Scope

HJ 1214-2021 is the English-translated version of 水质 可吸附有机卤素(AOX)的测定 微库仑法.

HJ 1214-2021 — This standard specifies the microcoulometric method for the determination of organic halides (excluding organic fluorides) that can be adsorbed on activated carbon in water.

This standard applies to the determination of adsorbable organic halogens in surface water, groundwater, domestic sewage and industrial wastewater. The determination results of this standard include adsorbable organic halogens in suspended solids in water samples.

When the sampling volume is 100 ml, the detection limit of this method is 0.007 mg/L, and the lower limit of determination is 0.028 mg/L.

Found what you need?

Document preview — HJ 1214-2021

National Standard of the People's Republic of China

Issued by: Ministry of Ecology and Environment

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 Terms and Definitions
  • 4 Principles of the method
  • 5 Interference and cancellation
  • 6 Reagents and materials
  • 7 Instruments and equipment
  • 8 Samples
  • 9 Analysis steps
  • 10 Result calculation and representation
  • 11 Accuracy
  • 12 Quality Assurance and Quality Control
  • 13 Waste Disposal

1 Scope of application

This standard specifies the microcoulometric method for the determination of organic halides (excluding organic fluorides) that can be adsorbed on activated carbon in water.

This standard applies to the determination of adsorbable organic halogens in surface water, groundwater, domestic sewage and industrial wastewater.

The determination results of this standard include adsorbable organic halogens in suspended solids in water samples.

When the sampling volume is 100 ml, the detection limit of this method is 0.007 mg/L, and the lower limit of determination is 0.028 mg/L.

2 Normative references

This standard refers to the following documents or clauses thereof. For dated references, only the dated version applies to this standard.

For undated references, the latest edition (including all amendments) applies to this standard.

HJ/T 91 Technical Specification for Surface Water and Sewage Monitoring HJ 91.1 Technical Specification for Wastewater Monitoring HJ 164 Technical Specification for Groundwater Environmental Monitoring HJ 493 Technical Regulations for the Preservation and Management of Water Quality Samples

3 Terms and Definitions

The following terms and definitions apply to this standard.

Adsorbable organically bound halogens (AOX) Under the test conditions specified in this standard, chlorine, bromine, iodine (not contains the equivalent total amount of organic fluorides), the results are in chlorine.

dissolved organic carbon (DOC) Total organic carbon content in water samples after filtration through a 0.45 m microporous membrane.

4 Principles of the method

The water sample is acidified by nitric acid, adsorbed by activated carbon, and then rinsed with sodium nitrate-nitric acid solution to remove the inorganic halide adsorbed on the activated carbon.

The adsorbed activated carbon is pyrolyzed and burned in the oxygen flow to generate hydrogen halide gas, and the hydrogen halide gas is passed into the micro-coulomb cell, and the micro-coulomb method is used.

The amount of halide ions was measured, and the result was expressed as the mass concentration of chlorine.

5 Interference and cancellation

5.1 The free chlorine in the water sample can lead to high AOX measurement results, and sodium sulfite solution (6.14) should be added immediately after sampling to eliminate dryness.

disturb.

5.2 When the inorganic chloride concentration in the sample is greater than 1 g/L, it should be diluted and measured; the sample with a lower AOX value contains more than 1 g/L Chloride ions may lead to high determination results. Adding the same concentration of chloride ions (such as NaCl) to laboratory blank samples can eliminate dryness.

disturb.

5.3 Inorganic bromide and iodide in water samples will lead to high determination results.

5.4 When the DOC in the sample exceeds 10 mg/L, it should be diluted and measured.

5.5 The organic bromide and organic iodide in the sample generate high-valence oxides of bromine and iodine during the combustion process, and this part of AOX cannot be measured, which will result in a low measurement result.

5.6 Alcohols, aromatic compounds and hydroxy acids can cause low results.

5.7 For samples containing living cells (such as microorganisms, algae, etc.), the determination results are high due to their own chloride content.

Add nitric acid to acidify the water sample, destroy the algae and microbial cells, dissolve the chloride contained in the water sample, and then go through the elution step to eliminate the interference.

