GB 9706.271-2022Medical electrical equipment - Part 2-71: Particular requirements for the basic safety and essential performance of functional near-infrared spectroscopy(NIRS) equipment (English PDF)

Also coversGB9706.271-2022

Open the GB 9706.271-2022 preview as PDF

Preview — first pages of GB 9706.271-2022

This is a limited preview

Buy now to download the full PDF

Issued by

State Administration for Market Regulation, China National Standardization Administration

Level / Type

National · Mandatory

Issue date

December 29, 2022

Implementation date

January 1, 2026

Scope

GB 9706.271-2022 (Medical electrical equipment - Part 2-71: Particular requirements for the basic safety and essential performance of functional near-infrared spectroscopy(NIRS) equipment) is available as an English-translated PDF.

GB 9706.271-2022 — replace. This document applies to functional near-infrared spectroscopy for the intended purpose of aiding diagnosis by itself or as part of an ME system The basic safety and basic performance of the equipment (hereinafter referred to as ME equipment). The scope of this document does not include. a) as part of ME EQUIPMENT for the measurement of hemoglobin oxygen saturation in microvessels (capillaries, arterioles and venules) and degree; b) near-infrared spectroscopy (NIRS) tissue oximeters not intended to be used to obtain the output of a functional near-infrared spectroscopy device; c) Pulse oximeters that are not intended to be used to obtain the output of functional near-infrared spectroscopy equipment, see YY0784 for requirements; d) For frequency domain and time domain equipment for functional near-infrared spectroscopy, different inspection procedures than those specified in this document may be used; e) A functional near-infrared spectroscopic device for measuring changes in the concentration of chromophores other than oxygenated and deoxygenated hemoglobin, using different The inspection procedures specified in this document.

Document preview — GB 9706.271-2022

National Standard of the People's Republic of China

ICS
11.040.55
Classification
C 40

Issued by: State Administration for Market Regulation, China National Standardization Administration

Contents

  • Preface
  • Introduction

foreword

This document is in accordance with the provisions of GB/T 1.1-2020 "Guidelines for Standardization Work Part 1.Structure and Drafting Rules for Standardization Documents"

drafting.

This document is part 2-71 of GB 9706 "Medical Electrical Equipment". GB 9706 has released the following parts.

--- Part 1.General requirements for basic safety and essential performance;

--- Part 1-3.General requirements for basic safety and essential performance Collateral standard. Radiation protection of diagnostic X-ray equipment;

--- Part 2-1.Particular requirements for the basic safety and essential performance of electron accelerators with energies ranging from 1 MeV to 50 MeV;

--- Part 2-2.Special requirements for basic safety and basic performance of high-frequency surgical equipment and high-frequency accessories;

--- Part 2-3.Special requirements for basic safety and basic performance of short-wave therapy equipment;

--- Part 2-4.Particular requirements for basic safety and basic performance of cardiac defibrillators;

--- Part 2-5.Particular requirements for basic safety and basic performance of ultrasonic physiotherapy equipment;

--- Part 2-6.Particular requirements for basic safety and basic performance of microwave therapy equipment;

--- Part 2-8.Particular requirements for basic safety and essential performance of therapeutic X-ray equipment with energy from 10kV to 1MV;

--- Part 2-11.Particular requirements for basic safety and essential performance of gamma beam therapy equipment;

--- Part 2-12.Particular requirements for basic safety and basic performance of critical care ventilators;

--- Part 2-13.Special requirements for basic safety and basic performance of anesthesia workstations;

--- Part 2-16.Particular requirements for basic safety and essential performance of hemodialysis, hemodiafiltration and hemofiltration equipment;

--- Part 2-17.Particular requirements for basic safety and basic performance of automatically controlled brachytherapy afterloading equipment;

--- Part 2-18.Particular requirements for basic safety and essential performance of endoscopic equipment;

--- Part 2-19.Particular requirements for basic safety and basic performance of infant incubators;

--- Part 2-22.Special requirements for basic safety and essential performance of laser equipment for surgery, plastic surgery, therapy and diagnosis;

--- Part 2-24.Particular requirements for basic safety and essential performance of infusion pumps and infusion controllers;

