12 Nitrogen. Purity >=99.99%.

4.13 Filtration. 0.22 µm Teflon or other material equivalent filter.

5 Instruments and equipment

5.1 solid phase extraction device. automatic or manual, flow rate can be adjusted.

5.2 Concentration device. KD concentrator or nitrogen blowing concentrator and other equipment equivalent performance.

5.3 Liquid Chromatography/Tandem Mass Spectrometer. Equipped with electrospray ionization source (ESI).

5.4 Column. C18 or equivalent reversed-phase high performance liquid chromatography column with reference specifications of 100 mm × 2.1 mm, 1.7 µm.

5.5 Microinjectors. 10 µl, 50 µl, 100 µl, 250 µl.

5.6 Common laboratory equipment and equipment.

6 Samples

6.1 Sample collection

Samples were collected with reference to the relevant provisions of HJ/T 91 and HJ/T 164.

Fresh water was collected with a pre-washed clean and dry milled brown glass bottle (250 ml).

6.2 Sample storage

Collected water samples 0 °C ~ 4 °C dark preservation, 3 days analysis is completed.

6.3 Preparation of the sample

6.3.1 Direct injection method

Mix 1.0 ml of water, add the internal standard solution 5.0 µl (4.1.4), mix and filter through the 0.22 µm filter (4.13), placed in the sample Bottle, to be tested.

6.3.2 Solid phase extraction method

The solid phase extraction column (4.11) was activated with 10 ml of methanol (4.3) and 10 ml of experimental water to ensure the stigma infiltration. Measure 100 Ml water sample, using a sulfuric acid solution (4.9) or sodium hydroxide solution (4.10) to adjust the pH to neutral water, water samples below 10 ml/min (About 3 to 4 drops/second) flow rate through the column. The volume of the waste water sample can be appropriately reduced according to the actual situation. Mix the solution with 10 ml of methanol/water (4.4) Leave the column to remove the weaker impurities on the column. Followed by purging with nitrogen (4.12) and drying the column. And then 10 ml of a Alcohol (4.3) The column was eluted at a flow rate of about 3 ml/min (about 1 drop/sec), and the eluate was received in the collection tube. The eluate was blown through nitrogen (4.12, Pay attention to keep the liquid slightly fluctuate) to dry, dry with acetonitrile/water solution (4.5) to 1.0 ml, and finally add the internal standard liquid 5.0 µl (4.1.4), after mixing 0.22 µm filter (4.13), placed in the sample bottle, to be tested.

6.4 Preparation of blank sample

6.4.1 Lab blank

The samples were prepared in the same manner as in the water treatment (6.3.1) to prepare a blank sample with direct injection.

The solid phase extraction method was prepared according to the same procedure of water sample treatment (6.3.2) instead of the sample.

6.4.2 Full program blank

Before sampling, follow the sample collection (6.1) and sample preservation (6.2) method, with the experimental water preparation of the whole program blank sample, and with it the sampling process is brought to the sampling site. The whole program blanks the sample with the actual sample and reaches the laboratory at the same time.

7 Analysis steps

7.1 Instrument reference conditions

7.1.1 Liquid Chromatographic Reference Conditions

Mobile phase. 0.002 mol/L ammonium acetate aqueous solution (A phase), acetonitrile (B phase), gradient elution program in Table 1.

Flow rate. 0.3 ml/min.

Column temperature. 40 ° C.

Injection volume. 10 µl.

Table 1 Liquid Chromatography Mobile Phase Gradient Elution Procedure

Time (min) A% B%

6.5 80 20

7.1.2 Mass Spectral Reference Conditions

Negative ion mode

Capillary voltage. 2.8 kV Ion source temperature. 120 ° C Atomization temperature. 350 ° C Atomized gas flow rate. 800 L/h Reverse blow flow rate. 10 L/h Collision air flow rate. 0.10 ml/min Multi-ion reaction monitoring mode (MRM), the specific conditions in Table 2.

