HJ 1013-2018Specifications and test procedures for nonmethane hydrocarbons continuous emission monitoring system in stationary sources (English PDF)

固定污染源废气非甲烷总烃连续监测系统技术要求及检测方法

Preview PDF

This is a limited preview

Buy now to download the full PDF (30 pages)

Need quotation for your management?

Get a formal quotation for HJ 1013-2018 in 30 minutes.

Issued by

Ministry of Ecological Environment

Level / Type

National · Mandatory

Issue date

December 29, 2018

Implementation date

July 1, 2019

Scope

HJ 1013-2018 is the English-translated version of 固定污染源废气非甲烷总烃连续监测系统技术要求及检测方法.

HJ 1013-2018 — This standard specifies the composition structure, technical requirements, performance indicators, detection methods of nonmethane hydrocarbons continuous emission monitoring system in stationary sources.

This standard applies to the design, production, testing of nonmethane hydrocarbons continuous emission monitoring system in stationary sources.

Found what you need?

Document preview — HJ 1013-2018

National Standard of the People's Republic of China

Issued by: Ministry of Ecological Environment

Contents

  • Foreword...3
  • 1 Scope...4
  • 2 Normative references...4
  • 3 Terms and definitions...4
  • 4 System composition and structure...5
  • 5 Technical requirements...7
  • 6 Performance indicators...11
  • 7 Testing method...13
  • 8 Quality assurance...27
  • 9 Testing items...30

Foreword

This standard is hereby formulated, to implement the "Environmental Protection

Law of the People's Republic of China" and "The Air Pollution Prevention Law

of the People's Republic of China", protect the ecological environment and

human health, regulate the performance and quality of the nonmethane

hydrocarbons continuous emission monitoring system in stationary sources,

implement monitoring for the emission pollutants from the stationary

atmospheric pollution source.

This standard specifies the main technical requirements, test items, test

methods for the nonmethane hydrocarbons continuous emission monitoring

system in stationary sources.

Appendices A and C of this standard are normative appendixes. Appendix B is

an informative appendix.

This standard is issued for the first time.

This standard was formulated by the Department of Ecological Environment

Monitoring, the Department of Regulations and Standards of the Ministry of

Ecological Environment.

Drafting organization of this standard. China Environmental Monitoring Station.

This standard was approved by the Ministry of Ecological Environment, on

December 29, 2018.

This standard will be implemented on July 1, 2019.

This standard is interpreted by the Ministry of Ecological Environment.

Specifications and test procedures for nonmethane

hydrocarbons continuous emission monitoring

system in stationary sources

1 Scope

This standard specifies the composition structure, technical requirements,

performance indicators, detection methods of nonmethane hydrocarbons

continuous emission monitoring system in stationary sources.

This standard applies to the design, production, testing of nonmethane

hydrocarbons continuous emission monitoring system in stationary sources.

2 Normative references

This standard refers to the following documents or their clauses. For undated

reference documents, their valid versions are applicable to this standard.

GB 3836.1 Explosive atmospheres - Part 1.Equipment - General

requirements

GB/T 4208 Degrees of protection provided by enclosure (IP code)

GB/T 16157 The determination of particulates and sampling methods of

gaseous pollutants emitted from exhaust gas of stationary source

HJ 38 Stationary source emission - Determination of total hydrocarbons,

methane and nonmethane hydrocarbons - Gas chromatography

HJ 75 Specifications for continuous emissions monitoring of SO2, NOX, and

particulate matter in the flue gas emitted from stationary sources

HJ 76 Specifications and test procedures for continuous emission monitoring

system for SO2, NOX and particulate matter in flue gas emitted from

stationary

3 Terms and definitions

The following terms and definitions apply to this standard.

3.1

Nonmethane hydrocarbons (NMHC)

Under the conditions specified in the HJ 38 standard, the sum of other

gaseous organic compounds than methane, which respond on the hydrogen

flame ionization detector, as calculated in carbon, unless otherwise specified.

3.2

Nonmethane hydrocarbons continuous emission monitoring system

(NMHC-CEMS)

All equipment, which is required to continuously monitor the emission

concentration and volume of non-methane hydrocarbons in exhaust gas

from stationary sources, as referred to as NMHC-CEMS.

3.3

Analysis cycle time

The time interval, between the two sets of measurement results, which is

given when the system is running continuously.

4 System composition and structure

4.1 System composition

The stationary pollution source NMHC-CEMS consists of a non-methane

hydrocarbon monitoring unit, a waste gas parameter monitoring unit, a data

collection and processing unit, as shown in Figure 1.The system measures the

concentration of non-methane hydrocarbons in the exhaust gas AND the

exhaust gas parameters (temperature, pressure, flow rate or flow, humidity,

etc.).

