Valid

GB/T 38485-2021Determination for trace gene residues of microorganisms—Microdroplet digital PCR (English PDF)

微生物痕量基因残留测定 微滴数字PCR法

Preview PDF

This is a limited preview

Buy now to download the full PDF (17 pages)

Need quotation for your management?

Get a formal quotation for GB/T 38485-2021 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

December 31, 2021

Implementation date

July 1, 2022

Scope

GB/T 38485-2021 is the English-translated version of 微生物痕量基因残留测定 微滴数字PCR法.

GB/T 38485-2021 is the Chinese national method for determining the trace gene residues of microorganisms by microdroplet digital PCR. It gives the principle, the reagents, the apparatus, the extraction of the nucleic acid, the preparation of the droplets, the amplification and reading, the calculation of the result and the test report.

It applies to the determination of characteristic gene residues below 100 pg/mL, the level at which an ordinary PCR no longer answers. Biotechnology plants, food and pharmaceutical laboratories use it. Issued on 31 December 2021, it has applied since 1 July 2022.

Found what you need?

Document preview — GB/T 38485-2021

National Standard of the People's Republic of China

ICS
07.080
Classification
A 21

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative References4
  • 3 Terms and Definitions4
  • 4 Principle4
  • 5 Reagents or Materials5
  • 6 Instruments and Equipment6
  • 7 Preparation of Reaction System6
  • 8 Microdroplet Digital PCR Reaction8
  • 9 Result Analysis and Expression9

1 Scope

This document specifies the microdroplet digital PCR method for the determination of trace gene residues (below 100 pg/mL) of microorganisms.

This document is applicable to the determination of trace characteristic gene residues (below 100 pg/mL) of Saccharomyces cerevisiae and Lactobacillus plantarum in processed products.

2 Normative References

The contents of the following documents constitute indispensable clauses of this document through the normative references in this text. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.

GB/T 6682 Water for Analytical Laboratory Use - Specification and Test Methods

3 Terms and Definitions

The following terms and definitions are applicable to this document.

3.1 Trace Gene Residue

Trace gene residue refers to trace (below 100 pg/mL) nucleic acid that is difficult to remove in the subsequent separation and purification of microbial processed products.

4 Principle

Adopt the droplet microfluidic technology to divide the target nucleic acid PCR amplification system remaining in the biological sample to be tested into dozens to tens of thousands of microliter water-in-oil microdroplets containing one copy of the nucleic acid molecule. In accordance with the conventional PCR system, amplify it; read out the signal through the florescent probe. The number of the generated microdroplet positive signals represents the copy number of target nucleic acid molecules in the original sample, through which, the number of target nucleic acid molecules can be directly counted, and the trace gene residues in the initial sample can be quantified.

In a constant-temperature shaker, at 30 °C ± 1 °C and 180 r/min, culture it overnight to obtain the seed solution. Take 0.5 mL of the seed solution and transfer it to 10 mL of MRS liquid medium; in a constant-temperature shaker, at 30 °C ± 1 °C and 180 r/min, culture it, until OD600 is 0.3 ~ 0.4 (about 2 h ~ 3 h) to obtain a pure culture of lactobacillus plantarum CGMCC 1.2437.

5 Reagents or Materials

5.8 Lactobacillus plantarum CGMCC 1.2437 pure culture. pick a single colony of lactobacillus plantarum CGMCC 1.2437 in the activated plate; inoculate it into 25 mL of MRS liquid medium.

6 Instruments and Equipment

6.3 Thermal cycler.

6.4 Electronic balance. with an accuracy of 0.01 g.

6.5 Pipette. with a measuring range of 0.1 µL ~ 2.5 µL, 2 µL ~ 20 µL, 10 µL ~ 100 µL and 100 µL ~ 1,000 µL.

6.6 Constant-temperature shaker. the temperature can achieve (28 ± 1) °C and (37 ± 1) °C; the speed is at 125 r/min ~ 250 r/min.

6.7 pH meter. with an accuracy of 0.1.

6.8 UV-visible spectrophotometer. the detectable wavelength includes (600 ± 20) nm; equipped with a 1 cm cuvette.

7 Preparation of Reaction System

7.1 Preparation of Saccharomyces Cerevisiae Residue Gene Detection Reaction System

7.1.1 DNA template preparation for test sample

Take 1.00 g of the test sample. In accordance with the instructions of the nucleic acid extraction kit, carry out DNA extraction. Finally, use 50 µL of sterile water for DNA elution. The extracted DNA sample is tested for nucleic acid concentration and purity with a nucleic acid quantometer.

7.1.2 DNA template preparation for positive control

Take 100 µL of the lactobacillus plantarum pure culture (5.8); use sterile water to dilute it, until OD600 is 0.2 (at this moment, the bacterial concentration is about 2 x 10^8 CFU/mL). Adopt the method of gradient dilution; continue to use sterile water to dilute it, until the bacterial concentration is 1 x 10^3 CFU/mL. Take 100 µL of the diluted lactobacillus plantarum pure culture; add 1.00 g of the test sample as a positive control. Use a DNA extraction kit to perform number of effective microdroplets shall not be lower than the number of microdroplets specified by the microdroplet digital PCR equipment, otherwise, the test sample need to be re-tested.

In accordance with the operating instructions of the microdroplet digital PCR instrument, within 30 min, perform PCR amplification, or place it in a refrigerator at 4 °C for PCR amplification within 4 h. The amplification can be performed on the digital PCR or a thermal cycler, and the heating and cooling rate is <= 2.5 °C/s; see the amplification procedure in Table 3.After the amplification is completed, read the fluorescence signals of all microdroplets for the subsequent data analysis.

9 Result Analysis and Expression

9.1 Threshold Setting

In accordance with the end-point fluorescence value of the negative amplification system in the system, set the threshold value of florescence. The threshold value needs to clearly distinguish between negative and positive microdroplets in the amplification system.

9.2 Detection Result Analysis of Control Group

There was no amplification phenomenon in the negative control group and the blank control, and obvious amplification phenomenon appeared in the positive control group, which indicates that this detection with the microdroplet digital PCR method is normal and reliable, and there is no need to repeat the detection.

9.3 Specimen Detection Result Analysis

The result of specimen detection is analyzed in accordance with the following principles.

a) Both the positive control and the test sample manifested obvious amplification

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download GB/T 38485-2021

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
17 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 38485-2021

$135.00