GB/T 24198-2009Ferronickel - Determination of nickel, silicon, phosphorus, manganese, cobalt, chromium, and copper contents - Wavelength dispersive X-ray fluorescence spectrometry (Routine method) (English PDF)
镍铁 镍、硅、磷、锰、钴、铬和铜含量的测定 波长色散X-射线荧光光谱法(常规法)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 24198-2009 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
July 8, 2009
Implementation date
April 1, 2010
Scope
GB/T 24198-2009 is the English-translated version of 镍铁 镍、硅、磷、锰、钴、铬和铜含量的测定 波长色散X-射线荧光光谱法(常规法).
GB/T 24198-2009 specifies the determination of the nickel, silicon, phosphorus, manganese, cobalt, chromium and copper contents of ferronickel by wavelength dispersive X-ray fluorescence spectrometry, as a routine method.
Clause 1 makes it applicable to the determination of electric furnace, induction furnace, converter and other as-cast conditions or nickel rolling, and puts the content range of each element in Table 1; a warning adds that personnel using the standard should have formal laboratory work experience, that the standard does not indicate all possible safety issues, and that appropriate safety and health measures and compliance with the relevant national regulations are the user's responsibility.
The standard runs Scope, Normative references, Principle, Reagents and materials, Instrument and equipment, Sampling and sample preparation, Instrument preparation, Analysis steps, Calculation of results, Precision and Test report, with two normative annexes, one on the instrument precision test and one giving the acceptance flow chart for sample analysis values.
Clause 2 cites GB/T 4010, GB/T 6379.1 and GB/T 6379.2, while clause 6 refers sampling and sample preparation to GB/T 20066 and GB/T 4010 and clause 7 requires the instrument working environment to comply with GB/T 16597.
The principle in clause 3 is that the primary X-ray emitted from the X-ray tube is irradiated onto the flat, smooth surface of the sample, that the resulting characteristic X-ray is dispersed by the crystal, and that the detector measures the fluorescence intensity at the 2 theta angle corresponding to the selected characteristic wavelength; the sentence that would state how the mass fractions follow from this is cut off on the page.
The practical clauses call for P10 gas, a mixture of 90 percent argon and 10 percent methane, for the flow proportional counter, together with certified reference material or reference material, for sample cutting and surface polishing equipment and a spectrometer whose analytical accuracy and stability meet Annex A, with a scintillation counter for heavy elements and a flow proportional counter for light ones, for a vacuum generally below 30 Pa and below 20 Pa for light elements, and for a sample at least 5 mm thick whose analytical surface covers the whole sample box mask.
Clause 8 then sets the measurement conditions, the plotting of the calibration curve on a series of certified reference or reference materials of similar smelting process and similar chemical composition, each measured at least twice, the confirmation of the curve against the certified or standard value by equation 3, and the standardization of the instrument before unknown samples are analysed; results are calculated from the calibration curve by equation 8 or equation 9, and the precision in clause 10 comes from a common test in 12 laboratories evaluated according to GB/T 6379.1 and GB/T 6379.2 and set out in Table 3.
The foreword names the China Iron and Steel Industry Association as the proposer and puts the standard under the jurisdiction of the national technical committee on pig iron and ferroalloy; the contents of the test report clause are cut off on the page. It was issued on 2009-07-08 and has applied since 2010-04-01.
Found what you need?
Document preview — GB/T 24198-2009
National Standard of the People's Republic of China
- ICS
- 77.100
- Classification
- H 11
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Principle5
- 4 Reagents and materials5
- 5 Instrument and equipment6
- 6 Sampling and sample preparation7
- 7 Instrument preparation7
- 8 Analysis steps7
- 9 Calculation of results11
- 10 Precision11
- 11 Test report12
- Annex A (Normative) Instrument precision test14
- Annex B (Normative) Sample analysis value acceptance flow chart15
Foreword
Annex A and Annex B of this Standard are normative.
This Standard was proposed by China Iron and Steel Industry Association.
This Standard shall be under the jurisdiction of National Technical Committee on Pig Iron and Ferroalloy of Standardization Administration of China.
1 Scope
This Standard specifies the determination of nickel silicon phosphorus manganese cobalt chromium and copper contents by wavelength dispersive X-ray fluorescence spectrometry.
