GBZ/T 307.2-2018Determination of cadmium in urine -- Part 2: Inductively coupled plasma mass spectrometry method (English PDF)

尿中镉的测定 第2部分:电感耦合等离子体质谱法

Preview PDF

This is a limited preview

Buy now to download the full PDF (2 pages)

Need quotation for your management?

Get a formal quotation for GBZ/T 307.2-2018 in 30 minutes.

Issued by

National Health and Family Planning Commission

Level / Type

National · Recommended

Issue date

August 16, 2018

Implementation date

January 1, 2019

Scope

GBZ/T 307.2-2018 is the English-translated version of 尿中镉的测定 第2部分:电感耦合等离子体质谱法.

GBZ/T 307.2-2018 — This part of GBZ/T 307 specifies inductively coupled plasma mass spectrometry for the determination of cadmium in urine. This section applies to the determination of cadmium in urine of occupational contacts.

Found what you need?

Document preview — GBZ/T 307.2-2018

National Standard of the People's Republic of China

ICS
13.100
Classification
C 60

Issued by: National Health and Family Planning Commission

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 Principle
  • 4 Instruments
  • 5 Reagent
  • 6 Sample collection, transportation and storage
  • 7 Analysis steps
  • 7.1 Preparation of urinary cadmium standard series
  • 7.2 Pretreatment of urinary cadmium samples and sample blanks
  • 7.3 Determination of urinary cadmium standard series, samples and sample blanks
  • 7.4 Quality control of the inspection process
  • 8 Calculation
  • 9 Description

Foreword

This standard is formulated in accordance with the Law of the People's Republic GBZ/T 307 "Determination of cadmium in urine" is divided into two parts.

-- Part 1. Graphite furnace atomic absorption spectrometry;

-- Part 2. Inductively coupled plasma mass spectrometry.

This part is the second part of GBZ/T 307.

This part is drafted in accordance with the rules given in GB/T 1.1-2009.

This section is mainly drafted by. China Center for Disease Control and Prevention, Occupational Health and Poison Control Institute, Guangdong Provincial Occupational Disease Prevention and Treatment Institute, Jiangsu Provincial Center for Disease Control and Prevention, Shandong Institute of Occupational Health and Occupational Diseases.

1 Scope

This part of GBZ/T 307 specifies inductively coupled plasma mass spectrometry for the determination of cadmium in urine.

This section applies to the determination of cadmium in urine of occupational contacts.

2 Normative references

The following documents are indispensable for the application of this document. For dated references, only the dated version applies to this document.

For undated references, the latest edition (including all amendments) applies to this document.

General rules for biological monitoring of GBZ/T 295 occupational population

3 Principle

Urine sample (hereinafter referred to as urine sample) is directly diluted with 0.5% nitric acid solution, with (Rh, 10.0 µg/L) as an internal standard, with inductance the concentration of cadmium in the diluent was determined by coupled plasma mass spectrometry.

4 Instruments

4.1 Covered polyethylene plastic bottle. 100 mL.

4.2 Volumetric flask. 10 mL.

4.3 Polyethylene centrifuge tube. 15 mL.

4.4 Micropipette. The range is from 100 µL to 1000 µL.

4.5 Inductively Coupled Plasma Mass Spectrometer. equipped with an autosampler.

Instrument operation reference conditions.

a) cooling gas flow rate. 12.5 L/min;

b) auxiliary gas flow rate. 0.70 L/min;

c) atomizing gas flow rate. 0.88 L/min;

d) peristaltic pump speed. 30 r/min;

e) Nebulizer temperature. 3 °C.

Note. The above conditions are provided for the convenience of standard users. If the conditions are changed, the same effect should be achieved.

5 Reagent

5.1 Deionized water.

5.2 Nitric acid, excellent grade pure, rho20=1.42 g/mL.

5.3 Nitric acid solution, 0.5% (volume fraction).

5.4 Blank urine sample. Normal human urine that does not touch cadmium, add nitric acid in a ratio of 100 1 and mix.

5.5 Standard solution. Adopt nationally recognized standard solution of cadmium and strontium.

6 Sample collection, transportation and storage

6.1 Collect 100 mL of urine sample in polyethylene plastic bottle, add nitric acid in a ratio of 100.1, mix well, and measure urinary muscle as soon as possible.

anhydride.