Table 2 Monitoring conditions for the multi-ion reaction of the target compound

The compound parent ion (m/z) Ion (M/z) Dwell time/s tap hole voltage/V collision voltage/V 199 * 141 20 15 2-methyl-4-chlorophenoxyacetic acid 0.02 219 * 161 16 12 2,4-dichlorophenoxyacetic acid 0.02 213 * 141 20 15 2- (2-methyl-4-chlorophenoxy Yl) propionic acid 215 143 0.02 233 * 161 16 12 2- (2,4-dichlorophenoxy) - propionic acid 235 163 0.02 253 * 195 16 12 2,4,5-trichlorophenoxyacetic acid 0.02 267 * 195 16 10 2- (2,4,5-trichlorophenoxy) Yl) -propionic acid 269 197 0.02 247 * 161 10 10 4- (2,4-Dichlorophenoxy) -Butyric acid 249 163 0.02 227 * 141 15 12 4- (2-methyl-4-chlorophenoxy Yl) butyric acid 229 143 0.02 225 * 167 15 12 2,4-dichlorophenoxyacetic acid-13C6 227 169 0.02 Note. with the * for the quantitative ion pair, for different mass spectrometry equipment, the parameters may be different, should be measured before the mass spectrometry parameters to optimize the best.

7.1.3 Instrument tuning

According to the instrument manual in the specified time and frequency of liquid chromatography/tandem mass spectrometer for the quality of the instrument and the sensitivity correction, to ensure that the instrument is in the best test state.

In the course of the use of the instrument, if the instrument found that the quality of significant deviation or a significant decline in sensitivity, the instrument should be immediately re-Perform quality and sensitivity correction.

7.2 Calibration

7.2.1 Drawing of calibration curves

(4.1.2) in a mixed solution of acetonitrile/water (4.5), at least 5 concentration points were prepared.

The standard concentrations of phenoxy carboxylic acids were 1.0 µg/L, 2.0 µg/L, 5.0 µg/L, 10.0 µg/L, 50.0 µg/L and 100 µg/L, respectively (This is the reference concentration), in each standard solution of the standard solution by adding 5.0 µl standard internal standard liquid (4.1.4), so that the internal standard mass concentration for 50.0 µg/L, stored in a brown sample vial, to be tested.

From the low concentration to the high concentration, the standard series solution was sequentially injected with the concentration of the target component in the standard series solution as the abscissa the calibration curve is established by plotting the ratio of the peak area (or peak height) to the peak area of the internal standard (or peak height) and the product of the internal standard concentration.

The calibration curve has a correlation coefficient >=0.995, otherwise the calibration curve is redrawn.

Y = ax b (1)

Where. x - target component concentration, µg/L Y - the product of the peak area (or peak height) ratio of the target component and the internal standard substance and the concentration of the internal standard substance A - calibration curve slope B - calibration curve intercept

7.2.2 Liquid Chromatography/Mass Spectrometry

Fig.1 Total ion chromatogram (concentration. 50 µg/L) of liquid chromatography/mass spectrometry of 8 phenoxy carboxylic acid herbicides and internal standard

Note. 1. 2-Methyl-4-chlorophenoxyacetic acid; 2,3. 2,4-dichlorophenoxyacetic acid, 2,4-dichlorophenoxyacetic acid-13C6; 4. 2- (2- 4-chlorophenoxy) propionic acid;

5. 2- (2,4-dichlorophenoxy) -propionic acid; 6. 2,4,5-trichlorophenoxyacetic acid; 7. 2- (2,4,5-trichlorophenoxy) - propionic acid; 8. 4- (2,4-dichlorophenoxy) -but Acid; 9. 4- (2-methyl-4-chlorophenoxy) butyric acid

7.3 Determination

7.3.1 Specimen determination

The same instrumental analysis conditions were used to determine the calibration curve.

7.3.2 Blank test

The measurement of the blank sample (6.4) was carried out according to the same instrument analysis condition (7.1) as the test sample.

Calculation and representation of results

8.1 Qualitative analysis of the target compound

Each of the tested components was selected for 1 parent ion and 2 daughter ions for monitoring. Under the same experimental conditions, the components of the sample to be tested the absolute value of the relative standard deviation should be less than 2.5% compared to the retention time of the target component in the standard sample; and the sample spectrum in the figure, the relative abundance (Ksam) of the qualitative ions of each component is the relative abundance of the qualitative ions in the standard solution spectrum close to the concentration (Kstd), the deviation does not exceed the range specified in Table 3, and it can be determined that there is a corresponding analyte in the sample.