4.2 System structure

The NMHC-CEMS's system structure mainly includes sample collection and

transmission devices, pretreatment equipment, analysis instruments, data

collection and transmission equipment, other auxiliary equipment, etc.

Depending on the measurement method and principle of the system, the

system may consist of all or part of the above structure.

4.2.3 Analytical instruments

The analytical instrument is used to measure and analyze the collected waste

gas samples from the pollution source. For specific technical requirements, see

5.4.3.

4.2.4 Data collection and transmission equipment

Data collection and transmission equipment is used to collect, process, store

monitoring data. It can transmit monitoring data and equipment working status

information, according to the commands from central computer. See 5.4.4 for

specific technical requirements.

5 Technical requirements

5.1 Appearance requirements

5.2 Working conditions

NMHC-CEMS shall work normally under the following conditions.

5.3.2 Insulation strength

When the ambient temperature is (15 ~ 35) °C and the relative humidity is <=

85%, the system shall last for 1 min, under 1500 V (effective value), 50 Hz sine

wave test voltage; there shall be no breakdown or arcing.

5.3.3 The system shall have a leakage protection device and good grounding

measures, to prevent damage to the system, which is caused by lightning

strikes.

5.4 Functional requirements

5.4.1 Requirements for sample collection and transmission devices
5.4.1.1 The sample collection and transmission device shall select the materials,

which are resistant to high temperature, corrosion, non-absorptive, non-reactive

with the pollutants under test; meanwhile, it shall not affect the normal

measurement of the pollutants under test.

5.4.1.6 The sampling pump shall have sufficient suction capacity, to overcome

the negative pressure of the flue; ensure that the sampling flow is accurate,

reliable, and relatively stable.

5.4.2 Pretreatment equipment
5.4.2.1 The pretreatment equipment and its components shall be easy to clean

and replace. It shall be made of materials, which do not absorb and do not react

with the pollutants under test.

5.4.2.2 In order to prevent particles from polluting the analyzer, a fine filter can

be set, before the gas sample enters the analyzer. The fine filter shall be able

to filter particles, which have a particle size of (0.5 ~ 2) µm at least.

5.4.3 Requirements of analytical instrument
5.4.4.4 It has a digital signal output function.
5.4.4.5 It has Chinese data collection, recording, processing and control

software. For data collection, recording, processing requirements, see

Appendix B in HJ 76.

5.4.6 Requirements of calibration function
5.4.6.1 The system shall be able to be calibrated manually and/or automatically.
5.4.6.2 For the system, which adopts the extraction measurement method, it

shall have a fixed and easy-to-operate calibration gas full-system calibration

function.

6 Performance indicators

6.1 Laboratory testing

6.1.1 Analysis cycle

System analysis cycle. <= 2 min.

6.1.5 24h drift

24 h zero drift and span drift. Not more than ± 3% of full scale.

6.1.6 The influence of environmental temperature changes

When the ambient temperature changes within the span of (15 ~ 35) °C, the

change of non-methane hydrocarbon's indication value. Not more than ±5% of

full scale.

6.1.10 Response factor

When the system measures non-methane hydrocarbons, the relative mass

response factors of other VOCs, as relative to propane, must meet a certain

span of requirements, as shown in Table 1.

6.2 On-site detection of pollutant discharge

6.2.1 Analysis cycle

System analysis cycle. <= 3 min.

6.2.2 24h drift

24 h zero drift and span drift. Not more than ± 3% of full scale.

6.2.3 Accuracy

When using the reference method to make measurement, the average

concentration of the non-methane hydrocarbon is.

6.2.4 Performance indicators of exhaust gas parameters

The performance indicator requirements of exhaust gas parameters (oxygen,

flow rate, flue gas temperature, humidity) shall meet the relevant requirements

of HJ 76.

7 Testing method

7.1 Laboratory testing requirements and methods

7.1.1 General requirements
7.1.1.1 Take at least 3 sets of systems of the same model for simultaneous

testing, at the designated laboratory site.

7.1.1.5 If the test is interrupted due to a system failure, restart the test after the

system returns to normal; the completed test indicators and data will be invalid.

During the test period, the number of system failures per unit (set) is <= 2.

7.1.1.6 For systems with automatic calibration function, it may set any cycle;

the system will perform automatic calibration; during the testing, the automatic

calibration cycle shall be set to >= 24 h.

