Scope
HJ 1044-2019 is the English-translated version of 环境空气 二氧化硫的自动测定 紫外荧光法.
HJ 1044-2019 — This standard specifies the ultraviolet fluorescence method for automatic determination of sulfur dioxide in ambient air. This standard applies to the automatic determination of sulfur dioxide in ambient air.
When the measuring range of the instrument is (0 to 500) nmol/mol, the detection limit of this method in the reference state is 3 µg/m3. The lower limit is 12 µg/m3; the detection limit of the method under standard conditions is 3 µg/m3, and the lower detection limit is 12 µg/m3.
Found what you need?
Document preview — HJ 1044-2019
National Standard of the People's Republic of China
- ICS
- 13.040.20
- Classification
- Z 15
Issued by: Ministry of Ecology and Environment
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Methodology
- 4 Interference and cancellation
- 5 Reagents and materials
- 6 Instruments and equipment
- 7 Analysis steps
- 8 Calculation and Representation of Results
- 9 Precision and accuracy
- 10 Quality Assurance and Quality Control
- 11 Notes
Foreword
In order to implement the "Environmental Protection Law of the People's Republic of China" Environment, protect human health, standardize the automatic determination method of sulfur dioxide in ambient air, and formulate this standard.
This standard specifies the ultraviolet fluorescence method for automatic determination of sulfur dioxide in ambient air.
This standard is issued for the first time.
This standard is formulated by the Department of Eco-Environmental Monitoring, Laws and Standards Department of the Ministry of Ecology and Environment.
This standard was drafted. Dalian Ecological Environment Monitoring Center, Liaoning Province.
Verification units of this standard. Harbin Environmental Monitoring Center Station, Hangzhou Environmental Monitoring Center Station, Anshan Ecology, Liaoning Province Environmental Monitoring Center, Guangzhou Environmental Monitoring Center Station, Shenyang Ecological Environment Monitoring Center in Liaoning Province and Qingdao Ecological Environment Center in Shandong Province Environmental Monitoring Center.
This standard was approved by the Ministry of Ecology and Environment on October 24, 2019.
This standard will be implemented from April 24, 2020.
This standard is explained by the Ministry of Ecology and Environment.
Automatic determination of sulfur dioxide in ambient air by ultraviolet fluorescence method WARNING. Sulfur dioxide is a toxic gas. Prevent leakage during operation and do protective work as required.
1 Scope
This standard specifies the ultraviolet fluorescence method for automatic determination of sulfur dioxide in ambient air.
This standard applies to the automatic determination of sulfur dioxide in ambient air.
When the measuring range of the instrument is (0 to 500) nmol/mol, the detection limit of this method in the reference state is 3 µg/m3.
The lower limit is 12 µg/m3; the detection limit of the method under standard conditions is 3 µg/m3, and the lower detection limit is 12 µg/m3.
2 Normative references
This standard refers to the following documents or clauses therein. For undated references, the valid version applies to this standard.
HJ 193 Ambient air gaseous pollutants (SO2, NO2, O3, CO) continuous automatic monitoring system installation and acceptance technology specification HJ 654 Ambient air gaseous pollutants (SO2, NO2, O3, CO) continuous automatic monitoring system technical requirements and inspection Test method HJ 663 Technical Specification for Ambient Air Quality Evaluation (Trial) HJ 818 Ambient air gaseous pollutants (SO2, NO2, O3, CO) continuous automatic monitoring system operation and quality control technology Technical specifications
3 Method principle
The sample air enters the instrument reaction chamber through the particulate filter at a constant flow rate, and the sulfur dioxide molecules are affected by the wavelength Excited sulphur dioxide molecules are generated after exposure to ultraviolet light at.200 nm to 220 nm, and the wavelength is emitted during the return to the ground state for 240 nm to 420 nm fluorescence, the concentration of sulfur dioxide in the sample air is proportional to the fluorescence intensity within a certain concentration range.
4 Interference and cancellation
4.1 When the sample air contains 2155 µg/m3 methane, it will affect the measurement result of sulfur dioxide by 3 µg/m3. It is normal the interference of methane in the ambient air is negligible.
4.2 When the sample air contains 6939 µg/m3 hydrogen sulfide, the effect on the sulfur dioxide measurement result is not higher than 1 µg/m3, which is normal.
The interference of hydrogen sulfide in the ambient air is negligible.
4.3 When the sample air contains 123 µg/m3 of nitric oxide, it will affect the measurement result of sulfur dioxide by 3 µg/m3, which is normal.
The interference of nitric oxide in the ambient air is negligible.
4.4 When the sample air contains aromatic hydrocarbons, it will affect the measurement result of sulfur dioxide and can be removed by a hydrocarbon remover.
5 Reagents and materials
5.1 Zero gas. Zero gas is generated by a zero gas generating device, and can also be provided by a zero gas cylinder. The performance index of zero gas should meet HJ 654 Requirements. If synthetic air is used, the oxygen concentration shall be (20.9 ± 2.0)% of the synthetic air.
5.2 Standard gas. Certified reference material of sulfur dioxide, the unit is µmol/mol.
5.3 Filter membrane. The material is polytetrafluoroethylene, and the pore size is <=5 µm.
6 Instruments and equipment
6.1 Injecting pipeline. should be polytetrafluoroethylene, fluorinated polyethylene propylene, stainless steel that does not chemically react with sulfur dioxide or borosilicate glass.
