GB 9706.263-2020Medical electrical equipment - Part 2-63: Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment (English PDF)

Also coversGB9706.263-2020

Open the GB 9706.263-2020 preview as PDF

Preview — first pages of GB 9706.263-2020

This is a limited preview

Buy now to download the full PDF

Issued by

State Administration for Market Regulation, China National Standardization Administration

Level / Type

National · Mandatory

Issue date

November 17, 2020

Implementation date

May 1, 2023

Scope

GB 9706.263-2020 (Medical electrical equipment - Part 2-63: Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment) is available as an English-translated PDF.

GB 9706.263-2020 — replace. This section applies to the basic safety and basic performance of extraoral imaging dental X-ray machines (hereinafter referred to as ME equipment). The range includes ME system for this type of ME equipment. Note 1.Including oral surface body layer equipment, cephalometric equipment and dental volume reconstruction equipment defined in.201.3.203 (hereinafter referred to as DVR). Note 2.DVR includes dental CBCT (cone beam computed tomography), and other names in some regions, such as DVT (digital volume tomography); DVR also includes tomosynthesis imaging. Note 3.It may include imaging of other anatomical parts (such as hands) required for dental treatment (such as orthodontic treatment). Note 4.It may include imaging of anatomical parts of interest to ENT (ear, nose, throat) professional physicians. The scope of this part is limited to the following X-ray equipment. ---X-ray tube components containing high-voltage transformer components and ---The geometric relationship between the X-ray source, the patient's required imaging location, and the X-ray image receiver is preset, and During the period of use, the operator cannot change arbitrarily. Note 5.Intraoral imaging dental X-ray machines are not within the scope of this section. Note 6.The distance from the focus to the image receiver and the distance from the focus to the subject are preset in the dental X-ray machine for extraoral imaging. Note 7.Due to the above restrictions, dental X-ray machines that are not within the scope of this section can use the applicable clauses of IEC 60601-2-54 with this section. ME equipment within the scope of GB 9706.18, IEC 60601-2-54, GB 9706.24, IEC 60601-2-65 or GB 9706.23 And the ME system is not within the scope of this special standard. The scope of application of this part also does not include radiation therapy simulation equipment and bone or group Density absorption equipment does not include dental fluoroscopy equipment. Within the specified scope, the provisions of this special standard will replace and replace GB 9706.3-2000 Medical Electrical Equipment Part 2.Diagnosis Special requirements for the safety of high-voltage generators for X-ray generators and GB 9706.14-1997 Medical Electrical Equipment Part 2.X-ray Special requirements for the safety of auxiliary equipment of cable equipment. Note 8.The requirements for X-ray generators and ancillary equipment, namely GB 9706.3 and GB 9706.14 described above, have been included in GB 9706.1-2020 Or special standards. Therefore, GB 9706.3 and GB 9706.14 are not applicable to the GB 9706 standard system to which the extraoral imaging dental X-ray machine belongs section. All the requirements of the combined X-ray tube assembly have been covered by this section, so unless the X-ray tube assembly is replaceable, GB 9706.11 also does not apply to the ME equipment specified in this special standard. Note 9.The requirements for dental X-ray machines in the previous version of the parallel standard GB 9706.12 and the special standard GB 9706.11 have been extracted and included in this section Scored. Note 10.In the scope of this special standard, the X-ray tube assembly of the X-ray equipment refers to the X-ray combined handpiece.

Document preview — GB 9706.263-2020

National Standard of the People's Republic of China

ICS
11.040.50
Classification
C 43

Issued by: State Administration for Market Regulation, China National Standardization Administration

Contents

  • 201.1 Scope, purpose and related standards1
  • 201.2 Normative references3
  • 201.3 Terms and Definitions3
  • 201.4 General requirements4
  • 201.5 General requirements for ME equipment testing5
  • 201.6 Classification of ME equipment and ME systems5
  • 201.7 Marking, marking and documentation of ME equipment5
  • 201.8 Protection of ME equipment against electric shock hazard7
  • 201.9 Protection of ME equipment and ME systems against mechanical hazards8
  • 201.10 Protection against unwanted or excessive radiation hazards (sources)8
  • 201.11 Protection against over-temperature and other hazards (sources)9
  • 201.12 Control and instrument accuracy and protection against hazardous outputs9
  • 201.13 Dangerous situations and fault states9
  • 201.14 Programmable Medical Electrical System (PEMS)9
  • 201.15 Structure of ME equipment9
  • 201.16 ME System9
  • 201.17 Electromagnetic compatibility of ME equipment and ME systems9
  • 202 Electromagnetic Compatibility-Requirements and Testing9
  • 203 Radiation protection of diagnostic X-ray equipment9
  • Appendix18
  • Appendix C (informative appendix) External marking of ME equipment, ME system or its components19
  • Appendix AA (informative appendix) special guide and principle explanation20
  • References25

Foreword

GB 9706 "Medical Electrical Equipment" is divided into the following parts.