6 Reagents and Materials

Unless otherwise specified, analytical reagents that meet national standards were used in the analysis. The experimental water was distilled water or pure water without AOX.

6.1 Nitric acid. rho=1.42 g/ml.

6.2 Hydrochloric acid. rho=1.19 g/ml.

6.3 Sulfuric acid. rho=1.84 g/ml.

6.4 Sodium nitrate (NaNO3).

6.5 Sodium sulfite (Na2SO3).

6.6 p-Chlorophenol (C6H5ClO).

6.7 2-Chlorobenzoic acid (C7H5ClO2).

6.8 Potassium iodide (KI).

6.9 Soluble starch.

6.10 Nitric acid solution. c(HNO3)=0.02 mol/L.

Measure 1.36 ml of nitric acid (6.1) and dilute to 1000 ml with water.

6.11 Hydrochloric acid solution. c(HCl)=0.010 mol/L.

Measure 0.83 ml of hydrochloric acid (6.2) and dilute to 1000 ml with water.

6.12 Sodium nitrate-nitric acid solution. c=0.2 mol/L.

Weigh 17.00 g of sodium nitrate (6.4) and dissolve it in water, add 25 ml of nitric acid (6.1), and dilute to 1000 ml with water. Sodium nitrate - nitric acid solution the liquid can be stored for 3 months in the amber glass bottle.

6.13 Sodium nitrate-nitric acid eluent. c=0.01 mol/L, pH~1.7.

Take 50 ml of sodium nitrate-nitric acid solution (6.12) and dilute to 1000 ml with water. Sodium nitrate-nitric acid eluent is safe in amber glass vials Save 1 month.

6.14 Sodium sulfite solution. c(Na2SO3)=1 mol/L.

Weigh 126 g of sodium sulfite (6.5), dissolve it in water, and dilute to 1000 ml with water. This solution is refrigerated at 4 °C and can be stored for 1 moon.

6.15 p-chlorophenol stock solution. rho(Cl)=200 µg/ml.

Accurately weigh 0.0725 g of p-chlorophenol (6.6) and dissolve it in water, transfer it to a 100 ml volumetric flask, and dilute to the mark with water. p-chlorophenol storage the prepared solution is stored in a glass bottle and refrigerated at 4°C for 1 month. Commercially available certified standard solutions can also be used, refer to standard solutions the certificate is saved.

6.16 p-chlorophenol intermediate solution. rho(Cl)=10.0 µg/ml.

Take 5.00 ml of the p-chlorophenol stock solution (6.15), transfer it to a 100 ml volumetric flask, dilute to the mark with water, and prepare for immediate use.

6.17 Use solution of p-chlorophenol. rho(Cl)=1.00 µg/ml.

Take 10.00 ml of p-chlorophenol intermediate solution (6.16), transfer it into a 100 ml volumetric flask, dilute to the mark with water, and prepare for immediate use.

6.18 2-Chlorobenzoic acid stock solution. rho(Cl)=250 µg/ml.

Accurately weigh 0.1104 g of 2-chlorobenzoic acid (6.7), dissolve it in water, transfer it to a 100 ml volumetric flask, and make up to the mark with water. 2-Chloro the benzoic acid stock solution is stored in a glass bottle and refrigerated at 4°C for 1 month.

6.19 2-chlorobenzoic acid intermediate solution. rho(Cl)=10.0 µg/ml.

Take 4.00 ml of 2-chlorobenzoic acid stock solution (6.18), transfer it into a 100 ml volumetric flask, dilute to the mark with water, and prepare for immediate use.

6.20 2-chlorobenzoic acid working solution. rho(Cl)=1.00 µg/ml.

Take 10.00 ml of the 2-chlorobenzoic acid intermediate solution (6.19), transfer it into a 100 ml volumetric flask, dilute to the mark with water, and prepare for immediate use.

Note. 2-Chlorobenzoic acid is difficult to dissolve in water and can be prepared 1 day in advance.