--- Part 2-25.Specific requirements for basic safety and basic performance of electrocardiographs;

--- Part 2-26.Special requirements for basic safety and basic performance of EEG machines;

--- Part 2-27.Special requirements for basic safety and basic performance of ECG monitoring equipment;

--- Part 2-28.Particular requirements for basic safety and essential performance of medical diagnostic X-ray tube assemblies;

--- Part 2-29.Particular requirements for basic safety and basic performance of radiotherapy simulators;

--- Part 2-36.Particular requirements for the basic safety and essential performance of in vitro induced lithotripsy equipment;

--- Part 2-37.Particular requirements for basic safety and essential performance of ultrasonic diagnostic and monitoring equipment;

--- Part 2-39.Particular requirements for basic safety and basic performance of peritoneal dialysis equipment;

--- Part 2-43.Particular requirements for basic safety and essential performance of X-ray equipment for interventional operation;

--- Part 2-44.Particular requirements for basic safety and essential performance of X-ray computed tomography equipment;

--- Part 2-45.Particular requirements for basic safety and basic performance of mammography equipment and mammography stereotaxic devices;

--- Part 2-54.Special requirements for basic safety and basic performance of X-ray photography and fluoroscopy equipment;

--- Part 2-55.Particular requirements for basic safety and basic performance of respiratory gas monitors;

--- Part 2-60.Particular requirements for basic safety and essential performance of dental equipment;

--- Part 2-63.Particular requirements for basic safety and basic performance of dental X-ray machines for extraoral imaging;

--- Part 2-65.Particular requirements for basic safety and basic performance of dental X-ray machines for intraoral imaging;

--- Part 2-71.Particular requirements for basic safety and essential performance of functional near-infrared spectroscopy (NIRS) equipment;

--- Part 2-75.Particular requirements for basic safety and essential performance of photodynamic therapy and photodynamic diagnostic equipment;

--- Part 2-83.Particular requirements for the basic safety and essential performance of household phototherapy equipment;

--- Part 2-90.Particular requirements for basic safety and essential performance of high-flow respiratory therapy equipment.

This document is modified to adopt IEC 80601-2-71.2015 "Medical Electrical Equipment Part 2-71.Functional Near Infrared Spectroscopy (NIRS)

Particular Requirements for Basic Safety and Essential Performance of Equipment".

The technical differences between this document and IEC 80601-2-71.2015 and their reasons are as follows.

--- Replaced IEC 60825-1.2014 with the normative reference GB 7247.1-2012 (see.201.2,.201.3,.201.7.2.101,

--- Replaced IEC 60601-1-6 (see.201.2,.201.3) with the normative referenced YY/T 9706.106 to adapt to my country's national conditions;

--- Replaced IEC 60601-1 (see.201.2,.201.3) with the normatively quoted GB 9706.1 to adapt to my country's national conditions;

--- Replaced ISO 80601-2-61 (see.201.1.1,.201.2) with normatively quoted YY0784 to adapt to my country's national conditions;

--- Increased the normative reference document YY9706.257-2021 (see.201.2) to adapt to the national conditions of our country;

---Added the classification information specified in.201.6 of YY9706.257-2021 (see.201.7.2.101, Table.201.C.101) to apply

Functional near-infrared spectroscopy equipment for incoherent light sources;

--- It is clear that the filter for data processing in signal stability is a low-pass filter (see.201.12.1.101.6) to avoid ambiguity;

---Changed the calculation formula of the standard deviation deltaS in the signal-to-noise ratio (see.201.12.1.101.8), the international standard formula was edited incorrectly, and it was modified

make corrections;

--- Increased the test method of average optical power in the pulse output mode (see.201.12.1.101.2), so as to be applicable to all

In the case of pulse modulation;

--- Increase the shutter switching time requirements (see.201.12.1.107) to ensure the test accuracy of the response time.

The following editorial changes have been made to this document.

--- Deleted "References";

--- Deleted the "index of term definitions".

Please note that some contents of this document may refer to patents. The issuing agency of this document assumes no responsibility for identifying patents.

Introduction

Safety standards for medical electrical equipment, also known as the 9706 series of standards, consist of general standards, collateral standards, specific standards, guidelines and interpretations.