Where. Ksam - the relative abundance of qualitative ions in a sample in a sample,% A2 - the peak area (or peak height) of the qualitative ion pair of a group in the sample A1 - the peak area (or peak height) of the quantitative ion pair of a component in the sample (3) where. Kstd - the relative abundance of qualitative ions in a standard sample,% Astd2 - the peak area (or peak height) of the qualitative ion pair of a group in the standard sample Astd1 - the peak area (or peak height) of the quantitative ion pair of a component in the standard sample

Table 3 Qualitative confirmation of the maximum allowable deviation of the relative abundance of ions

unit.%

A qualitative ion of a component in a standard sample Relative ion abundance (Kstd) Kstd> 50 20 9 precision and accuracy

9.1 Precision

9.1.1 Direct injection method

Six labs were used for the concentration of phenoxy carboxylic acid herbicides at concentrations of 1.0 µg/L, 10.0 µg/L and 50.0 µg/L, respectively.

6 repeated measurements.

The relative standard deviations in the laboratory were 2.4% ~ 13.9%, 1.1% ~ 7.3%, 2.0% ~ 6.6% respectively.

The relative standard deviations were 1.5% ~ 6.8%, 0.6% ~ 4.8%, 0.7 ~ 4.2%, respectively.

The reproducibility limits were. 0.2µg/L ~ 0.3µg/L, 0.9µg/L ~ 1.4µg/L, 5.1µg/L ~ 6.3µg/L.

The reproducibility limits were. 0.2µg/L ~ 0.3µg/L, 1.1µg/L ~ 1.5µg/L, 5.3µg/L ~ 7.4µg/L.

9.1.2 Solid phase extraction

6 laboratories with a concentration of 0.050µg/L of phenoxy carboxylic acid herbicides, 0.500µg/L and 1.00µg/L of the uniform blank sample Line six times to repeat the determination.

The relative standard deviations in the laboratory were 1.4% ~ 8.9%, 0.6% ~ 3.3% and 0.4% ~ 1.8% respectively.

The relative standard deviations were 1.2% ~ 2.1%, 0.2% ~ 0.8%, 0.2% ~ 0.6%, respectively.

The limit of reproducibility was 0.005µg/L ~ 0.007µg/L, 0.018µg/L ~ 0.031µg/L, 0.019µg/L ~ 0.032µg/L.

The reproducibility limits were. 0.005µg/L ~ 0.007µg/L, 0.019µg/L ~ 0.033µg/L, 0.02µg/L ~ 0.033µg/L.

9.2 Accuracy

9.2.1 Direct injection method

Six laboratories were tested for surface water and waste water, and the spiked concentrations were 10.0 µg/L and 50.0 µg/L, respectively. Spiked recovery rate the results were. 88.0% ~ 102%, 87.5% ~ 102%, the final recoveries were 90.8% ± 5.2% ~ 99.8% ± 3.6% 93.3% ± 4.2% ~ 97.5% ± 8.0%.

9.2.2 Solid phase extraction

Six laboratories carried out spike recovery experiments on surface water. The spiked concentrations were 0.100 µg/L and 1.00 µg/L, respectively.

The results showed that the titer was 2.00µg/L and 20.0µg/L, respectively. The recoveries of spiked were 74.2% 89.2%, 75.2% ~ 89.9%, and the final recoveries were 75.9% ± 1.4% ~ 87.5% ± 3.0% and 76.4 ± 1.4% ~ 89.3% ± 1.0%.

The accuracy and accuracy of the results are given in Appendix B.

10 Quality assurance and quality control

10.1 Blank analysis

Each analysis is done at least one lab blank and full program blank. To check for possible interference.

The concentration of each target compound in the laboratory blank and the program blank should not exceed the detection limit of the method, otherwise it should be resampled.

10.2 Determination of parallel samples

A parallel sample is to be analyzed for every 20 samples or for each batch (less than 20 samples/batch). The relative standard deviation of the parallel samples should be controlled 20% or less.

10.3 Matrix spikes

Approximately 20 samples or per batch (less than 20 samples/batch) need to be a matrix plus standard sample. The range of recoveries should be determined by direct injection Between 85% and 110%, the recoveries of the solid phase extraction were between 70% and 120%.

10.4 Calibration

10.4.1 Initial calibration

The initial calibration is required for the initial use of the instrument, or when the instrument is repaired, the column is replaced, or the continuous calibration is not acceptable. That is to establish the standard Quasi-curve.

10.4.2 Continuous calibration

Continuous calibration was performed using the intermediate concentration point of the analytical calibration curve, 1 for every 20 samples or per batch (less tha...

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 11 pages — is available in the English PDF.

How to Download HJ 770-2015

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
11 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

HJ 770-2015

$300.00