6.21 Check solution.

Accurately pipette 1.00 ml, 5.00 ml, 10.0 ml, 20.0 ml and 30.0 ml of p-chlorophenol working solution (6.17) or 2-chlorobenzoic acid, respectively Use liquid (6.20) into a 100 ml volumetric flask, and make up to the mark with water. The concentrations of AOX contained in this verification solution are. 0.010 mg/L, 0.050 mg/L, 0.100 mg/L, 0.200 mg/L and 0.300 mg/L, ready for immediate use.

6.22 Starch indicator. rho=10 g/L.

Weigh 1.0 g of soluble starch (6.9), make a paste with a small amount of water, slowly add 100 ml of warm water, and continue to cook until the solution is clear and cooled.

After use, it is ready for immediate use.

6.23 Activated carbon. iodine value >= 1050, chloride content 7 Instruments and equipment

7.1 Sampling container

Rigid glass bottle, 500 ml.

7.2 Adsorbable organic halogen analyzer

7.2.1 Instrument composition

The main components of the instrument should include burners, detectors and drying tubes, and a complete set of commercial adsorbable organic halogen analyzers can also be used.

7.2.2 Burners

7.2.2.1 Combustion furnace. the heating temperature is adjustable, the maximum temperature is not lower than 950 °C, both horizontal furnace and vertical furnace can be used.

7.2.2.2 Combustion tube. a quartz tube with a diameter of 2 cm to 4 cm and a length of about 30 cm.

7.2.3 Detector

Microcoulometer, capable of detecting 1 g Cl- with a relative standard deviation of 8 samples

8.1 Collection and storage of samples

Collect and save water samples according to the relevant regulations of HJ/T 91, HJ 91.1, HJ 164 and HJ 493.Sampling, transportation and storage of water samples Glassware should be used for storage, and the sampling volume should be >=500 ml.

To avoid oxidizing agents such as free chlorine in the sample, add 10 ml of sodium sulfite solution (6.14) per liter of water sample immediately after sampling.

Note. To test whether the water sample contains free chlorine. take a small amount of water sample into a clean test tube, add a small amount of potassium iodide (6.8), and then add 3 to 5 drops of starch to indicate (6.22), if it shows blue, it proves that there is free chlorine in the water, and starch-potassium iodide test paper can also be used to test.

After sampling, add 2 ml of nitric acid (6.1) per liter of water to make pH 9 Analysis steps

9.1 Instrument Reference Conditions

9.1.1 Adjust the temperature of the combustion furnace (7.2.2.1) not to be lower than 950 °C.

9.1.2 Adjust the flow rate of oxygen (6.27) to about 150 ml/min. If a complete set of commercial instruments is used, refer to the information provided by the instrument manufacturer for relevant parameters.

material.

9.2 Calibration of the Microcoulomb Counter

Before measurement, check whether the instrument is in normal working condition, and use at least one of the following methods to check.

a) Directly inject 50 l~100 l hydrochloric acid solution (6.11) or the calibration sample specified by the instrument manufacturer into the coulomb cell, and measure its concentration

9.3 Determination

9.3.1 Sample determination

Connect the combustion pipe (7.2.2.2) and the drying pipe (7.2.4), and inject sulfuric acid (6.3) into the drying pipe. The prepared sample (8.2.2 or 8.2.3) Push it into the heating zone in the combustion tube, and titrate and count by Coulomb cell.

Note. Relevant factors in the experiment may affect the measurement results, such as. the amount of activated carbon, sample volume, combustion furnace volume, residence time, combustion temperature and gas If the relevant parameters are changed, the performance and status of the instrument should be checked according to 12.2.1.

9.3.2 Blank test

Carry out the determination of the blank sample (8.3) according to the same procedure as the sample determination (9.3.1).

10 Result calculation and presentation

10.1 Result calculation

The mass concentration (mg/L) of AOX in the sample is expressed as the mass concentration of chlorine, and is calculated according to formula (4).

10.2 Result representation

The number of decimal places of the measurement results is consistent with the detection limit of the method, and the maximum number of significant figures is 3.

