HJ 683-2014Ambient air - Determination of aldehyde and ketone compounds - High performance liquid chromatography (English PDF)

Preview PDF

This is a limited preview

Buy now to download the full PDF (11 pages)

Need quotation for your management?

Get a formal quotation for HJ 683-2014 in 30 minutes.

Issued by

Ministry of Environmental Protection

Level / Type

National · Mandatory

Issue date

January 13, 2014

Implementation date

April 1, 2014

Scope

HJ 683-2014 (Ambient air - Determination of aldehyde and ketone compounds - High performance liquid chromatography) is available as an English-translated PDF.

HJ 683-2014 is the Chinese standard "Ambient air - Determination of aldehyde and ketone compounds - High performance liquid chromatography". Its scope clause reads: This Standard specifies high performance liquid chromatography for the determination of aldehyde and ketone compounds in ambient air.

This Standard applies to the determination of thirteen aldehyde and ketone compounds in ambient air, including formaldehyde, acetaldehyde, acrolein, acetone, propionaldehyde, crotonaldehyde, methacrolein, 2-butanone, butyraldehyde, benzaldehyde, valeraldehyde, m-tolualdehyde and n-hexaldehyde.

Other aldehyde and ketone compounds can also be used in the analysis of this Standard provided that they pass the suitability validation. When the sampling volume is 0.05 m3, the detection limit of this Method is 0.28 µg/m3 ~ 1.69 µg/m3 and the low-limit of determination is 1.12 µg/m3 ~ 6.76 µg/m3.

For details see Annex A. Its clauses include application scope; method principle; analytical procedures; quality assurance and quality control. It was issued by the Ministry of Environmental Protection on 2014-01-13 and took effect on 2014-04-01.

Found what you need?

Document preview — HJ 683-2014

National Standard of the People's Republic of China

Issued by: Ministry of Environmental Protection

Contents

  • Foreword...3
  • 1 Application Scope...4
  • 2 Normative References...4
  • 3 Method principle...4
  • 4 Interference and elimination...5
  • 5 Reagents and materials...5
  • 6 Instruments and equipment...6
  • 7 Samples...6
  • 8 Analytical procedures...7
  • 9 Calculation and expression of results...9
  • 10 Precision and accuracy...9
  • 11 Quality assurance and quality control...10
  • 12 Waste treatment...11

1 Application Scope

This Standard specifies high performance liquid chromatography for the determination of aldehyde and ketone compounds in ambient air.

This Standard applies to the determination of thirteen aldehyde and ketone compounds in ambient air, including formaldehyde, acetaldehyde, acrolein, acetone, propionaldehyde, crotonaldehyde, methacrolein, 2-butanone, butyraldehyde, benzaldehyde, valeraldehyde, m-tolualdehyde and n-hexaldehyde. Other aldehyde and ketone compounds can also be used in the analysis of this Standard provided that they pass the suitability validation.

When the sampling volume is 0.05 m3, the detection limit of this Method is 0.28 µg/m3

~ 1.69 µg/m3 and the low-limit of determination is 1.12 µg/m3 ~ 6.76 µg/m3. For details see Annex A.

2 Normative References

This Standard cites the following documents or clauses therein. For undated reference documents, their valid editions apply to this Standard.

HJ/T 55, Technical guidelines for fugitive emission monitoring of air pollutants

HJ/T 194, Manual methods for ambient air quality monitoring

3 Method principle

Use sampling tubes packed with coating 2,4-Dinitrophenylhydrazine (DNPH) to collect a certain volume of air sample. The aldehyde and ketone compounds in sample react with DNPH coated on silica gel in accordance with Equation (1) through the catalytic action of strong acid, generating stable, coloured hydrazine derivatives. After the elution of acetonitrile, use the ultraviolet (360 nm) or diode array detector of high performance liquid chromatographer for detection. Use the retention time for

Use sealing caps to seal both ends of sampling tubes; use tin foil or aluminum foil to wrap sampling tubes up for storage and transportation at low temperature (< 4°C). If analysis can’t be carried out in time, samples shall be stored for not more than 30 days at low temperature (< 4°C).

7.3 Preparation of samples

Add acetonitrile to elute sampling tubes; let acetonitrile to flow past sampling tubes naturally with the flow direction opposite to the direction of air flow. Collect the eluent in a 5 mL volumetric flask before adding acetonitrile to scale; use injection syringe (5.7) to absorb the eluent; filter through syringe filter (5.8); and transfer to a 2 mL brown bottle for determination. If the eluent can’t be analyzed in time after filtration, it can be stored for 30 days at 4°C.

7.4 Preparation of blank samples

7.4.1 Whole-course blanks

Each sampling shall include at least one whole-course blank. bring a sampling tube to site with its both ends open; continue for a sampling cycle without collecting sample in it; seal it in the same way as that of sampling tubes with samples; and bring it to laboratory. Prepare blank samples in accordance with the same procedures of 7.3.

7.4.2 Blank sampling tubes

Take the same batch of sampling tubes in laboratory to prepare blank samples in accordance with the same procedures of 7.3.

8 Analytical procedures

8.1 Reference chromatographic conditions

Mobile phase. acetonitrile/water. Gradient elution. maintain 60% acetonitrile for 20 min;

increase acetonitrile from 60% linearly to 100% within 20 ~ 30 min; then reduce acetonitrile to 60% within 30 ~ 32 min; and maintain for 8 min.

Test wavelength. 360 nm.

Flow rate. 1.0 mL/min.

Sample size. 20 µL.

8.2 Calibration

8.2.1 Preparation of standard series

Measure respectively 100 µL, 200 µL, 500 µL, 1 000 µL and 2 000 µL of standard working solution (5.4) to pour into a 10 mL volumetric flask; use acetonitrile to add to

For the details of the results of precision and accuracy see Annex B.

11 Quality assurance and quality control

11.1 Blank sampling tubes

At least 10% of each batch of sampling tubes shall be taken for blank value testing and the blank value shall meet the following requirements.

Formaldehyde less than 0.15 µg/tube;

Acetaldehyde less than 0.10 µg/tube;

Acetone less than 0.30 µg/tube;

Other matters less than 0.10 µg/tube.

11.2 Whole-course blanks

Each batch of samples shall include at least one whole-course blank and the determination result shall be lower than the method detection limit.

11.3 Parallel double sample

Each batch of samples shall include 10% of parallel double sample; when the sample quantity is less than 10, at least one parallel double sample shall be determined; and the relative deviation of the results of two parallel determinations shall be less than or equal to 25%.

11.4 Sampling flow

Observe from time to time whether the flow of sampler is stable during the sampling period. If the flow at the end of sampling is more than 15% different from the flow at the beginning of sampling, the samples shall be abandoned, and sampling shall be done once again.

11.5 Control of breakthrough capacity

The upper limit of aldoketone content in samples taken (in terms of formaldehyde) shall be less than 75% of the DNPH content of sampling tubes. The breakthrough capacity of aldoketones can be calculated in accordance with Equation (3).

where,

......

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 11 pages — is available in the English PDF.

Referenced standards

How to Download HJ 683-2014

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
11 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

HJ 683-2014

$115.00