It is applicable to the determination of electric furnace, induction furnace, converter and other as-cast conditions or nickel rolling. The range of each element is shown in Table 1.
2 Normative references
The provisions in following documents become the provisions of this Standard through reference in this Standard. For dated references, the subsequent amendments (excluding corrigendum) or revisions do not apply to this Standard, however, parties who reach an agreement based on this Standard are encouraged to study if the latest versions of these documents are applicable. For undated references, the latest edition of the referenced document applies.
GB/T 4010, Ferroallys sampling and preparation of samples for chemical analysis
GB/T 6379.1, Accuracy (trueness and precision) of measurement methods and results - Part 1: General principles and definitions (GB/T 6379.1-2004, ISO 5725-1:1994, IDT)
GB/T 6379.2, Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method (GB/T 6379.2-2004, ISO 5725-2:1994, IDT)
3 Principle
When the primary X-ray emitted from the X-ray tube is irradiated onto the surface of the flat, smooth sample, the resulting characteristic X-ray shall be crystallized by the crystal, and the detector shall measure the X-ray fluorescence intensity at the corresponding 2 theta angle of the selected characteristic wavelength. The mass fraction of nickel, silicon, phosphorus,
4 Reagents and materials
4.1 P10 gas (90% argon and 10% methane mixed gas) for flow proportional counter.
4.2 Certified reference material or reference material
Certified reference material (CRM) or reference material (RM).
5 Instrument and equipment
5.1 Sample preparation equipment
Sample cutting equipment and sample surface polishing equipment.
5.2 X-ray fluorescence spectrometer
The analytical accuracy and stability of the simultaneous or sequential wavelength dispersive spectrometer shall meet the requirements of Annex A.
5.2.4 Detector
Scintillation counter for heavy element analysis; flow proportional counter for light element analysis. It can also use a sealed proportional counter.
5.2.5 Vacuum system
During the measurement process, the vacuum system pressure is generally below 30 Pa and remains constant; for light elements, it shall be less than 20 Pa.
6 Sampling and sample preparation
Refer to the provisions of GB/T 20066 and GB/T 4010.
The size of the sample shall depend on the geometry of the sample box. The analytical surface shall be able to cover all the sample box mask, while ensuring the sample thickness of at least 5 mm.
7 Instrument preparation
7.1 Instrument working environment
The instrument working environment shall comply with GB/T 16597.
8 Analysis steps
8.1 Measurement conditions
Select the appropriate measurement conditions, according to the type of equipment used, the type of sample, elements for analysis, coexisting elements and their content range.
8.2 Plotting and confirmation of calibration curve
8.2.1 Calibration curve plotting
Under the selected working conditions, use X-ray fluorescence spectrometer to measure a series of certified reference material (CRM) or reference material (RM) of similar sample smelting process and similar chemical composition. Each sample shall be measured at least 2 times.
8.2.2 Confirmation of calibration curve accuracy
According to the selected analysis conditions, use X-ray fluorescence spectrometer to measure certified reference material (CRM) or reference material (RM) of similar sample smelting process and of similar chemical composition. There shall be no statistically significant difference between the analyzed value of the element for analysis and the certified value or standard value Ac. Determine whether there are significant differences according to equation (3).
8.3 Analysis of unknown sample
8.3.1 Standardization of instrument
Regularly confirm the analysis of standardized sample. When the instrument drifts, the instrument shall be drilled and calibrated by measuring the X-ray fluorescence intensity of the normalized sample.
9 Calculation of results
Calculate the contents of elements for analysis from the calibration curve, according to the fluorescence intensity measurement of unknown sample, by equation (8) or equation (9).
10 Precision
The precision of this Standard is the result of a common test conducted in 12 laboratories and determined according to GB/T 6379.1 and GB/T 6379.2. See Table 3 for precision.
11 Test report
The test report shall contain the following information. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 4010Ferroalloys sampling and preparation of samples for chemical analysis
- GB/T 6379.1Accuracy (trueness and precision) of measurement methods and results -- Part 1: General principles and definitions
- GB/T 6379.2Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
Similar standards
How to Download GB/T 24198-2009
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.