7.1.2 Standard material requirements
7.1.2.1 Calibration gas. Commercially available certified calibration gas, which

has an uncertainty of <= 2.0%.

7.1.3 Laboratory testing methods
7.1.3.1 Analysis cycle

The analysis cycle time refers to the time interval, between the two sets of

measurement results, which are given, during continuous operation of NMHC-

CEMS. There are 3 measurements in 3 consecutive days. The daily analysis

cycle time shall meet the requirements of Table 3.

7.1.3.3 Repeatability

After the system under test is running stably, lead in the span calibration gas.

Record the measured value, after the displayed value is stable. Repeat the

above operation, at least 6 times, by the use of the span calibration gas of the

same concentration. Calculate the repeatability (relative standard deviation) of

the system under test, according to formula (2). The results shall meet the

requirements of Table 3.

7.2.2 On-site testing methods
7.2.2.1 Analysis cycle

The on-site testing method of the analysis cycle time is consistent with the

laboratory testing method, as shown in 7.1.3.1.The results shall meet the

requirements of Table 4.

7.2.2.2 24h drift

The detection methods of on-site 24h zero drift and span drift are consistent

with the laboratory detection methods, as shown in 7.1.3.5.The results shall

meet the requirements of Table 4.

7.2.2.4 Exhaust gas parameters

Refer to HJ 76, for the detection method of exhaust gas parameters (oxygen,

flow rate, smoke temperature, humidity).

8 Quality assurance

8.1 Installation quality assurance

8.1.1 The installation location and on-site supporting environmental conditions

shall meet the requirements of the HJ 75 standard.

8.2 Testing quality assurance

8.2.1 The on-site testing shall be carried out, under the condition that the

stationary pollution source discharges pollutants normally. During the initial

inspection and re-inspection, a dedicated person must be responsible for

supervising the working conditions AND maintain it relatively stable during the

test.

8.2.4 For complete extraction and dilution extraction systems, when performing

zero and span calibration, in principle, the zero gas and calibration gas must

pass through the same path as the sample gas (such as sample collection

device, transmission pipeline, pretreatment, etc.).

8.3 Daily operation quality assurance

8.3.1 Regular calibration

9 Testing items

For the stationary pollution source waste gas NMHC-CEMS, the laboratory

testing and on-site testing items are as shown in Table 3 and Table 4.The on-

site testing items of exhaust gas parameters are as shown in Table 5.See

Appendix B, for the relevant record form.

HJ

ENVIRONMENTAL PROTECTION STANDARD OF THE

PEOPLE'S REPUBLIC OF CHINA

Specifications and test procedures for nonmethane

hydrocarbons continuous emission monitoring

system in stationary sources

Issued on. DECEMBER 29, 2018

Implemented on. JULY 01, 2019

Issued by. Ministry of Ecological Environment

Table of Contents

Foreword... 3

1 Scope... 4

2 Normative references... 4

3 Terms and definitions... 4

4 System composition and structure... 5

5 Technical requirements... 7

6 Performance indicators... 11

7 Testing method... 13

8 Quality assurance... 27

9 Testing items... 30

Appendix A (Normative) NMHC-CEMS daily report, monthly report, annual

report... 33

Appendix B (Informative) Original record of laboratory testing and on-site

testing... 38

Appendix C (Normative) Calculation of conversion of volume concentration to

mass concentration in carbon... 47

Foreword

This standard is hereby formulated, to implement the "Environmental Protection

Law of the People's Republic of China" and "The Air Pollution Prevention Law

of the People's Republic of China", protect the ecological environment and

human health, regulate the performance and quality of the nonmethane

hydrocarbons continuous emission monitoring system in stationary sources,

implement monitoring for the emission pollutants from the stationary

atmospheric pollution source.

This standard specifies the main technical requirements, test items, test

methods for the nonmethane hydrocarbons continuous emission monitoring

system in stationary sources.

Appendices A and C of this standard are normative appendixes. Appendix B is

an informative appendix.

This standard is issued for the first time.

This standard was formulated by the Department of Ecological Environment

Monitoring, the Department of Regulations and Standards of the Ministry of

Ecological Environment.

Drafting organization of this standard. China Environmental Monitoring Station.

This standard was approved by the Ministry of Ecological Environment, on

December 29, 2018.

This standard will be implemented on July 1, 2019.

This standard is interpreted by the Ministry of Ecological Environment.

Specifications and test procedures for nonmethane

hydrocarbons continuous emission monitoring

system in stationary sources

1 Scope

This standard specifies the composition structure, technical requirements,

performance indicators, detection methods of nonmethane hydrocarbons

continuous emission monitoring system in stationary sources.