---Part 1.General requirements for basic safety and basic performance;

---Part 1-3.General requirements for basic safety and basic performance Parallel standard. Radiation protection of diagnostic X-ray equipment;

---Part 2-1.Specific requirements for basic safety and basic performance of electron accelerators with energy from 1MeV to 50MeV;

---Part 2-2.Special requirements for the basic safety and basic performance of high-frequency surgical equipment and high-frequency accessories;

---Part 2-3.Special requirements for basic safety and basic performance of short-wave treatment equipment;

---Part 2-4.Special requirements for the basic safety and basic performance of cardiac defibrillators;

---Part 2-5.Special requirements for the basic safety and basic performance of ultrasonic physical therapy equipment;

---Part 2-6.Special requirements for the basic safety and basic performance of microwave therapy equipment;

---Part 2-8.Special requirements for the basic safety and basic performance of therapeutic X-ray equipment with an energy of 10kV to 1MV;

---Part 2-11.Special requirements for the basic safety and basic performance of gamma beam therapy equipment;

---Part 2-12.Special requirements for the basic safety and basic performance of intensive care ventilators;

---Part 2-13.Special requirements for basic safety and basic performance of anesthesia workstations;

---Part 2-16.Specific requirements for basic safety and basic performance of hemodialysis, hemodiafiltration and hemofiltration equipment;

---Part 2-17.Special requirements for basic safety and basic performance of automatic control brachytherapy after-installation equipment;

---Part 2-18.Specific requirements for basic safety and basic performance of endoscopic equipment;

---Part 2-19.Special requirements for basic safety and basic performance of infant incubators;

---Part 2-22.Special requirements for basic safety and basic performance of laser equipment for surgery, plastic surgery, treatment and diagnosis;

---Part 2-24.Special requirements for basic safety and basic performance of infusion pumps and infusion controllers;

---Part 2-25.Special requirements for basic safety and basic performance of electrocardiograph;

---Part 2-26.Specific requirements for the basic safety and basic performance of EEG machines;

---Part 2-27.Special requirements for basic safety and basic performance of ECG monitoring equipment;

---Part 2-28.Special requirements for the basic safety and basic performance of medical diagnostic X-ray tube components;

---Part 2-29.Specific requirements for the basic safety and basic performance of radiotherapy simulators;

--- Part 2-36.Special requirements for the basic safety and basic performance of in vitro lithotripsy equipment;

---Part 2-37.Special requirements for the basic safety and basic performance of ultrasonic diagnostic and monitoring equipment;

---Part 2-39.Specific requirements for the basic safety and basic performance of peritoneal dialysis equipment;

---Part 2-43.Special requirements for the basic safety and basic performance of interventional X-ray equipment;

---Part 2-44.Special requirements for the basic safety and basic performance of X-ray computed tomography equipment;

---Part 2-45.Special requirements for the basic safety and basic performance of mammography equipment and mammography stereotaxic devices;

---Part 2-54.Specific requirements for the basic safety and basic performance of X-ray photography and fluoroscopy equipment;

---Part 2-60.Special requirements for basic safety and basic performance of dental equipment;

---Part 2-63.Special requirements for the basic safety and basic performance of extraoral imaging dental X-ray machines;

---Part 2-65.Specific requirements for the basic safety and basic performance of intraoral imaging dental X-ray machines;

---Part 2-66.Special requirements for basic safety and basic performance of hearing equipment and hearing equipment system.

This part is part 2-63 of GB 9706.

This section was drafted in accordance with the rules given in GB/T 1.1-2009.

This part uses the redrafting law to amend and adopt IEC 60601-2-63.2017 ``Medical Electrical Equipment Part 2-63.Extraoral Imaging Dental

Special requirements for the basic safety and basic performance of X-ray machines.

The technical differences between this part and IEC 60601-2-63.2017 and the reasons are as follows.

---Regarding normative reference documents, this section has made adjustments with technical differences to adapt to my country's technical conditions and adjustments.

The situation is collectively reflected in.201.2 "Normative Reference Documents", and the specific adjustments are as follows.

