Valid

HJ 1225-2021Ambient air-Automatic determination of ozone-Chemiluminescence method (English PDF)

环境空气 臭氧的自动测定 化学发光法

Preview PDF

This is a limited preview

Buy now to download the full PDF (193 pages)

Need quotation for your management?

Get a formal quotation for HJ 1225-2021 in 30 minutes.

Issued by

Ministry of Ecology and Environment

Level / Type

National · Mandatory

Issue date

January 1, 1970

Implementation date

January 1, 1970

Scope

HJ 1225-2021 is the English-translated version of 环境空气 臭氧的自动测定 化学发光法.

HJ 1225-2021 — This standard specifies a chemiluminescence method for the automatic determination of ozone in ambient air. This standard applies to the automatic determination of ozone in ambient air.

When the instrument range is 0 nmol/mol ~ 500 nmol/mol, the detection of this method under the reference state (298.15 K, 101.325 kPa) the limit is 4 µg/m3, and the lower limit of determination is 16 µg/m3.

Found what you need?

Document preview — HJ 1225-2021

National Standard of the People's Republic of China

Issued by: Ministry of Ecology and Environment

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 Terms and Definitions
  • 4 Principles of the method
  • 5 Interference and cancellation
  • 6 Reagents and materials
  • 7 Instruments and equipment
  • 8 Analysis steps
  • 9 Result calculation and representation
  • 10 Accuracy
  • 11 Quality Assurance and Quality Control
  • 12 Notes

1 Scope of application

This standard specifies a chemiluminescence method for the automatic determination of ozone in ambient air.

This standard applies to the automatic determination of ozone in ambient air.

When the instrument range is 0 nmol/mol ~ 500 nmol/mol, the detection of this method under the reference state (298.15 K, 101.325 kPa) the limit is 4 µg/m3, and the lower limit of determination is 16 µg/m3.

2 Normative references

This standard refers to the following documents or clauses thereof. For dated references, only the dated version applies to this standard.

For undated references, the latest edition (including all amendments) applies to this standard.

HJ 193 Technical specification for installation and acceptance of continuous automatic monitoring system for ambient air gaseous pollutants (SO2, NO2, O3, CO) HJ 654 Technical requirements and detection methods of continuous automatic monitoring system for ambient air gaseous pollutants (SO2, NO2, O3, CO) HJ 818 Ambient air gaseous pollutants (SO2, NO2, O3, CO) continuous automatic monitoring system operation and quality control technical specifications

3 Terms and Definitions

The following terms and definitions apply to this standard.

ozone transfer standard According to relevant operating procedures, it can be accurately reproduced or accurately analyzed, and can be traced to a higher level or higher authoritative standard ozone concentration of transportable equipment. Ozone transfer standards are used to transfer the amount of ozone primary standards or to calibrate ozone analyzers at monitoring sites.

Note. The ozone transfer standard is divided into the generation type ozone transfer standard and the analytical ozone transfer standard according to the working principle. According to the position in the step-by-step transmission of ozone value the position is divided into two-level transfer standard, three-level transfer standard and four-level transfer standard.

productive ozone transfer standard this type of transfer standard only contains an ozone generator and does not contain a UV photometer. The odor that occurs is adjusted by adjusting the power of the generator.

The oxygen concentration cannot be measured in real time for the generated ozone concentration. The Generative Ozone Transfer Standard applies only to the quality of on-site ozone analyzers.

Quantitative control work, not suitable for calibrating analytical ozone transfer standards.

analytical ozone transfer standard this type of delivery standard contains a UV photometer that measures the ozone concentration generated by the ozone generator in real time. Analytical ozone delivery standards can be Used to calibrate analytical ozone transfer standards, generative ozone transfer standards and field ozone analyzers. Some analytical ozone delivery standards come with the ozone generator can measure the ozone concentration while generating ozone, and perform real-time feedback adjustment to the ozone generator.

4 Principles of the method

The sample air enters the instrument reaction chamber through the particulate filter at a constant flow, and the ozone is mixed with the excess nitric oxide for an instant reaction After luminescence, the ozone concentration in the sample air is proportional to the luminescence intensity within a certain concentration range.

5 Interference and cancellation

Particulate matter in the air may accumulate in the sampling line or reaction chamber and interfere with the measurement of ozone.

Install a particulate filter to eliminate the effects of particulate matter.

6 Reagents and Materials

6.1 Nitric oxide. x=10000 µmol/mol

The error is within ±10%. It is prepared with nitrogen with a purity of not less than 99.999% and stored in a steel cylinder.

6.2 Filter membrane. The material is polytetrafluoroethylene, and the pore size is less than or equal to 5 µm.

6.3 Zero gas. It is generated by zero gas generator, and can also be provided by zero gas cylinder. The performance index should meet the relevant requirements of HJ 654.If using Synthetic air, in which the concentration of oxygen shall be 20.9% ± 2.0% of the synthetic air.

6.4 Nitrogen. purity >=99.999%.

7 Instruments and equipment

7.1 Sampling line

It should be made of polytetrafluoroethylene, borosilicate glass and other materials that do not chemically react with ozone.