This standard applies to the design, production, testing of nonmethane

hydrocarbons continuous emission monitoring system in stationary sources.

2 Normative references

This standard refers to the following documents or their clauses. For undated

reference documents, their valid versions are applicable to this standard.

GB 3836.1 Explosive atmospheres - Part 1.Equipment - General

requirements

GB/T 4208 Degrees of protection provided by enclosure (IP code)

GB/T 16157 The determination of particulates and sampling methods of

gaseous pollutants emitted from exhaust gas of stationary source

HJ 38 Stationary source emission - Determination of total hydrocarbons,

methane and nonmethane hydrocarbons - Gas chromatography

HJ 75 Specifications for continuous emissions monitoring of SO2, NOX, and

particulate matter in the flue gas emitted from stationary sources

HJ 76 Specifications and test procedures for continuous emission monitoring

system for SO2, NOX and particulate matter in flue gas emitted from

stationary

3 Terms and definitions

The following terms and definitions apply to this standard.

3.1

Nonmethane hydrocarbons (NMHC)

Under the conditions specified in the HJ 38 standard, the sum of other

gaseous organic compounds than methane, which respond on the hydrogen

flame ionization detector, as calculated in carbon, unless otherwise specified.

3.2

Nonmethane hydrocarbons continuous emission monitoring system

(NMHC-CEMS)

All equipment, which is required to continuously monitor the emission

concentration and volume of non-methane hydrocarbons in exhaust gas

from stationary sources, as referred to as NMHC-CEMS.

3.3

Analysis cycle time

The time interval, between the two sets of measurement results, which is

given when the system is running continuously.

4 System composition and structure

4.1 System composition

The stationary pollution source NMHC-CEMS consists of a non-methane

hydrocarbon monitoring unit, a waste gas parameter monitoring unit, a data

collection and processing unit, as shown in Figure 1.The system measures the

concentration of non-methane hydrocarbons in the exhaust gas AND the

exhaust gas parameters (temperature, pressure, flow rate or flow, humidity,

etc.).

4.2 System structure

The NMHC-CEMS's system structure mainly includes sample collection and

transmission devices, pretreatment equipment, analysis instruments, data

collection and transmission equipment, other auxiliary equipment, etc.

Depending on the measurement method and principle of the system, the

system may consist of all or part of the above structure.

4.2.3 Analytical instruments

The analytical instrument is used to measure and analyze the collected waste

gas samples from the pollution source. For specific technical requirements, see

5.4.3.

4.2.4 Data collection and transmission equipment

Data collection and transmission equipment is used to collect, process, store

monitoring data. It can transmit monitoring data and equipment working status

information, according to the commands from central computer. See 5.4.4 for

specific technical requirements.

5 Technical requirements

5.1 Appearance requirements

5.2 Working conditions

NMHC-CEMS shall work normally under the following conditions.

5.3.2 Insulation strength

When the ambient temperature is (15 ~ 35) °C and the relative humidity is <=

85%, the system shall last for 1 min, under 1500 V (effective value), 50 Hz sine

wave test voltage; there shall be no breakdown or arcing.

5.3.3 The system shall have a leakage protection device and good grounding

measures, to prevent damage to the system, which is caused by lightning

strikes.

5.4 Functional requirements

5.4.1 Requirements for sample collection and transmission devices
5.4.1.1 The sample collection and transmission device shall select the materials,

which are resistant to high temperature, corrosion, non-absorptive, non-reactive

with the pollutants under test; meanwhile, it shall not affect the normal

measurement of the pollutants under test.

5.4.1.6 The sampling pump shall have sufficient suction capacity, to overcome

the negative pressure of the flue; ensure that the sampling flow is accurate,

reliable, and relatively stable.

5.4.2 Pretreatment equipment
5.4.2.1 The pretreatment equipment and its components shall be easy to clean

and replace. It shall be made of materials, which do not absorb and do not react

with the pollutants under test.

5.4.2.2 In order to prevent particles from polluting the analyzer, a fine filter can

be set, before the gas sample enters the analyzer. The fine filter shall be able

to filter particles, which have a particle size of (0.5 ~ 2) µm at least.

5.4.3 Requirements of analytical instrument
5.4.4.4 It has a digital signal output function.
5.4.4.5 It has Chinese data collection, recording, processing and control

software. For data collection, recording, processing requirements, see

Appendix B in HJ 76.

