GB/T 20485.32-2021Methods for the calibration of vibration and shock transducers - Part 32: Resonance testing - Testing the frequency and the phase response of accelerometers by means of shock excitation (English PDF)
振动与冲击传感器校准方法 第32部分:谐振测试 用冲击激励测试加速度计的频率和相位响应
This is a limited preview
Buy now to download the full PDF (24 pages)
Need quotation for your management?
Get a formal quotation for GB/T 20485.32-2021 in 30 minutes.
Issued by
State Administration for Market Regulation, China National Standardization Administration
Level / Type
National · Recommended
Issue date
August 20, 2021
Implementation date
March 1, 2022
Scope
GB/T 20485.32-2021 is the English-translated version of 振动与冲击传感器校准方法 第32部分:谐振测试 用冲击激励测试加速度计的频率和相位响应.
GB/T 20485.32-2021 is the Chinese standard "Methods for the calibration of vibration and shock transducers - Part 32: Resonance testing - Testing the frequency and the phase response of accelerometers by means of shock excitation".
Its scope clause reads: This part of GB/T 20485 specifies the instruments and procedures for testing the frequency and phase response of accelerometers with shock excitation.
This section applies to piezoelectric, piezoresistive and variable capacitance accelerometers with a damping ratio of less than 1, and the frequency range is up to 150kHz.
The accelerometer under test that requires this method to obtain frequency and phase response needs to have a good mechanical installation as stated in ISO 5348 Characteristics, and the mass of the reference impact ball exceeds at least three times the mass of the measured accelerometer.
The phase response of the measured accelerometer obtained by this method assumes that the accelerometer has input and output signals at a frequency of 0 Hz. There is zero phase shift between.
Note 1.The user may notice that the frequency and phase response of the same accelerometer used in the field may be different, depending on the quality of the test structure, Consistency and installation method.
This method can only qualitatively evaluate the frequency and phase response of the accelerometer. Note 2.Compared to focusing on sufficient dynamic range characteristics, users may not pay much attention to the initial part of the measured accelerometer frequency and phase response.
Resolution. The biggest advantage of obtaining accelerometer frequency and phase response through this method is to obtain the best number of spectral lines in the initial part of the above-mentioned characteristics. Its clauses include factors affecting measurement repeatability.
Found what you need?
Document preview — GB/T 20485.32-2021
National Standard of the People's Republic of China
- Classification
- J 04
Issued by: State Administration for Market Regulation, China National Standardization Administration
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Factors affecting measurement repeatability
- 4 Equipment and other devices
- 4.1 Environmental conditions
- 4.2 Reference Impact Ball
- 4.2.1 General
- 4.2.2 Reference impact ball size
- 4.2.3 Refer to the selection of the diameter range of the impact ball
- 4.2.4 Refer to the requirements of the impact ball mounting surface and thread tolerance
- 4.3 Hit the ball
- 4.4 Signal Analyzer
- 4.5 Conditioning amplifier
- 4.6 Auxiliary devices and methods
- 5 Step
- 5.1 Assemble the test structure
- 5.1.1 Installing the accelerometer under test
- 5.1.2 Installation reference impact ball
- 5.1.3 Install the guide tube of the impact ball
- 5.2 Connecting the instrument
- 5.3 Setting up the signal analyzer
- 5.3.1 Analyzer initialization
- 5.3.2 Setting measurement parameters
- 5.3.3 Setting display
- 5.3.4 Set Mark
- 5.4 Test
- 6 Results processing
- 6.1 Recording the time signal
- 6.2 Time signal processing
- 7 Results report
1 Scope
This part of GB/T 20485 specifies the instruments and procedures for testing the frequency and phase response of accelerometers with shock excitation.
This section applies to piezoelectric, piezoresistive and variable capacitance accelerometers with a damping ratio of less than 1, and the frequency range is up to 150kHz.
The accelerometer under test that requires this method to obtain frequency and phase response needs to have a good mechanical installation as stated in ISO 5348
Characteristics, and the mass of the reference impact ball exceeds at least three times the mass of the measured accelerometer.
The phase response of the measured accelerometer obtained by this method assumes that the accelerometer has input and output signals at a frequency of 0 Hz.
There is zero phase shift between.
Note 1.The user may notice that the frequency and phase response of the same accelerometer used in the field may be different, depending on the quality of the test structure,
Consistency and installation method. This method can only qualitatively evaluate the frequency and phase response of the accelerometer.
Note 2.Compared to focusing on sufficient dynamic range characteristics, users may not pay much attention to the initial part of the measured accelerometer frequency and phase response.
Resolution. The biggest advantage of obtaining accelerometer frequency and phase response through this method is to obtain the best number of spectral lines in the initial part of the above-mentioned characteristics.
2 Normative references
The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this article
Pieces. For undated reference documents, the latest version (including all amendments) is applicable to this document.
ISO 2041 Mechanical vibration, shock and condition monitoring vocabulary (Mechanicalvibration, shockandconditionmonito- ring-Vocabulary)
ISO 5347-22 Vibration and shock sensor calibration method Part 22.Accelerometer resonance test general method (Methods forthecalibrationofvibrationandshockpick-ups-Part 22.Accelerometerresonancetesting-General methods)
ISO 5348 Mechanical vibration and shock-mechanical installation of accelerometers (Mechanical vibration and shock-Mechanical mountingofaccelerometers).
3 Factors affecting measurement repeatability
The uncertainty limits of frequency response measurement are as follows.
For the resonant frequency of the measured accelerometer, the absolute uncertainty is equal to the resolution of the frequency analysis, which is the long time record of the accelerometer signal
The reciprocal of degrees. This method suggests that the minimum number of spectral lines in the frequency domain is 400.Assuming that the resonant frequency is in the middle of the frequency band, the standard of the resonant frequency is not accurate
The fixed degree is about 0.5%.
Note 1.This uncertainty is assumed to be uniformly distributed within the frequency resolution range.
For the damping ratio of the measured accelerometer, the uncertainty depends on the signal-to-noise ratio measured in the time domain.
Suppose that when the measurement is performed, the maximum value of the signal is close to the upper limit of the dynamic range of the measuring instrument and the typical damping ratio of the piezoelectric accelerometer is about
0.01, the standard uncertainty of the damping ratio measurement is about 1%.
Note 2.The signal analyzer used for damping ratio measurement should have a dynamic range of at least 80dB. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.
Referenced standards
Normative references
ISO 2041 · ISO 5347 · ISO 5348
How to Download GB/T 20485.32-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 20485.33-2018 — Methods for the calibration of vibration and shock transducers -- Part 33: Testing of magnetic field sensitivity
GB/T 20485.31-2011 — Methods for the calibration of vibration and shock transducers -- Part 31: Testing of transverse vibration sensitivity
GB/T 20485.41-2015 — Methods for the calibration of vibration and shock transducers -- Part 41: Calibration of laser vibrometers
Secure payment via Stripe
Payments accepted
GB/T 20485.32-2021
$210.00