---General standard. The safety standard that medical electrical equipment should be generally applicable, that is, the equipment that meets the definition of medical electrical equipment should meet this

Basic standard requirements.

--- Parallel standards. safety standards that should be generally applicable to medical electrical equipment, but in most cases are limited to certain specific functions or special

It is only the equipment that needs to meet the requirements of such standards.

---Special standards. Safety standards applicable to a certain type of medical electrical equipment, and not all medical electrical equipment have specific standards

standard.

---Guidelines and interpretations. explanations for the application guidelines and clauses of the relevant requirements in the standards involved.

In the GB 9706 series of standards, in addition to the published standards and this document, the planned standards are as follows.

--- Part 2-66.Particular requirements for basic safety and essential performance of hearing devices and hearing device systems. The purpose is to build listening

Particular requirements for basic safety and essential performance of equipment and hearing device systems.

This document deals with the basic safety and essential performance requirements for functional near-infrared spectroscopy (NIRS) devices. This document has been amended and supplemented by

GB 9706.1-2020 "Medical Electrical Equipment Part 1.General Requirements for Basic Safety and Essential Performance".

Items marked with an asterisk (*) in this document have specific guidelines and rationale related to the project in Appendix AA. Appendix AA

Relevant content not only contributes to the correct application of this document, but also speeds up the revision of the standard due to changes in clinical practice or technical development in a timely manner.

standard process. Appendix CC gives the correspondence between this document and the basic principles of YY/T 0467-2016.

Medical Electrical Equipment Part 2-71.Functionality

Basic safety and

Basic performance-specific requirements

201.1 Scope, purpose and relevant standards

Except as described below, Chapter 1 of the general standard applies.

201.1.1 Scope

replace.

This document applies to functional near-infrared spectroscopy for the intended purpose of aiding diagnosis by itself or as part of an ME system

The basic safety and basic performance of the equipment (hereinafter referred to as ME equipment).

The scope of this document does not include.

a) as part of ME EQUIPMENT for the measurement of hemoglobin oxygen saturation in microvessels (capillaries, arterioles and venules)

and degree;

b) near-infrared spectroscopy (NIRS) tissue oximeters not intended to be used to obtain the output of a functional near-infrared spectroscopy device;

c) Pulse oximeters that are not intended to be used to obtain the output of functional near-infrared spectroscopy equipment, see YY0784 for requirements;

d) For frequency domain and time domain equipment for functional near-infrared spectroscopy, different inspection procedures than those specified in this document may be used;

e) A functional near-infrared spectroscopic device for measuring changes in the concentration of chromophores other than oxygenated and deoxygenated hemoglobin, using different

The inspection procedures specified in this document.

201.1.2 Purpose

replace.

The purpose of this document is to specify the basic safety and essential performance specific requirements for functional near-infrared devices.

201.1.3 Collateral Standards

Supplement.

This document refers to Chapter 2 of the General Standard and applicable collateral standards listed in.201.2 of this document.

GB 9706.103 and YY/T 9706.110 are not applicable. All other published GB 9706.1 collateral standards apply.

201.1.4 Particular standards

replace.

In the GB 9706 series of standards, specific standards may modify, replace or delete the requirements contained in general standards or collateral standards to apply

Other BASIC SAFETY and ESSENTIAL PERFORMANCE requirements may also be added depending on the ME EQUIPMENT considered.

The requirements of a particular standard take precedence over those of a general standard.

In this document, GB 9706.1-2020 is referred to as the general standard. Collateral standards are indicated by their respective document numbers.

The numbering of chapters and clauses in this document corresponds to the general standard by adding the prefix "201" (for example,.201.1 in this document corresponds to the general standard

Chapter 1), or correspond to the applicable collateral standard by adding the prefix "20x", where x is the number of the international standard corresponding to the parallel standard

The last digit (for example, 202.4 in this document corresponds to the content of Chapter 4 in the collateral standard YY9706.102-2021, and 203.4 in this document corresponds to

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.

Referenced standards

How to Buy GB 9706.271-2022

  1. 1Add to cart. Click the "Buy GB 9706.271-2022" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB 9706.271-2022

$425.00

$360.00for partners