5.4.6 Requirements of calibration function
5.4.6.1 The system shall be able to be calibrated manually and/or automatically.
5.4.6.2 For the system, which adopts the extraction measurement method, it

shall have a fixed and easy-to-operate calibration gas full-system calibration

function.

6 Performance indicators

6.1 Laboratory testing

6.1.1 Analysis cycle

System analysis cycle. <= 2 min.

6.1.5 24h drift

24 h zero drift and span drift. Not more than ± 3% of full scale.

6.1.6 The influence of environmental temperature changes

When the ambient temperature changes within the span of (15 ~ 35) °C, the

change of non-methane hydrocarbon's indication value. Not more than ±5% of

full scale.

6.1.10 Response factor

When the system measures non-methane hydrocarbons, the relative mass

response factors of other VOCs, as relative to propane, must meet a certain

span of requirements, as shown in Table 1.

6.2 On-site detection of pollutant discharge

6.2.1 Analysis cycle

System analysis cycle. <= 3 min.

6.2.2 24h drift

24 h zero drift and span drift. Not more than ± 3% of full scale.

6.2.3 Accuracy

When using the reference method to make measurement, the average

concentration of the non-methane hydrocarbon is.

6.2.4 Performance indicators of exhaust gas parameters

The performance indicator requirements of exhaust gas parameters (oxygen,

flow rate, flue gas temperature, humidity) shall meet the relevant requirements

of HJ 76.

7 Testing method

7.1 Laboratory testing requirements and methods

7.1.1 General requirements
7.1.1.1 Take at least 3 sets of systems of the same model for simultaneous

testing, at the designated laboratory site.

7.1.1.5 If the test is interrupted due to a system failure, restart the test after the

system returns to normal; the completed test indicators and data will be invalid.

During the test period, the number of system failures per unit (set) is <= 2.

7.1.1.6 For systems with automatic calibration function, it may set any cycle;

the system will perform automatic calibration; during the testing, the automatic

calibration cycle shall be set to >= 24 h.

7.1.2 Standard material requirements
7.1.2.1 Calibration gas. Commercially available certified calibration gas, which

has an uncertainty of <= 2.0%.

7.1.3 Laboratory testing methods
7.1.3.1 Analysis cycle

The analysis cycle time refers to the time interval, between the two sets of

measurement results, which are given, during continuous operation of NMHC-

CEMS. There are 3 measurements in 3 consecutive days. The daily analysis

cycle time shall meet the requirements of Table 3.

7.1.3.3 Repeatability

After the system under test is running stably, lead in the span calibration gas.

Record the measured value, after the displayed value is stable. Repeat the

above operation, at least 6 times, by the use of the span calibration gas of the

same concentration. Calculate the repeatability (relative standard deviation) of

the system under test, according to formula (2). The results shall meet the

requirements of Table 3.

7.2.2 On-site testing methods
7.2.2.1 Analysis cycle

The on-site testing method of the analysis cycle time is consistent with the

laboratory testing method, as shown in 7.1.3.1.The results shall meet the

requirements of Table 4.

7.2.2.2 24h drift

The detection methods of on-site 24h zero drift and span drift are consistent

with the laboratory detection methods, as shown in 7.1.3.5.The results shall

meet the requirements of Table 4.

7.2.2.4 Exhaust gas parameters

Refer to HJ 76, for the detection method of exhaust gas parameters (oxygen,

flow rate, smoke temperature, humidity).

8 Quality assurance

8.1 Installation quality assurance

8.1.1 The installation location and on-site supporting environmental conditions

shall meet the requirements of the HJ 75 standard.

8.2 Testing quality assurance

8.2.1 The on-site testing shall be carried out, under the condition that the

stationary pollution source discharges pollutants normally. During the initial

inspection and re-inspection, a dedicated person must be responsible for

supervising the working conditions AND maintain it relatively stable during the

test.

8.2.4 For complete extraction and dilution extraction systems, when performing

zero and span calibration, in principle, the zero gas and calibration gas must

pass through the same path as the sample gas (such as sample collection

device, transmission pipeline, pretreatment, etc.).

8.3 Daily operation quality assurance

8.3.1 Regular calibration

9 Testing items

For the stationary pollution source waste gas NMHC-CEMS, the laboratory

testing and on-site testing items are as shown in Table 3 and Table 4.The on-

site testing items of exhaust gas parameters are as shown in Table 5.See

Appendix B, for the relevant record form.

......

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 30 pages — is available in the English PDF.

Referenced standards

How to Download HJ 1013-2018

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
30 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

HJ 1013-2018

$445.00