7.2 Particulate filter

Installed between the sampling manifold and the instrument inlet. The material of the filter membrane of the particulate filter should meet the requirements of 6.2, except for the filter membrane Parts should be made of polytetrafluoroethylene, borosilicate glass and other materials that do not chemically react with ozone. If the instrument has a built-in particulate filter, then no external particulate filter is required.

7.3 Ozone Analyzer

The performance index should meet the requirements of HJ 654.The ozone measurement system is shown in Figure 1.

7.4 Ozone Delivery Criteria

Analytical ozone transfer standards with ozone generators are generally used, and generative ozone transfer standards can also be used.

Measuring range. 0 nmol/mol ~ 500 nmol/mol, the minimum display unit is 1 nmol/mol, the performance index of the instrument should conform to HJ 654 phase off requirements.

The ozone flow rate output by the ozone transfer standard should be at least 1 L/min greater than the instrument set flow rate of the ozone analyzer being calibrated.

8 Analysis steps

8.1 Installation and debugging of the instrument

After the newly purchased ozone analyzer (7.3) is installed, various parameters should be set according to the operation manual for debugging. Debug indicators include zero points Noise, detection limit, span noise, indication error, span precision, 24 h zero drift and 24 h span drift, debugging according to HJ 193 implement.

8.2 Inspection

During the operation of the ozone analyzer (7.3), perform zero check, span check, linearity check and flow check according to HJ 818, if If the inspection result is unqualified, the instrument should be calibrated and repaired if necessary. After the repair is completed, the instrument should be checked for linearity, and recalibrate.

8.3 Calibration

8.3.1 Determining the Instrument Range

The range of the ozone analyzer (7.3) should be determined according to the actual concentration of ozone in different seasons. When the ozone concentration is lower than 20% of the range , a lower range should be selected.

8.3.2 Calibration procedure

8.3.2.1 Pass the zero gas into the ozone analyzer (7.3), and after the reading is stable, adjust the output value of the ozone analyzer (7.3) to be equal to zero.

8.3.2.2 Connect the ozone transfer standard (7.4) outlet to the ozone analyzer (7.3) inlet so that the ozone transfer standard (7.4) occurs the ozone concentration is 80% of the operating range of the ozone analyzer (7.3). After the reading is stable, adjust the output value of the ozone analyzer (7.3). like Use the generation type ozone transfer standard to make the output value of the ozone analyzer (7.3) equal to the ozone concentration generated by the ozone transfer standard (7.4);

If using an analytical ozone transfer standard, make the output of the ozone analyzer (7.3) equal to the actual ozone measured by the ozone transfer standard (7.4) concentration.

8.4 Determination of samples

Pass the sample air into the ozone analyzer (7.3), which automatically determines and records the ozone concentration.

9 Result calculation and presentation

9.1 Result calculation

The mass concentration of ozone is calculated according to formula (1).

9.2 Result representation

The measurement result retains integer bits.

10 Accuracy

10.1 Precision

Six laboratories performed six replicate determinations of ozone at concentrations of 16.0 µg/m3, 147 µg/m3 and 964 µg/m3 in the reference state.

The intra-laboratory relative standard deviations were 5.5%-10%, 0.7%-1.4%, and 0.1%-0.2%; the inter-laboratory relative standard deviations were 10%, 1.9% and 1.9%; repeatability limits are 4 µg/m3, 6 µg/m3 and 4 µg/m3 respectively; reproducibility limits are 6 µg/m3 and 16 µg/m3 respectively and 31 µg/m3.

10.2 Correctness

Six laboratories performed six replicate determinations of ozone at concentrations of 16.0 µg/m3, 147 µg/m3 and 964 µg/m3 in the reference state.

The relative errors are -10% to 16%, -1.7% to 2.7% and -1.8% to 1.4%, respectively; the final values of the relative errors are 1.5% ± 21%, 0.4%±3.8% and 0.4%±2.6%.

11 Quality Assurance and Quality Control

11.1 The relevant requirements for the calibration of the ozone analyzer (7.3) shall be implemented in accordance with HJ 818.

11.2 Quantity transfer and quantity traceability of ozone transfer standard (7.4) shall be implemented in accordance with HJ 818.

11.3 The filter membrane holder of the particulate filter should be cleaned at least once every six months; the filter membrane is generally replaced every two weeks.

The frequency of membrane replacement should be increased according to the actual situation.

11.4 The air-tightness of the sampling pipeline should be checked monthly, cleaned every six months, and replaced if necessary.

12 Notes

12.1 When the ozone analyzer (7.3) and the nitrogen dioxide (nitrogen oxide) analyzer are used at the same time, drain the outlet of the ozone analyzer (7.3) Connect a purification tank with oxidant and activated carbon to remove excess nitric oxide and generated nitrogen dioxide in the reaction gas. The purification tank should be replaced regularly Change.

12.2 After replacing the sampling system components or filter membranes, sample air should be collected at a normal flow rate for at least 10 minutes and subjected to saturated adsorption treatment.

The measurement data generated during this period are not considered valid data. This process can also be carried out in the laboratory.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 193 pages — is available in the English PDF.

Referenced standards

How to Download HJ 1225-2021

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
193 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

HJ 1225-2021

$135.00