GB/T 18281.2-2024Sterilization of health care products - Biological indicators - Part 2: Biological indicators for ethylene oxide sterilization processes (English PDF)
医疗保健产品灭菌 生物指示物 第2部分:环氧乙烷灭菌用生物指示物
This is a limited preview
Buy now to download the full PDF (9 pages)
Need quotation for your management?
Get a formal quotation for GB/T 18281.2-2024 in 30 minutes.
Issued by
State Administration for Market Regulation
Level / Type
National · Recommended
Issue date
November 28, 2024
Implementation date
December 1, 2025
Scope
GB/T 18281.2-2024 is the English-translated version of 医疗保健产品灭菌 生物指示物 第2部分:环氧乙烷灭菌用生物指示物.
GB/T 18281.2-2024 is the Chinese standard "Sterilization of health care products - Biological indicators - Part 2: Biological indicators for ethylene oxide sterilization processes".
Its scope clause reads: This Document specifies the requirements and test methods for test microorganisms, bacterial suspensions, contaminated carriers, and biological indicators to be used in evaluating the performance of sterilizers and sterilization processes.
The sterilizer and sterilization process use pure ethylene oxide gas or a mixture of ethylene oxide gas and dilution gas, with a sterilization temperature range of 29 °C to 65 °C. NOTE 1.Requirements for validation and control of ethylene oxide sterilization are found in ISO 11135 and ISO 14937.
NOTE 2.Some national or regional regulations may specify workplace safety requirements. Its clauses include terms and definitions; general requirements; test microorganisms; bacterial suspensions; carriers and primary packaging; contaminated carriers and biological indicators; bacterial count and resistance.
It was issued by the State Administration for Market Regulation on 2024-11-28 and took effect on 2025-12-01.
Found what you need?
Document preview — GB/T 18281.2-2024
National Standard of the People's Republic of China
- ICS
- 11.080.01
Issued by: State Administration for Market Regulation
Contents
- Foreword...3
- Introduction...5
- 1 Scope...7
- 2 Normative References...7
- 3 Terms and Definitions...7
- 4 General Requirements...8
- 5 Test Microorganisms...8
- 6 Bacterial Suspensions...8
- 7 Carriers and Primary Packaging...8
- 8 Contaminated Carriers and Biological Indicators...9
- 9 Bacterial Count and Resistance...9
1 Scope
This Document specifies the requirements and test methods for test microorganisms, bacterial suspensions, contaminated carriers, and biological indicators to be used in evaluating the performance of sterilizers and sterilization processes. The sterilizer and sterilization process use pure ethylene oxide gas or a mixture of ethylene oxide gas and dilution gas, with a sterilization temperature range of 29 °C to 65 °C.
NOTE 1.Requirements for validation and control of ethylene oxide sterilization are found in ISO 11135 and ISO 14937.
NOTE 2.Some national or regional regulations may specify workplace safety requirements.
2 Normative References
The provisions in following documents become the essential provisions of this Document through reference in this Document. For the dated documents, only the versions with the dates indicated are applicable to this Document; for the undated documents, only the latest version (including all the amendments) is applicable to this Document.
GB/T 18281.1-2024 Sterilization of health care products - Biological indicators - Part 1.
General requirements (ISO 11138-1.2017, IDT)
ISO 11138-1 Sterilization of health care products – Biological indicators – Part 1.General requirements
3 Terms and Definitions
For the purposes of this Document, the terms and definitions given in ISO 11138-1 apply.
The URLs for the terminology databases maintained by ISO and IEC for standardization are as follows.
--- ISO Online Browsing Platform. http.//wwwi.so.org/obp/.
4 General Requirements
The requirements in ISO 11138-1 apply.
5 Test Microorganisms
5.1 The test microorganism shall be a Bacillus atrophaeus spore, Bacillus subtilis spore, or other microbial strains that demonstrate equivalent performance to the requirements of this
Document.
NOTE 1.Some strains of Bacillus subtilis have been reclassified as Bacillus atrophaeus.
NOTE 2.Bacillus atrophaeus ATCC 9372, NCTC 10073, NCIMB 8058, DSM 2277, NRRL B-4418, and ICP77.18 have been found to be suitable strains1).
5.2 If a test microorganism other than Bacillus atrophaeus is used, the suitability of the resistance of the test microorganism shall be determined.
6 Bacterial Suspensions
The requirements in ISO 11138-1 apply.
7 Carriers and Primary Packaging
7.1 The suitability of carriers and primary packaging materials used for biological indicators
for ethylene oxide sterilization shall be demonstrated to meet the requirements of 5.2 and
Appendix B in GB/T 18281.1-2024.
7.2 The exposure conditions for determining compliance shall be.
a) Minimum exposure temperature. >=55 °C;
b) Sterilization factor. Ethylene oxide gas mass concentration >=800 mg/L at relative humidity >=70%;
c) Maximum exposure temperature. As stated by the biological indicator manufacturer;
1) Bacillus atrophaeus ATCC 9372, NCTC 10073, NCIMB 8058, DSM 2277, NRRL B-4418, and CIP 77.18 are several examples of suitable commercially available products. This information is provided for the convenience of users of this Document and does not imply endorsement of these products.
d) Exposure time. >=6 h.
NOTE. Although within the practical limitations of the ethylene oxide sterilization process, the exposure conditions shall also represent a realistic challenge to the carrier.
8 Contaminated Carriers and Biological Indicators
The requirements in ISO 11138-1 apply.
9 Bacterial Count and Resistance
9.1 The manufacturer shall declare that the resistance characteristics of the biological indicator meet the requirements of 6.4 in GB/T 18281.1-2024.
9.2 The increment of viable count per unit on the biological indicator shall be nominally stated as an integer multiple of <=0.1 × 10n (e.g., per milliliter of bacterial suspension, per contaminated carrier, or per biological indicator).
9.3 The viable count on the contaminated carrier and biological indicator shall be >=1.0 × 10^6.
9.4 Resistance shall be expressed as the D value at 54 °C, in min. The D value of each batch of biological indicator or contaminated carrier shall be nominally stated as the D value at 54 °C (in min), rounded-off to one digit after the decimal point.
9.5 When testing with a mixed test gas (see Appendix B) under the conditions of Appendix A, the D value of suspensions, contamination carriers, or biological indicators containing Bacillus atrophaeus spores shall be >=2.5 min at 54 °C. The D value for other microorganisms shall support their application.
9.6 When testing with 100% ethylene oxide test gas (see Appendix B) under the conditions of
Appendix A, the D value of suspensions, contamination carriers, or biological indicators containing Bacillus atrophaeus spores shall be >=2.0 min at 54 °C. The D value for other microorganisms shall support their application.
9.7 The resistance characteristics specified in this Document and any other part of ISO 11138 shall be defined using the specific key variables of the used sterilization process.
9.8 The D value shall be determined according to the methods in Appendixes C and D of GB/T
18281.1-2024.
9.9 The determination of D value and survival-kill reaction characteristics requires the use of a resistance meter, and application shall be conducted under specified process parameters (see
Appendix A).
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 9 pages — is available in the English PDF.
Referenced standards
Normative references
GB/T 18281.1-2024 · ISO 11138
How to Download GB/T 18281.2-2024
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 18281.3-2024 — Sterilization of health care products - Biological indicators - Part 3: Biological indicators for moist heat sterilization processes
GB/T 18281.4-2024 — Sterilization of health care products - Biological indicators - Part 4: Biological indicators for dry heat sterilization processes