* Replace IEC 60601-1.2012 (see.201.2 and related provisions) with GB 9706.1-2020 which is modified to adopt international standards;

* Replace IEC 60601-1-2 with YY0505 which is equivalent to adopting international standards (see.201.1.3 and related provisions);

* Replace IEC 60336 (see.201.7.9.3) with YY/T 0063 equivalent to the international standard;

* Replace IEC 60601-1-3.2008/AMD1.2013 with GB 9706.12 equivalent to the international standard (see.201.2 and related provisions);

--- Deleted the content of "IEC 60601-1-3" and "AnnexB" in 203.6.3.1, and added the description of "priority number system", which is in harmony with the current domestic standards.

This section has made the following editorial changes.

---The foreword and introduction of international standards have been deleted;

--- The note in 203.4.101.1 has been deleted, which is in harmony with the current domestic standards;

--- Deleted the words "replacement" and "addition" in the normative references;

---All terms are represented in bold;

---The term index is deleted.

Please note that certain contents of this document may involve patents. The issuing agency of this document is not responsible for identifying these patents.

This part is proposed and managed by the State Drug Administration.

Medical electrical equipment-Part 2-63.Extraoral imaging dentistry

X-ray machine basic safety and basic performance special requirements

201.1 Scope, purpose and related standards

In addition to the following, Chapter 1 of the general standard 1) applies.

1) The general standard is GB 9706.1-2020 "Medical Electrical Equipment Part 1.General Requirements for Basic Safety and Basic Performance".

201.1.1 Scope

replace.

This section applies to the basic safety and basic performance of extraoral imaging dental X-ray machines (hereinafter referred to as ME equipment). The range includes

ME system for this type of ME equipment.

Note 1.Including oral surface body layer equipment, cephalometric equipment and dental volume reconstruction equipment defined in.201.3.203 (hereinafter referred to as DVR).

Note 2.DVR includes dental CBCT (cone beam computed tomography), and other names in some regions, such as DVT (digital volume tomography);

DVR also includes tomosynthesis imaging.

Note 3.It may include imaging of other anatomical parts (such as hands) required for dental treatment (such as orthodontic treatment).

Note 4.It may include imaging of anatomical parts of interest to ENT (ear, nose, throat) professional physicians.

The scope of this part is limited to the following X-ray equipment.

---X-ray tube components containing high-voltage transformer components and

---The geometric relationship between the X-ray source, the patient's required imaging location, and the X-ray image receiver is preset, and

During the period of use, the operator cannot change arbitrarily.

Note 5.Intraoral imaging dental X-ray machines are not within the scope of this section.

Note 6.The distance from the focus to the image receiver and the distance from the focus to the subject are preset in the dental X-ray machine for extraoral imaging.

Note 7.Due to the above restrictions, dental X-ray machines that are not within the scope of this section can use the applicable clauses of IEC 60601-2-54 with this section.

ME equipment within the scope of GB 9706.18, IEC 60601-2-54, GB 9706.24, IEC 60601-2-65 or GB 9706.23

And the ME system is not within the scope of this special standard. The scope of application of this part also does not include radiation therapy simulation equipment and bone or group

Density absorption equipment does not include dental fluoroscopy equipment.

Within the specified scope, the provisions of this special standard will replace and replace GB 9706.3-2000 Medical Electrical Equipment Part 2.Diagnosis

Special requirements for the safety of high-voltage generators for X-ray generators and GB 9706.14-1997 Medical Electrical Equipment Part 2.X-ray

Special requirements for the safety of auxiliary equipment of cable equipment.

Note 8.The requirements for X-ray generators and ancillary equipment, namely GB 9706.3 and GB 9706.14 described above, have been included in GB 9706.1-2020

Or special standards. Therefore, GB 9706.3 and GB 9706.14 are not applicable to the GB 9706 standard system to which the extraoral imaging dental X-ray machine belongs

section.

All the requirements of the combined X-ray tube assembly have been covered by this section, so unless the X-ray tube assembly is replaceable,

GB 9706.11 also does not apply to the ME equipment specified in this special standard.

Note 9.The requirements for dental X-ray machines in the previous version of the parallel standard GB 9706.12 and the special standard GB 9706.11 have been extracted and included in this section

Scored.

Note 10.In the scope of this special standard, the X-ray tube assembly of the X-ray equipment refers to the X-ray combined handpiece.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.

How to Buy GB 9706.263-2020

  1. 1Add to cart. Click the "Buy GB 9706.263-2020" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB 9706.263-2020

$425.00

$360.00for partners