GB 23101.4-2008[GB/T 23101.4-2008] Implants for surgery -- Hydroxyapatite -- Part 4: Determination of coating adhesion strength (English PDF)
外科植入物 羟基磷灰石 第4部分:涂层粘结强度的测定
This is a limited preview
Buy now to download the full PDF (12 pages)
Need quotation for your management?
Get a formal quotation for GB 23101.4-2008 in 30 minutes.
Issued by
General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China
Level / Type
National · Mandatory
Issue date
December 30, 2008
Implementation date
March 1, 2010
Scope
GB 23101.4-2008 is the English-translated version of 外科植入物 羟基磷灰石 第4部分:涂层粘结强度的测定.
GB 23101.4-2008 is the Chinese standard "[GB/T 23101.4-2008] Implants for surgery -- Hydroxyapatite -- Part 4: Determination of coating adhesion strength". Its scope clause reads: GB 23101 of the measurement methods specified in this part of hydroxyapatite coating adhesion strength.
Its clauses include terms and definitions; determination of coating adhesion strength; hydroxyapatite coating tensile test report.
Found what you need?
Document preview — GB 23101.4-2008
National Standard of the People's Republic of China
- ICS
- 11.040.40
- Classification
- C 35
Issued by: General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions
1 Scope
GB 23101 of the measurement methods specified in this part of hydroxyapatite coating adhesion strength.
2 Normative references
The following documents contain provisions which, through reference in this Part of GB 23101, constitute provisions of this part. For dated reference documents
Member, all subsequent amendments (not including errata content) or revisions do not apply to this section, however, encouraged to reach under this section
Parties to research agreement to use the latest versions of these documents. For undated reference documents, the latest versions apply to this section.
Check tension and (or) pressure testing machine force measuring system. GB/T 16825.1-2002 static uniaxial testing machines - Part 1
Test and calibration (ISO 7500-1.1999, IDT)
3 Terms and Definitions
The following terms and definitions apply to this part of GB 23101.
3.1
By sintering the powder grains integrated into cohesion hydroxyapatite crystals block.
3.2
By heat, vapor or liquid phase technique in a non-metal or metal deposited on the surface of distinct layers of different materials.
3.3
Having a crystal structure of the International Commission for Diffraction Data (ICDD) powder diffraction card PDF9-432 characterized by chemical composition.
Note. Chemical formula. Ca5 (OH) (PO4) 3.
3.4
By heating the powder particle surface area and a significant reduction in the overall volume of the powder and densified, and followed by improving the mechanical properties of ceramics production process.
4 Determination of coating adhesion strength
4.1 Principle
Coating adhesion strength is measured on a cylindrical test assembly by applying a uniaxial tensile load, the test assembly consists of a hydroxyl
Hydroxyapatite coating sample uncoated bond a member.
4.2 Test Equipment
4.2.1 mechanical testing machine
Load capacity of not less than 30kN, accuracy no less than ± 2% of full scale mechanical testing machine. The tensile load shall be applied perpendicular to the applied coating on flat surface.
GB 23101.4-2008/ISO 13779-4.2002 (According to GB/T 16825.1 test and calibration of mechanical measuring system)
4.2.2 Load component
The load test machine is transmitted to the apparatus test assembly.
Load component should ensure that the shaft and the shaft of the test assembly test deviation between <=1 ° and ± 0.1mm, thus ensuring vertical surface coating test
To the loading axis. Eccentric loading resulting assembly results invalid.
Figure 1 is a combination of one of the yoke and pin from the loading component instance to meet these requirements. The samples were coated and uncoated parts
Each consists of two vertical pins fixed to allow for axial load deviation is minimized. The specific dimensions charging assembly as specified in Table 1, the appropriate
Close to the expected maximum bond or cohesive strength of the coating test samples 50MPa.
1 --- test module;
2 --- pin;
3 --- yoke;
4 --- binder;
5 --- coating to be tested.
Figure 1 so that the smallest deviation from axial loading loading component diagram
4.2.3 Test assembly
Test component parts (the coated samples and uncoated parts) are bonded together with a polymeric binder.
Uncoated substrate samples with the coating member should be the same material, preferred materials are shown in Table 1.
Size uncoated parts shall be coated sample size and the same dimensions and tolerances are shown in Table 1.
Loading and testing of components shown in Figure 2, the provisions applicable to the expected size of the component adhesive or cohesive strength of up to 50MPa coating test.
Note 1.The accuracy of this test method is very sensitive to the following factors. Preparation of coated sample, the polymer binder and the use of coated samples and uncoated
Appropriate member of the school straight. Poor alignment can lead to eccentric loads and bending stress on the test assembly, the test result is invalid.
Note 2.The bonding surface uncoated parts shall be roughened to aid polymer adhesive bond.
GB 23101.4-2008/ISO 13779-4.2002
1 --- for hydroxyapatite coating deposited on the surface.
Symbols and dimensions defined in Table 1.
4.2.4 polymeric binder
Bond strength greater than 30MPa at least a polymeric binder, or at least higher than the coating adhesion or cohesive strength 5MPa, whichever is larger value. Polymer binder used in all the test report shall be marked.
If the coating is porous portion extended to the coating/substrate interface should operate through careful selection of the coating and to ensure sufficient tack adhesive poly
Polymer binder will not penetrate through the coating to the substrate. Binder penetrate the substrate through the coating will cause the measured result is invalid.
Penetration of the binder can be assessed by examining the cross-section of the test piece.
