GB 5009.293-2023National food safety standard - Determination of monocaprylin in foods (English PDF)
Also coversGB5009.293-2023
Open the GB 5009.293-2023 preview as PDF
This is a limited preview
Buy now to download the full PDF
Issued by
NHC; SAMR
Level / Type
National · Mandatory
Issue date
September 6, 2023
Implementation date
September 6, 2024
Scope
GB 5009.293-2023 (National food safety standard - Determination of monocaprylin in foods) is available as an English-translated PDF.
GB 5009.293-2023 — This standard specifies the gas chromatography and gas chromatography-mass spectrometry methods for the determination of capryl monoglyceride in food. This standard is applicable to the determination of capryl monoglyceride in wet flour products, pastries, baked food fillings and surface paste (only bean paste), meat sausages, and spicy strips. Method A - Gas chromatography
Document preview — GB 5009.293-2023
National Standard of the People's Republic of China
Issued by: NHC; SAMR
Contents
- 1 Scope3
- Method A - Gas chromatography3
- 2 Principles3
- 3 Reagents and materials3
- 4 Instruments and equipment4
- 5 Analysis steps5
- 6 Result calculation7
- 7 Precision7
- 8 Others8
- Method B - Gas chromatography-mass spectrometry8
- 9 Principles8
- 10 Reagents and materials8
- 11 Instruments and equipment9
- 12 Analysis steps9
- 13 Result calculation11
- 14 Precision11
- 15 Others11
- Appendix A Gel permeation chromatography reference analysis conditions12
- Appendix B Gas chromatogram and gas chromatography-mass spectrometry selected ion monitoring (SIM) chromatogram of capryl monoglyceride standard solution13
- Appendix C Monitoring ions and relative abundance of capryl monoglyceride15
1 Scope
This standard specifies the gas chromatography and gas chromatography-mass spectrometry methods for the determination of capryl monoglyceride in food.
This standard is applicable to the determination of capryl monoglyceride in wet flour products, pastries, baked food fillings and surface paste (only bean paste), meat sausages, and spicy strips.
Method A - Gas chromatography
2 Principles
Capryl monoglyceride in wet flour products is extracted with n-hexane saturated acetonitrile, degreased with acetonitrile-saturated n-hexane, measured by gas chromatography, qualitatively analyzed according to retention time, and quantitatively analyzed with the external standard method.
Capryl monoglyceride in bean paste, pastries, meat sausages and spicy strips is extracted with n-hexane saturated acetonitrile, purified by gel permeation chromatography, and measured by gas chromatography, qualitatively analyzed according to retention time, and quantitatively analyzed with external standard method.
3 Reagents and materials
Unless otherwise stated, the reagents used in this method are all chromatographically pure, and the water is first-grade water specified in GB/T 6682.
3.1 Reagents
3.1.1 Absolute ethanol (C2H6O), chromatographically pure.
3.1.2 Acetonitrile (CH3CN), chromatographically pure.
3.1.3 n-Hexane (C6H14), chromatographically pure.
3.1.4 Ethyl acetate (CH3COOC2H5), chromatographically pure.
3.1.5 Cyclohexane (C6H12), chromatographically pure.
3.2 Reagent preparation
3.2.1 n-hexane saturated acetonitrile, acetonitrile-saturated n-hexane: Measure 100 mL
of acetonitrile (3.1.2) into a 250 mL separatory funnel, add 100 mL of n-hexane, shake vigorously for several minutes, and let it stand; after it separates into layers, take the lower layer, then the n-hexane saturated acetonitrile solution is obtained; take the upper layer, then the acetonitrile-saturated n-hexane solution is obtained.
3.2.2 Ethyl acetate-cyclohexane (1+1): Mix ethyl acetate and cyclohexane in a volume
ratio of 1:1.
3.3 Standards
Capryl monoglyceride standard (capryl monoglyceride, C11H22O4, CAS number:
26402-26-6) with a purity of >=98%, or a standard with national authentication and a Reference Material Certificate.
3.4 Preparation of standard solution
3.4.1 Capryl monoglyceride standard stock solution (10.0 mg/mL): Accurately weigh
0.1 g (accurate to 0.0001 g) of standard capryl monoglyceride in a 10 mL volumetric flask, dissolve it in absolute ethanol, dilute to the mark, and mix well. Transfer the solution to a brown glass bottle and store it at -20 °C. It will be valid for 6 months.
3.4.2 Capryl monoglyceride standard intermediate solution (1.0 mg/mL): Take 1 mL of
capryl monoglyceride standard stock solution (10.0 mg/mL) into a 10 mL volumetric flask, dilute to volume with absolute ethanol, and mix well. Prepare it fresh just before use.
3.4.3 Capryl monoglyceride standard series working solution: Take 20 uL, 50 uL of
capryl monoglyceride standard intermediate solution (1.0 mg/mL), and 10 uL, 20 uL, 50 uL and 100 uL of capryl monoglyceride standard stock solution (10.0 mg/mL)
respectively; use absolute ethanol to prepare capryl monoglyceride standard series working solutions with concentrations of 0.0200 mg/mL, 0.0500 mg/mL, 0.100 mg/mL, 0.200 mg/mL, 0.500 mg/mL, and 1.00 mg/mL, respectively. Prepare the solutions fresh just before use.
4 Instruments and equipment
4.1 Gas chromatograph (GC): with flame ionization detector (FID).
at 30 °C to 40 °C until nearly dry; dilute to 2 mL with absolute ethanol, and pass through a 0.22 um organic filter membrane, for gas chromatography determination later.