6.2 Sample blank. randomly take 2 parts of the urine collection container with the same batch number as the sample collection, and add water equivalent to the sample collection amount as the sample.

Blank, transport and save at the same time as the sample.

6.3 Sample transportation. Store the collected samples and sample blanks in a clean container for refrigerated transport.

6.4 Sample storage. Store at 4 °C for 2 months.

7 Analysis steps

7.1 Preparation of urinary cadmium standard series

The cadmium monolith standard solution was diluted to 1.0 µg/mL cadmium standard application solution with 0.5% nitric acid solution, and then the concentration was 0.0 µg/L.

0.2 µg/L, 1.0 µg/L, 5.0 µg/L, 10.0 µg/L, 20.0 µg/L cadmium standard solution series (medium is 0.5% nitric acid solution).

7.2 Pretreatment of urinary cadmium samples and sample blanks

Remove the urinary cadmium sample or sample blank and return to the laboratory temperature. Shake well and mix well, take 0.50 mL, and place in 15 mL polyethylene in the centrifuge tube, add 4.50 mL of 0.5% nitric acid solution and shake it to be tested. The rest of the treatment steps are the same as above. At the same time as a reagent blank, the same sample deal with.

7.3 Determination of urinary cadmium standard series, samples and sample blanks

7.3.1 Determination of urinary cadmium standard series. The instrument was adjusted to the best measurement state, and each standard solution was determined, and each concentration was repeatedly measured twice.

The working curve is plotted against the corresponding cadmium concentration (µg/L) or the regression equation is calculated.

7.3.2 Determination of sample and sample blanks. Determine the sample and sample blanks using the operating conditions of the measurement work series, and subtract the measured count value.

After the reagent blank count value, the cadmium concentration (µg/L) was calculated from the standard curve or the regression equation.

7.4 Quality control of the inspection process

7.4.1 The blank measurement result should be less than the detection limit. When the test result is greater than the detection limit, it indicates that the sample is in the process of collection, transportation and storage.

If contaminated, batch samples should be discarded.

7.4.2 Before the urine sample is tested, two levels of urinary cadmium quality control samples shall be measured simultaneously. The quality control sample is processed and sampled.

Processing should be carried out in the same batch. After measuring the standard series of urinary cadmium and drawing a standard curve, first measure the quality control sample, such as the quality control sample.

The determination results fall within a given range and sample determination can be performed. The quality control sample should be repeatedly determined after every 10 samples were measured. When the measurement result when the given range is exceeded, the first 10 samples should be re-measured.

8 Calculation

Calculate the concentration of cadmium in the urine sample according to formula (1).

In the formula.

C - the concentration of cadmium in urine, in micrograms per gram of creatinine (µg/g creatinine);

C0 - the concentration of cadmium in the urine sample obtained from the working curve or the regression equation in micrograms per liter (µg/L);

F - sample dilution factor.

Cr - Urinary creatinine concentration in grams per liter (g/L).

9 Description

9.1 The detection limit of this method is 0.08 µg/L, the lower limit of quantification is 0.30 µg/L; the measurement range is 0.30 µg/L to.200.0 µg/L. relatively the standard deviation ranges from 1.0% to 1.6% (n=6).

9.2 Volumetric flasks and other supplies need to be soaked overnight in 1..1 nitric acid solution, washed sequentially with tap water, laboratory water, and added the background test is carried out by using 1% nitric acid solution as the method to be tested, and the test result should be less than the detection limit. Covered polyethylene plastic bottle with stoppered polyethylene Disposable products such as centrifuge tubes and micro-pump tips should also be used after sampling.

9.3 The quality control of the testing process shall be carried out in accordance with the requirements of GBZ/T 295.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 2 pages — is available in the English PDF.

How to Download GBZ/T 307.2-2018

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
2 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GBZ/T 307.2-2018

$85.00