Note. If the water sample is filtered, the test result only indicates the AOX concentration in the filtered water, which should be stated in the test report.

11 Accuracy

11.1 Precision

The 6 laboratories determined the concentrations of AOX in the range of 0.016 mg/L to 0.017 mg/L, 0.016 mg/L to 0.038 mg/L, 0.033 mg/L~5.63 mg/L, 0.043 mg/L~8.63 mg/L and 0.016 mg/L~0.879 mg/L of surface water, groundwater and sewage the actual samples of the effluent from the processing plant, pulp and papermaking wastewater and textile dyeing and finishing wastewater were repeatedly determined for 6 times.

The relative standard deviation ranges in the laboratory are. 2.0%-4.3%, 3.6%-9.4%, 2.8%-9.2%, 2.3%-7.8%, 1.2%-9.3%.

11.2 Correctness

Six laboratories carried out six repeated tests on standard samples containing AOX concentrations of 0.050 mg/L, 5.00 mg/L and 15.0 mg/L.

Certainly.

The relative errors of the oscillation adsorption method are. -14.8%~9.6%, -7.6%~0.7%, -6.7%~12.0%;

The final values of the relative errors of the oscillation adsorption method are. -1.9%±16.4%, -4.0%±5.4%, -1.0%±12.8%;

The relative errors of the column adsorption method are. -8.0% to 1.8%, -8.4% to 1.2%, and -10.7% to 0;

The final values of the relative errors of the column adsorption method were. -2.6%±8.1%, -3.2%±7.2%, -3.8%±7.2%.

The 6 laboratories determined the concentrations of AOX in the range of 0.016 mg/L to 0.017 mg/L, 0.016 mg/L to 0.038 mg/L, 0.033 mg/L~5.63 mg/L, 0.043 mg/L~8.63 mg/L and 0.016 mg/L~0.879 mg/L surface water 0.050 mg/L and 0.100 mg/L), groundwater (0.050 mg/L and 0.100 mg/L for spiking), sewage treatment plant effluent (spiking for 0.100 mg/L) 1.00 mg/L and 3.00 mg/L), pulp and papermaking wastewater (5.00 mg/L and 10.0 mg/L spiked), textile dyeing and finishing wastewater (spiked 5.00 mg/L and 10.0 mg/L) the actual samples were determined in 6 replicates.

The recovery rates of standard additions ranged from 88.6% to 99.6%, 89.6% to 102%, 88.0% to 103%, 83.0% to 102%, and 81.7% to 104%, 81.2%~101%, 82.0%~105%, 80.8%~101%, 82.4%~103%, 81.8%~109%.

12 Quality Assurance and Quality Control

12.1 Blank test

1 laboratory blank test is performed for every 10 consecutive samples or each batch (<=10 samples/batch), and the blank value should be lower than Otherwise, the adsorption step, the combustion step and the determination of microcoulombs should be checked separately.

12.2 Calibration

12.2.1 When the instrument is turned on continuously, the whole process measurement of the verification solution (6.21) shall be carried out once a month, and the measurement results shall be subjected to regression analysis.

The correlation coefficient should be greater than or equal to 0.999, and the slope should be in the range of 0.95 to 1.05.The full process determination of the verification solution is only used for instrument performance and condition inspection.

12.2.2 Each batch of samples should be checked for the intermediate point of a standard solution, or a standard solution with a similar AOX concentration to the sample to be tested should be prepared the whole process measurement is carried out at the point, and the relative error between the measurement result and the standard solution concentration should be within ±10%.

12.3 Parallel samples

Measure 10% of the parallel samples for each batch of samples. When the number of samples is less than 10, at least one parallel sample should be measured. Sample determination result >= 0.100 mg/L when the measurement result of the parallel sample is less than ±10%, the relative deviation of the measurement result of the parallel sample should be within ±10%; when the measurement result of the sample is less than 0.100 mg/L, the measurement result of the parallel sample is absolutely the difference should be

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 8 pages — is available in the English PDF.

Referenced standards

How to Download HJ 1214-2021

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
8 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

HJ 1214-2021

$150.00