5.1.3.2 Bean paste, pastries, meat sausages and spicy strips
Transfer bean paste, pastry, meat sausage, and spicy strip concentrate processed in 5.1.2 to a 10 mL volumetric flask, wash the eggplant-shaped evaporation flask twice with 6 mL of ethyl acetate-cyclohexane (3 mL each time), and combine the washing liquid into the same 10 mL volumetric flask; dilute to the mark with ethyl acetate-cyclohexane (1+1), and filter with a 0.22 um organic filter membrane. After the filtrate is purified by gel permeation chromatography, collect the eluate (for gel permeation chromatography conditions, please refer to Appendix A), rotary evaporate at 30 °C~40 °C until almost dry, dilute to 1 mL with absolute ethanol, and pass through a 0.22 um organic filter membrane, for gas chromatography determination later.
5.2 Instrument reference conditions
The reference conditions for gas chromatography are as follows:
a) Chromatographic column: (35%-phenyl)-methylpolysiloxane capillary column, 30 mx0.25 mm (inner diameter) x0.25 um (film thickness), or equivalent.
b) Temperature programming: The initial temperature is 60 °C, and hold for 1 minute; increase temperature to 340 °C at 20 °C/min, and hold for 5 minutes.
c) Inlet temperature: 340 °C.
d) Detector temperature: 340 degrees C, with a hydrogen flow of 30 mL/min, an air flow of 400 mL/min, and a makeup flow of 30 mL/min.
e) Carrier gas: nitrogen with a purity of >=99.999%, constant flow mode, and flow rate of 1.0 mL/min.
f) Injection method: Splitless injection; open the valve after 0.5 minutes.
g) Injection volume: 1.0 uL.
5.3 Preparation of standard curve
Inject the capryl monoglyceride standard series working solutions into the gas chromatograph respectively, and measure the corresponding peak area of capryl monoglyceride. Take the concentration of capryl monoglyceride in the standard working solution as the abscissa, the peak area of capryl monoglyceride as ordinate, and draw the standard curve. For the chromatogram of the capryl monoglyceride standard solution, see Figure B.1 in Appendix B.
8 Others The detection limit of this method is 0.050 g/kg, and the quantitation limit is 0.10 g/kg.
Method B - Gas chromatography-mass spectrometry
9 Principles Capryl monoglyceride in wet flour products is extracted with n-hexane saturated acetonitrile, degreased with acetonitrile-saturated n-hexane, and measured with gas chromatography-mass spectrometry; the selected ion monitoring mode (SIM) is used, the qualitative analysis is carried out according to the retention time and the abundance ratio of qualitative ion fragments, and external standard method is used for quantitative analysis.
Capryl monoglyceride in bean paste, pastries, meat sausages and spicy strips is extracted with n-hexane saturated acetonitrile, purified with gel permeation chromatography, and measured with gas chromatography-mass spectrometry; the selected ion monitoring mode (SIM) is used, the qualitative analysis is carried out according to the retention time and the abundance ratio of qualitative ion fragments, and external standard method is used for quantitative analysis.
10 Reagents and materials
10.1 Reagents Same as 3.1.
10.2 Reagent preparation Same as 3.2.
10.3 Standards Same as 3.3.
10.4 Preparation of standard solution
10.4.1 Capryl monoglyceride standard stock solution (10.0 mg/mL): Same as 3.4.1.
10.4.2 Capryl monoglyceride standard intermediate solution (1.0 mg/mL): Same as 3.4.2.
10.4.3 Capryl monoglyceride standard series working solution: Take 20 uL, 50 uL of capryl monoglyceride standard intermediate solution (1.0 mg/mL) and 10 uL, 20 uL, 50 uL and 100 uL of capryl monoglyceride standard stock solution (10.0 mg/mL)
respectively, and prepare capryl monoglyceride matrix standard solutions with the concentrations of 0.0200 mg/mL, 0.0500 mg/mL, 0.100 mg/mL, 0.200 mg/mL, 0.500 mg/mL, and 1.00 mg/mL, respectively. Prepare the solutions fresh just before use.
11 Instruments and equipment
11.1 Gas chromatography-mass spectrometer (GC-MS): equipped with electron impact ion source (EI).
11.2 Gel permeation chromatograph (GPC): See Appendix A for relevant parameters.
11.3 Electronic balance: The sensitivity is 0.1 mg and 0.01 g, respectively.
11.4 Vortex mixer.
11.5 Centrifuge.
11.6 Rotary evaporator.
11.7 Nitrogen blower.
11.8 Microporous filter membrane (organic phase): 0.22 um.
11.9 Ultrasonic generator: The power is 185 W.
12 Analysis steps
12.1 Sample pretreatment It is the same as 5.1, for gas chromatography-mass spectrometry measurement.
12.2 Gas chromatography reference conditions
12.2.1 Chromatographic column: (35%-phenyl)-methylpolysiloxane capillary column,
30 mx0.25 mm (inner diameter) x0.25 um (film thickness), or equivalent.
12.2.2 Temperature programming: The initial temperature is 60 °C, and hold for 1 minute; increase temperature to 340 °C at 20 °C/min, and hold for 5 minutes.
12.2.3 Inlet temperature: 340 degrees C.
12.2.4 Carrier gas: helium with a purity of >=99.999%, constant flow mode, and flow rate of 1.0 mL/min.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.
How to Buy GB 5009.293-2023
- 1Add to cart. Click the "Buy GB 5009.293-2023" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 5009.1-2025 — National food safety standard - Physical and chemical testing - General principles