4.3 coating sample preparation
Carefully prepared coating sample. Before assembly and test first checks all samples removed during the test found no special representative
Special coating sample. Hydroxyapatite coating on the coated samples not to be machined. Coating area to be tested shall be the nominal straight
Circular diameter of 25mm (tolerance Table 1) or an equivalent cross-sectional area. Measure and record the cross-sectional area and the sample size coating.
GB 23101.4-2008/ISO 13779-4.2002
If using a sample size or other cross-sectional area, and the resulting data should demonstrate a nominal diameter of 25mm circular sample data is relative
And the like.
4.4 School polymeric binder used and tested components straight
Polymer binder should be applied to the coating once the shaft center. It should control the amount of adhesive and the two parts are pressed together in order to maintain the force
Card in coated samples/uncoated part is not formed at the flange connection. Excess adhesive after curing, the application does not harm the integrity of the test assembly
A method for removing. The thickness of the adhesive on the contact surface and parallel test sectional uncoated parts with a binder/coating sample is contacted, and to ensure that
Uniform.
To ensure uniaxial testing component alignment during the curing polymeric binder, a required test straightening means rigidly fixed component. This
A device should be able to provide a constant load (typically 0.2MPa) to compensate for shrinkage of the adhesive during curing component.
Table 1 for bonding or adhesive strength of the coating is less than the 50MP Rao test fixture and sample size
Symbol test components/fixtures
Dimensions/mm
Titanium alloy (Ti6Al4V) 316 stainless steel
Ladder-type test piece 25 ± 0.05 25 ± 0.05
Straight test piece 25-0.02-0.041 25-0.02-0.041
Rao sample segment shank diameter 20-0.02-0.041 18-0.02-0.041
Chou pin hole diameter 9 + 8 + 0 0.015 0 0.015
Ba sample hole depth 14 ± 0.1 12 ± 0.1
Ladder-type test member 20 + 18 + 0 0.033 0 0.033
Straight test piece + 25 + 25 0.033 0.033 0 0
Ladder-type test piece 40 ± 0.1 35 ± 0.1
Straight test piece 45 ± 0.1 45 ± 0.1
K yoke inside of the depth of 40 ± 0.1 40 ± 0.1
4.5 Test procedure
The test assembly is placed in a tensile tester jig to match the direction of pull exerted by the longitudinal direction. Tensile load to (1.0 ± 0.01) mm/min constant speed is applied until the entire test assembly is fully extended. Record the maximum load is applied, accurate to 100N.
Use 10 × optical magnification examination to confirm the failure occurred within the coating rather than coating - adhesive contact surface.
4.6 Calculation of coating adhesion strength
Coating adhesion or cohesive strength sigma, in units of megapascals (MPa), with the following formula.
sigma = FA Where.
--- F. The maximum load, in units of bovine (the N);
A --- cross-sectional area (usually 490.87mm2), to calculate accurate ± 0.5mm2.
GB 23101.4-2008/ISO 13779-4.2002
5 hydroxyapatite coating tensile test report
The test report shall include the following.
a) based on national standards of this section;
b) indicate the test component materials, including polymeric binder component type and uncoated materials. Report for bonding uncoated table
Various surface roughness and surface treatment details of the test surface;
c) the number of test samples;
d) detailing the method used in the production of coated samples and coated samples with a variety of information related to connection, such as the date, lot number;
e) Each coated samples and uncoated components All dimensions, and cross-sectional area and thickness of the adhesive coating of hydroxyapatite;
f) All of the maximum load and fracture mode (for example, or between the coating and the substrate within the coating fracture);
g) the bond strength of the following values. maximum, minimum, average and standard deviation;
h) the extent of penetration of the binder into the substrate;
i) identify the fault is in the coating side of the substrate, rather than adhesive - substrate interface.
GB 23101.4-2008/ISO 13779-4.2002 references
[1] ISO 13779-1, Implantsforsurgery-Hydroxyapatite-Part 1.Ceramichydroxyapatite
[2] ISO 13779-2, Implantsforsurgery-Hydroxyapatite-Part 2.Coatingsofhydroxyapatite
[3] ASTMF1609.1995, Standardspecificationforcalciumphosphatecoatingsforimplantable materials
[4] ASTMF1185.1988, Standardspecificationforcompositionofceramichydroxyapatitefor surgicalimplants
[5] ASTM C633-79.1979, Standardtestmethodforadhesionorcohesionstrengthofflame sprayedcoatings
[6] BS5350-C15.1990, Methodsoftestforadhesives. Adhesivelybondedjoints. mechanical tests. Determinationofbondstrengthincompressiveshear
[7] JISH8666.1994, Testmethodsforceramicsprayedcoatings
[8] NFS94-072, Materialsforsurgicalimplants-Determinationoftensileadherenceofphosph- ocalciccoatingsforbiomedicalapplications
GB 23101.4-2008/ISO 13779-4.2002
GB 23101.2-2025 | GB/T 19701.1-2024 | GB/T 19701.2-2024 | GB/T 24627-2023 | GB/T 23101.4-2023 | […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 12 pages — is available in the English PDF.
Referenced standards
How to Download GB 23101.4-2008
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 23101.1-2008 — Implants for surgery -- Hydroxyapatite -- Part 1: Ceramic hydroxyapatite
GB 23101.2-2025 — Implants for surgery - Hydroxyapatite - Part 2: Thermally sprayed coatings of hydroxyapatite
GB 23101.3-2010 — Implants for surgery - Hydroxyapatite - Part 3: Chemical analysis and characterization of crystallinity ratio and phase purity