6.2 Particulate matter filter. installed between the sampling manifold and the instrument inlet. Particle filter other than the filter membrane it should be polytetrafluoroethylene, fluorinated polyethylene propylene, stainless steel or borosilicate glass that does not react with sulfur dioxide and other materials. The instrument does not require an external particulate filter if it has a built-in particulate filter.
6.3 Sulfur dioxide analyzer. The performance index shall meet the requirements of HJ 654. The sulfur dioxide measurement system is shown in Figure 1.
1-air inlet; 2-roof; 3-fan; 4-dehumidifying device; 5-injection pipeline; 6-way valve; 7-zero gas; 8-standard gas; 9- particulate filter;
10-Sulfur dioxide analyzer; 11-Hydrocarbon remover; 12-Reaction chamber; 13-Signal output; 14-Flow controller; 15-Pump; 16-Evacuation port;
17-Data output.
Figure 1 Schematic diagram of sulfur dioxide measurement system
7 Analysis steps
7.1 Installation and commissioning of the instrument
After the newly purchased instrument is installed, various parameters should be set and debugged according to the operation manual. Debugging indicators include zero noise, Minimum detection limit, range noise, indication error, range accuracy, 24 h zero point drift and 24 h range drift.
The detection methods and indicators are implemented in accordance with HJ 193.
7.2 Inspection
During the operation of the instrument, zero point check, range check and linearity check are required. The inspection method is in accordance with the appendix of HJ 818.
Record B is executed. If the inspection results are unsatisfactory, the instrument needs to be calibrated and repaired if necessary.
After the instrument is repaired, perform a linear check and recalibrate the instrument.
7.3 Calibration
7.3.1 Determine instrument range
Instrument range should be determined according to the actual concentration of sulfur dioxide in different seasons. When the sulfur dioxide concentration is below the range at 20%, a lower range should be selected.
7.3.2 Calibration procedure
7.3.2.1 Pass zero gas into the instrument and adjust the output value of the instrument to zero after the reading is stable.
7.3.2.2 Pass a standard gas with a concentration of 80% of the range into the instrument. After the reading is stable, adjust the output value of the instrument to be equal to the standard.
Gas concentration value.
7.4 Determination of samples
The sample air was passed into the instrument for automatic determination and recording of the volume concentration of sulfur dioxide.
8 Results calculation and representation
8.1 Calculation of results
When used for ambient air quality monitoring, fugitive emissions monitoring or indoor air quality monitoring, Result calculations are performed for the states required by the emission standards and emission standards.
The mass concentration of sulfur dioxide is calculated according to formula (1).
In the formula. rho--mass concentration of sulfur dioxide, µg/m3;
64--Molar mass of sulfur dioxide, g/mol;
Vm--molar volume of sulfur dioxide, 22.4 in the standard state and 24.5 in the reference state, L/mol;
phi--Volume concentration of sulfur dioxide, nmol/mol.
8.2 Results representation
The measurement results are reserved in whole digits, and the statistical method of monitoring data used for air quality evaluation is implemented in accordance with HJ 663.
9 Precision and accuracy
9.1 Precision
Six laboratories performed 6 sulfur dioxide standard gases at concentrations of 18 µg/m3, 137 µg/m3, and 784 µg/m3 Repeated determination. the relative standard deviations in the laboratory were 1.2% to 3.3%, 0.6% to 3.3%, and 0.3% to 1.2%;
The relative standard deviations are 4.0%, 2.3% and 0.8% respectively; the repeatability limits are 3 µg/m3, 8 µg/m3 and 16 µg/m3;
Reproducibility limits are 3 µg/m3, 10 µg/m3, and 24 µg/m3, respectively.
9.2 Accuracy
Six laboratories performed 6 sulfur dioxide standard gases at concentrations of 18 µg/m3, 137 µg/m3, and 784 µg/m3 Repeated determination. the relative errors are -6.0% to 2.4%, -4.3% to 1.6%, and 0.3% to 1.7%; the final value of the relative error is They are -0.3% ± 6.2%, -0.6% ± 5.2%, and 0.9% ± 1.2%.
10 Quality Assurance and Quality Control
10.1 The frequency and indicators of the zero check, range check, linearity check, flow rate check, and calibration of the instrument shall be performed in accordance with HJ 818.
Row.
10.2 The filter holder of the particulate filter is cleaned at least once every six months; the filter is generally replaced every two weeks, and the particulate matter is concentrated.
In areas with high temperature or high concentration periods, the frequency of replacement should be increased depending on the actual pollution of the filter membrane.
10.3 The sampling pipeline should be checked for air tightness monthly, cleaned every six months, and replaced if necessary.
11 Matters needing attention
After replacing the sampling system components and membranes, the sample air should be collected at normal flow for at least 10 minutes for saturated adsorption the measurement data generated during the process is not regarded as valid data. This process can also be performed in the laboratory.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 6 pages — is available in the English PDF.
Referenced standards
How to Download HJ 1044-2019
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
HJ 1409-2025 — Technical guidelines for environmental impact assessment - Marine ecology and environment
HJ 1410.1-2025 — Basic dataset of derivation of water quality criteria for aquatic organisms - Part 1: General principles
HJ 1410.10-2025 — Basic dataset of derivation of water quality criteria for aquatic organisms - Part 10: Mollusca
Secure payment via Stripe
Payments accepted
HJ 1044-2019
$120.00